Fluke 9500B Oscilloscope Calibrator

  • Fully automated, upgradeable calibration tool, offering hands-free, accurate oscilloscope calibration at a cost-effective price and performance level.
  • Ultimate leveled sine performance up to 6.0 GHz with pulse edges of only 70 ps.
  • Active Head Technologyâ„¢ enables adjustable output amplitude between 4.44 mV and 3.1V.
  • Controlled waveshape filtering ensures accurate energy distribution regardless of chosen amplitude.
  • Simultaneous output on 5 channels,no need for lead changing and complex signal multiplexers or impedance matching network.

Description

The Fluke 9500B Oscilloscope Calibrator is a high-performance, fully automated calibration solution designed for modern oscilloscopes.

Key Features:

  • Fully automated, upgradeable calibration tool, offering hands-free, accurate oscilloscope calibration at a cost-effective price and performance level.
  • Ultimate leveled sine performance up to 6.0 GHz with pulse edges of only 70 ps.
  • Active Head Technologyâ„¢ enables adjustable output amplitude between 4.44 mV and 3.1V.
  • Controlled waveshape filtering ensures accurate energy distribution regardless of chosen amplitude.
  • Simultaneous output on 5 channels,no need for lead changing and complex signal multiplexers or impedance matching network.
  • High Performance: Offers exceptional accuracy and performance to calibrate even the most demanding high-speed oscilloscopes.
  • Full Automation: Automates the calibration process, significantly reducing manual effort and increasing efficiency.
  • Upgradeable: Can be upgraded to support higher frequencies as your calibration needs evolve.
  • Multi-Channel Capability: Supports simultaneous output on multiple channels, streamlining the calibration process.
  • User-Friendly Interface: Intuitive software and easy-to-use controls make operation simple and efficient.

Applications:

  • Calibration Laboratories: Ideal for calibrating a wide range of oscilloscopes in calibration laboratories.
  • Electronics Manufacturing: Used in manufacturing environments for quality control and instrument calibration.
  • Research and Development: Supports research and development activities in various fields.

The Fluke 9500B is a state-of-the-art calibration solution that sets a new standard for oscilloscope calibration. Its combination of high performance, full automation, and user-friendly features makes it an indispensable tool for calibration professionals and anyone who relies on accurate oscilloscope measurements.

Additional information

Brand

Fluke Calibration

Specifications: 9500B Oscilloscope Calibrator

DC Voltage
Amplitude ± 1 mV to ± 200V into 1M Ω
± 1 mV to ± 5V into 50 Ω
Accuracy ± (0.025% + 25 µV)
Ranging 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Deviation ± 11.2%
Squarewave
Amplitude Range: 40 µV to 200V pk-pk into 1MΩ
40 µV to 5V pk-pk into 50Ω
Polarity: Positive, negative or symmetrical about ground
Accuracy (10 Hz to 10 kHz): <1 mV ± (1% + 10 µV)
1 mV-21 mV ± (0.10% + 10 µV)
Ranging:  1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Deviation: ± 11.2%
Rise/Fall Time <100V <150 ns
>=100V <200 ns
Aberrations <2% peak for first 500 ns
Frequency Range: 10 Hz to 100 kHz
Accuracy: ± 0.25 ppm
Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous

 

Low-Edge Pulse
Amplitude Range: 5 mV to 3V pk-pk into 50Ω
Accuracy: +50 ps to -150 ps
Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Rise/Fall Time 500 ps return to ground
Mk/Sp Ratio 1:9
Aberrations (into VSWR 1.2:1) <± pk in 8Ghz
<± 1.5% pk in 3Ghz
(First 10ns)
Frequency Range: 10 Hz to 2 Mhz
Accuracy: ± 0.25 ppm
Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
High-Edge Pulse
Amplitude Range: 1mV to 200V pk-pk into 1 MΩ
1mV to 5V pk-pk into 50Ω
Accuracy: ± 3%
Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Rise/Fall Time <100V <150 ns
>=100V <200 ns
Mk/Sp Ratio 1:1
Aberrations <2% peak for first 500ns
Frequency Range: 10 Hz to 100 kHz
Accuracy: ± 0.25 ppm
Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous

 

Fast-Edge (only available on 9530 Active Heads)
Amplitude Range: 5 mV to 3V pk-pk into 50Ω
Accuracy: ± 3%
Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Rise/Fall Time 150 ps return to ground
Mk/Sp Ratio 1:9
Aberrations <3% peak in 8 Ghz
<2% peak in 3 Ghz
(First Ins)
Frequency Range: 10 Hz to 2 Mhz
Accuracy: ± 0.25 ppm
Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
70ps Fast-Edge (only available on 9560 Active Heads)
Amplitude Range: 25 mV to 2V pk-pk
Accuracy: ± 3%
Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Rise/Fall Time 70 ps return to ground
Mk/Sp Ratio 1:9
Aberrations <± 4% peak in 20 Ghz
<3% peak in 8 Ghz
<1% peak in 3 Ghz
(First 700 ps)
Frequency Range: 10 Hz to 1 Mhz
Accuracy: ± 0.25 ppm
Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous

 

25ps Fast-Edge (only available on 9550 Active Heads)
Amplitude Range: 425 to 575 mV pk-pk into 50Ω
Accuracy: ± 2%
Rise/Fall Time 25 ps return to ground
Mk/Sp Ratio 1:9
Timing Markers
Styles Square/Sine, Pulse or Narrow Triangle
Square/Sine Period Square: 10 ns to 55s
Period Sine: 450 ps to 10ns
9500/600 = 1ns to 10ns
9560 = 180ps to 10ns
Pulse Period: 1 µs to 55s
Rise/Fall Time: 2.5% of period
Narrow Triangle
Ranging 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous for period of all waveshapes
Timing Accuracy Normal: ± 0.25 ppm
Timing Jitter 10 ps pk-pk
Deviation ± 45% for period
Amplitude 100 mV to 1V pk-pk
Sub-Division Every 10th marker can be set to higher amplitude for periods ± 1 µs for all waveshapes

 

Leveled Sine and Dual Sine
Frequency Range: 9500B/600 0.1 Hz to 600 MHz
9500B/1100 0.1 Hz to 1.1 GHz
9500B/3200 0.1 Hz to 3.2 GHz
9500B/2200+9560 0.1Hz to 6.4 GHz
Amplitude (Leveled Sine into 50Ù) 0.1 Hz – 550 MHz 4.44 mV to 5.560V pk-pk
550 MHz – 2.5 GHz 4.44 mV to 3.336V pk-pk
2.5 GHz – 3.2 GHz 4.44 mV to 2.224V pk-pk
3.2 GHz – 6.4 GHz 25mV to 2 V pk-pk
Accuracy: ± 1.5% at 50 kHz
Flatness (Leveled Sine relative to 50kHz) 0.1 Hz – 300 MHz ± 2%
300 MHz – 550 MHz ± 3%
550 MHz – 1.1 GHz ± 4%
1.1 GHz – 3.2 GHz ± 5%
3 GHz – 6.0 GHz ± 5%
Ranging 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Sine Purity 2nd Harmonic <-35 dBc
3rd Harmonic <-40 dBc
All Other Spurious Signals <-40 dBc (typical)
Input Impedance
Resistance Measurement (not available on 9550) Range: 10Ω – 150Ω and 50 kΩ – 12 MΩ
Accuracy: 10 – 40 (W) ± 0.5%
40 – 90 ± 0.1%
90 – 150 ± 0.5%
50k – 800k ± 0.5%
800k – 1.2M ± 0.1%
1.2M – 12M ± 0.5%
Capacitance Measurement (not available on 9550 or 9560 heads) Range: 1 pF to 95 pF
Accuracy: 1 pF – 35 pF 2% ± 0.25 pF
35 pF – 95 pF 3% ± 0.25 pF

 

Current
Amplitude DC: ± 100 µA to ± 100 mA
Squarewave: 100 µA to 100 mA pk-pk
Accuracy ± (0.25% + 0.5 µA)
Frequency      10 Hz to 100 kHz
Accuracy: ± 0.25 ppm
Ranging: 1, 2, 5 or 1, 2, 2.5, 4, 5 or continuous
Composite Video Output
Amplitude 1.0V, 0.7V and 0.3V
Pattern White, gray or black
Sync Polarity Positive or negative
Standards 625-line 50 Hz or 525-line 60 Hz

 

LFLinear Ramp
Waveforms 1V pk-pk sym. triangle
Ramp Time 1 ms to 1s
Overload Pulse
Amplitude 5V to 20V into 50Ω
Polarity Positive or negative
Duration 0.2s to 100s
Trigger Manual

 

Zero Skew
Unadjusted Skew ± 25 ps ch to ch
Adjusted Skew ± 5 ps ch to ch
Frequency Range 10 Hz to 100 MHz
Short/Open Output
Output Leakage Open Circuit: ± 50 pA
Short Circuit: ± 15 µV

 

Auxiliary Input
Signal Routing Rear i/p to any Active Head
Maximum Input Voltage: ± 40V pk-pk
Current: ± 400 mA pk-pk
Rate User Selectable: f (up to 120 MHz), f/10 or f/100
Free Run: 100 Hz
Reference Frequency Input Frequency Range: 1 MHz to 20 MHz in 1 MHz steps
Level: 90 mV to 1V pk-pk typ.
Lock Range: ± 50 ppm
Reference Frequency Output Frequency: 1 MHz or 10 MHz
Level: Into 50W: 1V pk-pk (typical)
General Specifications
Temperature Operating: 5°C to 40°C
Storage: 0°C to 50°C
Humidity (non-condensing) Operating: <90% over 5°C to 30°C
<75% over 30°C to 40°C
Storage:   <95% over 0°C to 50°C
Power, battery life Voltage: 95V to 132V rms or 209V to 264V rms
Frequency: 48 Hz to 63 Hz
Consumption: 400 VA
Warm up Time 20 minutes
Calibration Documentation 17025 accredited report of calibration included
Dimensions Model 9500B Mainframe:
H x W x D        133 x 427 x 440 mm
(5.24 x 16.8 x 17.3 ins)
Weight           12 kg (27 lbs) approx.
Modules 9510, 9530, 9550:
H x W x D        65 x 31 x 140 mm
(2.56 x 1.22 x 5.51 ins)
Weight:          0.45 kg (1-lb) approx.
Safety       Designed to UL3111 and EN61010-1-1:1993/A2:1995.
CE Marked
EMC (including options)
Emissions:  EN55011/22
Immunity:  EN50082-1:1992
FCC Rules part 15 sub-part J class B

 

Related Products