Fluke 9500B Oscilloscope Calibrator
- Fully automated, upgradeable calibration tool, offering hands-free, accurate oscilloscope calibration at a cost-effective price and performance level.
- Ultimate leveled sine performance up to 6.0 GHz with pulse edges of only 70 ps.
- Active Head Technologyâ„¢ enables adjustable output amplitude between 4.44 mV and 3.1V.
- Controlled waveshape filtering ensures accurate energy distribution regardless of chosen amplitude.
- Simultaneous output on 5 channels,no need for lead changing and complex signal multiplexers or impedance matching network.
- Description
- Additional information
- Specifications
Description
The Fluke 9500B Oscilloscope Calibrator is a high-performance, fully automated calibration solution designed for modern oscilloscopes.
Key Features:
- Fully automated, upgradeable calibration tool, offering hands-free, accurate oscilloscope calibration at a cost-effective price and performance level.
- Ultimate leveled sine performance up to 6.0 GHz with pulse edges of only 70 ps.
- Active Head Technologyâ„¢ enables adjustable output amplitude between 4.44 mV and 3.1V.
- Controlled waveshape filtering ensures accurate energy distribution regardless of chosen amplitude.
- Simultaneous output on 5 channels,no need for lead changing and complex signal multiplexers or impedance matching network.
- High Performance: Offers exceptional accuracy and performance to calibrate even the most demanding high-speed oscilloscopes.
- Full Automation: Automates the calibration process, significantly reducing manual effort and increasing efficiency.
- Upgradeable: Can be upgraded to support higher frequencies as your calibration needs evolve.
- Multi-Channel Capability: Supports simultaneous output on multiple channels, streamlining the calibration process.
- User-Friendly Interface: Intuitive software and easy-to-use controls make operation simple and efficient.
Applications:
- Calibration Laboratories: Ideal for calibrating a wide range of oscilloscopes in calibration laboratories.
- Electronics Manufacturing: Used in manufacturing environments for quality control and instrument calibration.
- Research and Development: Supports research and development activities in various fields.
The Fluke 9500B is a state-of-the-art calibration solution that sets a new standard for oscilloscope calibration. Its combination of high performance, full automation, and user-friendly features makes it an indispensable tool for calibration professionals and anyone who relies on accurate oscilloscope measurements.
Additional information
| Brand | Fluke Calibration |
|---|
Specifications: 9500B Oscilloscope Calibrator
| DC Voltage | ||
| Amplitude | ± 1 mV to ± 200V into 1M Ω ± 1 mV to ± 5V into 50 Ω |
|
| Accuracy | ± (0.025% + 25 µV) | |
| Ranging | 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous | |
| Deviation | ± 11.2% | |
| Squarewave | ||
| Amplitude | Range: 40 µV to 200V pk-pk into 1MΩ 40 µV to 5V pk-pk into 50Ω Polarity: Positive, negative or symmetrical about ground Accuracy (10 Hz to 10 kHz): <1 mV ± (1% + 10 µV) 1 mV-21 mV ± (0.10% + 10 µV) Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous Deviation: ± 11.2% |
|
| Rise/Fall Time | <100V <150 ns >=100V <200 ns |
|
| Aberrations | <2% peak for first 500 ns | |
| Frequency | Range: 10 Hz to 100 kHz Accuracy: ± 0.25 ppm Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous |
|
| Low-Edge Pulse | ||
| Amplitude | Range: 5 mV to 3V pk-pk into 50Ω Accuracy: +50 ps to -150 ps Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous |
|
| Rise/Fall Time | 500 ps return to ground | |
| Mk/Sp Ratio | 1:9 | |
| Aberrations (into VSWR 1.2:1) | <± pk in 8Ghz <± 1.5% pk in 3Ghz (First 10ns) |
|
| Frequency | Range: 10 Hz to 2 Mhz Accuracy: ± 0.25 ppm Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous |
|
| High-Edge Pulse | ||
| Amplitude | Range: 1mV to 200V pk-pk into 1 MΩ 1mV to 5V pk-pk into 50Ω Accuracy: ± 3% Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous |
|
| Rise/Fall Time | <100V <150 ns >=100V <200 ns |
|
| Mk/Sp Ratio | 1:1 | |
| Aberrations | <2% peak for first 500ns | |
| Frequency | Range: 10 Hz to 100 kHz Accuracy: ± 0.25 ppm Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous |
|
| Fast-Edge (only available on 9530 Active Heads) | |
| Amplitude | Range: 5 mV to 3V pk-pk into 50Ω Accuracy: ± 3% Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous |
| Rise/Fall Time | 150 ps return to ground |
| Mk/Sp Ratio | 1:9 |
| Aberrations | <3% peak in 8 Ghz <2% peak in 3 Ghz (First Ins) |
| Frequency | Range: 10 Hz to 2 Mhz Accuracy: ± 0.25 ppm Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous |
| 70ps Fast-Edge (only available on 9560 Active Heads) | |
| Amplitude | Range: 25 mV to 2V pk-pk Accuracy: ± 3% Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous |
| Rise/Fall Time | 70 ps return to ground |
| Mk/Sp Ratio | 1:9 |
| Aberrations | <± 4% peak in 20 Ghz <3% peak in 8 Ghz <1% peak in 3 Ghz (First 700 ps) |
| Frequency | Range: 10 Hz to 1 Mhz Accuracy: ± 0.25 ppm Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous |
| 25ps Fast-Edge (only available on 9550 Active Heads) | ||
| Amplitude | Range: 425 to 575 mV pk-pk into 50Ω Accuracy: ± 2% |
|
| Rise/Fall Time | 25 ps return to ground | |
| Mk/Sp Ratio | 1:9 | |
| Timing Markers | ||
| Styles | Square/Sine, Pulse or Narrow Triangle | |
| Square/Sine | Period Square:Â 10 ns to 55s Period Sine:Â 450 ps to 10ns 9500/600 = 1ns to 10ns 9560 = 180ps to 10ns |
|
| Pulse | Period: 1 µs to 55s Rise/Fall Time: 2.5% of period |
|
| Narrow Triangle | ||
| Ranging | 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous for period of all waveshapes | |
| Timing Accuracy | Normal: ± 0.25 ppm | |
| Timing Jitter | 10 ps pk-pk | |
| Deviation | ± 45% for period | |
| Amplitude | 100 mV to 1V pk-pk | |
| Sub-Division | Every 10th marker can be set to higher amplitude for periods ± 1 µs for all waveshapes | |
| Leveled Sine and Dual Sine | ||
| Frequency | Range:Â 9500B/600 0.1 Hz to 600 MHz 9500B/1100 0.1 Hz to 1.1 GHz 9500B/3200 0.1 Hz to 3.2 GHz 9500B/2200+9560 0.1Hz to 6.4 GHz |
|
| Amplitude (Leveled Sine into 50Ù) | 0.1 Hz – 550 MHz 4.44 mV to 5.560V pk-pk 550 MHz – 2.5 GHz 4.44 mV to 3.336V pk-pk 2.5 GHz – 3.2 GHz 4.44 mV to 2.224V pk-pk 3.2 GHz – 6.4 GHz 25mV to 2 V pk-pk Accuracy: ± 1.5% at 50 kHz |
|
| Flatness (Leveled Sine relative to 50kHz) | 0.1 Hz – 300 MHz ± 2% 300 MHz – 550 MHz ± 3% 550 MHz – 1.1 GHz ± 4% 1.1 GHz – 3.2 GHz ± 5% 3 GHz – 6.0 GHz ± 5% |
|
| Ranging | 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous | |
| Sine Purity | 2nd Harmonic <-35 dBc 3rd Harmonic <-40 dBc All Other Spurious Signals <-40 dBc (typical) |
|
| Input Impedance | ||
| Resistance Measurement (not available on 9550) | Range: 10Ω – 150Ω and 50 kΩ – 12 MΩ Accuracy: 10 – 40 (W) ± 0.5% 40 – 90 ± 0.1% 90 – 150 ± 0.5% 50k – 800k ± 0.5% 800k – 1.2M ± 0.1% 1.2M – 12M ± 0.5% |
|
| Capacitance Measurement (not available on 9550 or 9560 heads) | Range: 1 pF to 95 pF Accuracy: 1 pF – 35 pF 2% ± 0.25 pF 35 pF – 95 pF 3% ± 0.25 pF |
|
| Current | ||
| Amplitude | DC: ± 100 µA to ± 100 mA Squarewave: 100 µA to 100 mA pk-pk |
|
| Accuracy | ± (0.25% + 0.5 µA) | |
| Frequency |      10 Hz to 100 kHz Accuracy: ± 0.25 ppm Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous |
|
| Composite Video Output | ||
| Amplitude | 1.0V, 0.7V and 0.3V | |
| Pattern | White, gray or black | |
| Sync Polarity | Positive or negative | |
| Standards | 625-line 50 Hz or 525-line 60 Hz | |
| LFLinear Ramp | ||
| Waveforms | 1V pk-pk sym. triangle | |
| Ramp Time | 1 ms to 1s | |
| Overload Pulse | ||
| Amplitude | 5V to 20V into 50Ω | |
| Polarity | Positive or negative | |
| Duration | 0.2s to 100s | |
| Trigger | Manual | |
| Zero Skew | ||
| Unadjusted Skew | ± 25 ps ch to ch | |
| Adjusted Skew | ± 5 ps ch to ch | |
| Frequency Range | 10 Hz to 100 MHz | |
| Short/Open Output | ||
| Output Leakage | Open Circuit: ± 50 pA Short Circuit: ± 15 µV |
|
| Auxiliary Input | ||
| Signal Routing | Rear i/p to any Active Head | |
| Maximum Input | Voltage: ± 40V pk-pk Current: ± 400 mA pk-pk |
|
| Rate | User Selectable:Â f (up to 120 MHz), f/10 or f/100 Free Run:Â 100 Hz |
|
| Reference Frequency Input | Frequency Range: 1 MHz to 20 MHz in 1 MHz steps Level: 90 mV to 1V pk-pk typ. Lock Range: ± 50 ppm |
|
| Reference Frequency Output | Frequency:Â 1 MHz or 10 MHz Level:Â Into 50W: 1V pk-pk (typical) |
|
| General Specifications | ||
| Temperature | Operating: 5°C to 40°C Storage: 0°C to 50°C |
|
| Humidity (non-condensing) | Operating: <90% over 5°C to 30°C <75% over 30°C to 40°C Storage:  <95% over 0°C to 50°C |
|
| Power, battery life | Voltage:Â 95V to 132V rms or 209V to 264V rms Frequency:Â 48 Hz to 63 Hz Consumption:Â 400 VA |
|
| Warm up Time | 20 minutes | |
| Calibration Documentation | 17025 accredited report of calibration included | |
| Dimensions | Model 9500B Mainframe: H x W x D        133 x 427 x 440 mm (5.24 x 16.8 x 17.3 ins) Weight          12 kg (27 lbs) approx. Modules 9510, 9530, 9550: H x W x D        65 x 31 x 140 mm (2.56 x 1.22 x 5.51 ins) Weight:          0.45 kg (1-lb) approx. |
|
| Safety |       Designed to UL3111 and EN61010-1-1:1993/A2:1995. CE Marked EMC (including options) Emissions:  EN55011/22 Immunity:  EN50082-1:1992 FCC Rules part 15 sub-part J class B |
|