Fluke 5618B Small Diameter Industrial RTD

  • Secondary level performance with full ITS-90 calibration.
  • Sheath diameter of 3.2mm (0.125 in), ensuring reduced response time without compromising precision.
  • NVLAP-accredited calibration from -200°C to 500°C, lab code 200706-0.
  • Reduced measurement error in short immersion depth applications by conducting less heat between ambient and the sensor.
Category:

Description

The Fluke 5618B is a high-accuracy, portable PRT (Platinum Resistance Thermometer) specifically designed for demanding temperature calibration applications. It’s a robust and reliable instrument often used as a reference thermometer in calibration laboratories, industrial settings, and metrology applications. Its portability and accuracy make it valuable for on-site calibrations and comparisons.

Key Features:

  • High Accuracy: Offers exceptional temperature measurement accuracy, crucial for precise calibration work. Typical accuracy specifications are very tight, often in the millidegree Celsius range.
  • Wide Temperature Range: Covers a broad temperature range, allowing for calibration and measurement across a wide spectrum of temperatures. The specific range depends on the probe configuration.
  • Rugged and Portable: Designed for field use as well as laboratory work. Its robust construction ensures it can withstand the rigors of transportation and demanding environments.
  • Interchangeable PRT Probes: Works with a variety of interchangeable PRT probes, allowing users to select the optimal probe for their specific temperature range and application. This modularity adds flexibility.
  • Readout Options: Can be used with compatible readout devices or systems, allowing for data logging, display, and analysis. It’s designed to interface with precision measurement instruments.
  • Calibration Certificate: Supplied with a calibration certificate, providing traceability and confidence in the accuracy of the instrument.

Applications:

  • Calibration Laboratories: As a reference PRT for calibrating other temperature instruments and sensors.
  • Metrology: For high-precision temperature measurements in research and metrology applications.
  • Industrial Calibration: On-site calibration of temperature sensors and instruments in various industries.
  • Pharmaceutical and Biotech: Temperature calibration in sensitive pharmaceutical and biotech processes.

In summary, the Fluke 5618B is a high-precision, portable PRT designed for professionals who require accurate and reliable temperature measurements for calibration and metrology applications. Its robust design, interchangeable probes, and traceable accuracy make it a valuable tool in various industries.

Additional information

Brand

Fluke Calibration

Specifications: 5618B Small Diameter Industrial RTD

Specifications
Resistance Nominal 100 Ω at 0 °C
Temperature
Coefficient
0.003923 Ω/Ω/ °C nominal
Temperature
Range
–200 °C to 500 °C
(–200 °C to 300 °C for 5618B-6-X)
Drift Rate ± 0.1 °C when used periodically to 500 °C
Sheath Material 316 SST
Leads 22 AWG PTFE, 6’
Termination Specify
Hysteresis
 
Less than 0.01 °C at 0 °C when using –196 °C and
420 °C as the end points.
Time Constant
 
 
Four seconds maximum for 63.2 % response to
step change in water moving
at 3 fps.
Thermal EMF Less than 25 mV at 420 °C
Calibration
 
 
Includes manufacturer’s NVLAP-accredited
calibration w/ITS-90 coefficients, R vs. T values in
1 °C increments, lab code 200706-0
Size
 
 
5618B-12: 305 mm L x 3.2 mm dia. (12 x 1/8 in)
5618B-9: 229 mm L x 3.2 mm dia. (9 x 1/8 in)
5618B-6: 152 mm L x 3.2 mm dia. (6 x 1/8 in)
Calibrated
Accuracy† (k=2)
 
± 0.026 °C at –200 °C
± 0.066 °C at 0 °C
± 0.219 °C at 500 °C
†Includes
calibration
uncertainty and
drift specification.
NVLAP† Calibration Uncertainty
Temperature Expanded
Uncertainty
(k=2)
–196 °C 0.024 °C
–38 °C 0.011 °C
0 °C 0.010 °C
200 °C 0.018 °C
420 °C‡ 0.029 °C
Note: Calibration uncertainties depend on the uncertainties of the lab performing the calibration. Subsequent calibrations of this probe performed with different processes, at different facilities, or with changed uncertainty statements may state different uncertainties. †Lab code 200706-0 ‡300 °C for 5618B-6

 