Fluke 5649/5650 | Type R and Type S Thermocouple Standards

  • Type R and S Thermocouples.
  • Platinum and platinum-rhodium alloy.
  • Temperature range of 0 °C to 1450 °C.
  • Uncertainties as good as 0.15 °C over most of that range.
  • Calibrated at four fixed points (Sn, Zn, Al, Ag) and then extrapolated to 1450 °C.
  • Four version avaialbe, 20 and 25 inch lengths with and without CJC.
Category:

Description

The Fluke 5649/5650 are high-accuracy, industrial-grade thermocouple standards. They are used for precise temperature measurement in a variety of applications.

Key features:

  • Type R and S thermocouples: These thermocouple types are known for their high stability and accuracy at high temperatures.
  • Wide temperature range: They cover a temperature range from 0 °C to 1450 °C.
  • Low uncertainties: They offer uncertainties as low as 0.15 °C over most of their range.
  • Durable construction: The thermocouples are encased in a rugged alumina sheath for protection.
  • Various lengths and configurations: They are available in different lengths and with or without a reference junction to suit different needs.

Benefits:

  • High accuracy: They provide highly accurate temperature measurements, which are essential for many industrial processes.
  • Stability: They offer excellent stability over time, ensuring reliable measurements.
  • Versatility: They can be used in a wide range of applications, including high-temperature furnaces, heat treatment, and calibration.

Applications:

  • Calibration laboratories: Used as reference standards for calibrating other temperature sensors.
  • Industrial processes: Used for precise temperature control in processes such as heat treatment, ceramics, and metalworking.
  • Research and development: Used in scientific experiments where high accuracy and stability are required.

Overall, the Fluke 5649/5650 are high-quality thermocouple standards that provide accurate and reliable temperature measurements for demanding applications.

Additional information

Brand

Fluke Calibration

Specifications: 5649/5650 | Type R and Type S Thermocouple Standards

Specifications
Range 0 °C to 1450 °C
Type Platinum/13 % Rhodium vs. platinum (type R)
Platinum/10 % Rhodium vs. platinum (type S)
Calibration Fixed point calibration uncertainties
(k=2): ± 0.15 °C up to 962 °C,
increasing linearly to ± 2.0 °C at 1450 °C
Hot Junction
Sheath Dimensions
6.35 mm (0.25 in) diameter;
see Ordering Information for lengths
Reference Junction Sheath Dimensions 4.8 mm dia. x 210 mm long (0.188 x 8.25 in)
Long-Term Stability ± 0.5 °C to 1100 °C
± 2.0 °C to 1450 °C (over one year depending on usage)
Short-Term Stabilities ± 0.2 °C to 1100 °C
± 0.6 °C to 1450 °C
Immersion At least 152 mm (6 in) recommended
Protective Case Model 2609 case included
Weight 1 kg (2 lb)

 

Related Products