Fluke 5730A High Performance Multifunction Calibrator
- Calibrate up to long-scale 8.5 digit digital multimeters and RF voltmeters with wideband option.
- Improved AC voltage, AC current, and resistance specifications for increased test uncertainty ratios (TURs).
- 6.5-inch VGA touchscreen, a full-color interface, and Visual Connection Managementâ„¢ output terminals
- Full Met/CAL® compatibility with 5700A and 5720A procedures
- Description
- Additional information
- Specifications
Description
The Fluke 5730A is a high-performance, multifunction calibrator designed for precise and efficient electrical measurements.
Key Features & Benefits:
- Enhanced Accuracy: Offers exceptional accuracy for DC/LF measurements, making it suitable for calibrating demanding instruments like long-scale digital multimeters and RF voltmeters.
- Versatile Functionality: Provides a wide range of measurement capabilities, including:
- DC/LF Voltage: High-accuracy DC and low-frequency AC voltage outputs.
- DC/LF Current: Precise DC and low-frequency AC current outputs.
- Resistance: Measures resistance with high accuracy.
- Optional Wideband AC Voltage: Extends AC voltage measurements to higher frequencies.
- User-Friendly Interface: Features a large, full-color touchscreen display for easy navigation and operation.
- Improved Usability: Incorporates modern technology and design for enhanced user experience.
- Met/CAL® Compatibility: Seamlessly integrates with Fluke’s Met/CAL® calibration software for streamlined calibration processes.
Applications:
- Calibration Laboratories: Ideal for calibrating a wide range of electrical instruments, including DMMs, oscilloscopes, and more.
- Research and Development: Supports various research and development activities in fields like electronics, telecommunications, and aerospace.
- Manufacturing: Used in manufacturing environments for quality control and instrument calibration.
The Fluke 5730A represents the next generation of high-performance calibrators, offering improved accuracy, enhanced usability, and a wide range of measurement capabilities. It is a valuable tool for calibration laboratories, research institutions, and industries that require precise and reliable electrical measurements.
Additional information
Brand | Fluke Calibration |
---|
Specifications: 5730A High Performance Multifunction Calibrator
Technical Specifications | ||
Voltage DC | Range | 0 to ±1100 V |
Best 1 Year, 95% Specification | 3.5 ppm + 2.5 µV | |
Voltage AC | Range | 22 µV to 1100V |
10 Hz to 1 MHz | ||
Best 1 Year, 95% Specification: | 42 ppm + 8 µV | |
Resistance | Range | 0 to 100 MΩ, 18 values in x1 and x1.9 |
Best 1 Year, 95% Specification | 6.5 ppm | |
Current DC | Range | 0 to ±2.2A (0 to ±11A with 5725A; 0 to ±100 A with 52120A) |
Best 1 Year, 95% Specification | 35 ppm + 7 nA | |
Current AC | Range | 9 µA to 2.2A, 10 Hz to 10 kHz (9 µA to ±11A with 5725A; 9 µA to ± 120 A with 52120A) |
Best 1 Year, 95% Specification | 103 ppm + 8 nA | |
Wideband ac Voltage option | Range | 300 µV to 3.5V |
10 Hz to 50 MHz | ||
Best 1 Year, 99% Specification | 0.4% + 500 µV | |
General Specifications | ||
Warm up time | Twice the time since last warmed up, to a maximum of 30 minutes | |
Settling time | Less than 5 seconds for all functions and ranges except as noted | |
Standard interfaces | IEEE-488 (GPIB), RS-232, USB 2.0 Device, Ethernet, 5725A, 52120A, phase lock in (BNC), phase reference out (BNC) | |
Temperature performance | Operating | 0 °C to 50 °C |
Calibration:15 °C to 35 °C | 15 °C to 35 °C | |
Storage | -40 °C to 75 °C | |
Operating altitude | 2000 m maximum | |
Relative humidity | Operating | < 80% to 30 °C, < 70% to 40 °C, < 40% to 50 °C |
Storage | < 95%, non-condensing. | |
Safety | IEC 61010-1: 300 V CAT II, Pollution Degree 2 | |
Analog low isolation | 20 V | |
Electromagnetic environment | IEC 61326-1: Controlled | |
Line power | 47 Hz to 63 Hz; ±10% 100 V, 110 V, 115 V, 120 V, 200 V, 220 V, 230 V, 240 V | |
Power consumption | 300 VA | |
Dimensions | Height | 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet |
Width | 43.2 cm (17 in), standard rack width | |
Depth | 64.8 cm (25.5 in), overall; 59.4 cm (23.4 in), rack depth | |
Weight | 27 kg: (62 lbs.) | |
Absolute uncertainty definition | 5730A uncertainty specifications include stability, temperature coefficient, linearity, load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated | |
Specification confidence interval | 99% and 95% |