Fluke 5615 Secondary Reference Temperature Standards

  • Inconel™ 600 sheath that’s 305 mm (12 in) long and 6.35 mm (0.250 in) in diameter
  • Calibrated accuracy of ± 0.012 °C at 0 °C.
  • Reference-grade platinum sensing element.
  • NVLAP-accredited calibration, lab code 200706-0.
  • Durable sensors, ideal in the factory, field, or lab.
Category:

Description

The Fluke 5615 is a secondary reference temperature standard, sometimes called a Platinum Resistance Thermometer (PRT). This means it’s a very accurate thermometer used to calibrate other thermometers and temperature measurement devices.

Here’s a breakdown of what that means:

  • High Accuracy: The 5615 is designed to measure temperatures with very high precision. This is essential for ensuring the accuracy of other temperature measuring instruments.
  • Stability: It maintains its accuracy over time, providing consistent and reliable readings. This is crucial for long-term measurements and calibrations.
  • Platinum Sensing Element: At its core is a very pure platinum element. Platinum’s electrical resistance changes predictably with temperature, making it ideal for accurate temperature measurement.
  • Durable Construction: It’s built to withstand the demands of industrial environments, with a protective sheath (often made of Inconel) to protect the delicate sensing element.
  • Calibration: It comes with a calibration certificate from a recognized accreditation body (NVLAP), ensuring its accuracy is traceable to national standards.

Here’s where you might find a Fluke 5615:

  • Calibration Labs: Used to calibrate other temperature sensors and instruments.
  • Industrial Processes: Monitoring and controlling critical temperatures in manufacturing, pharmaceuticals, and other industries.
  • Research and Development: Ensuring accurate temperature measurements in experiments.

Additional information

Brand

Fluke Calibration

Specifications: 5615 Secondary Reference Temperature Standards

Specifications
Temperature range
 
5615-12 and 5615-9: –200 °C to 420 °C
5615-6: –200 °C to 300 °C
Nominal resistance at 0 °C 100 Ω ± 0.10 Ω
Temperature coefficient 0.0039250 Ω/Ω/°C
Accuracy[1] ± 0.024 °C at –200 °C
± 0.012 °C at 0 °C
± 0.035 °C at 420 °CC
Short-term repeatability[2] ± 0.009 °C at 0.010 °C
Drift[3] ± 0.007 °C at 0.010 °C
Sensor length 28 mm (1.1 in)
Sensor location 6.9 mm ± 3.3 mm from tip (0.27 in ± 0.13 in)
Sheath dimensions 5615-6: 152 mm x 4.76 mm (6.0 in x 0.188 in)
5615-9: 229 mm x 4.76 mm (9.0 in x 0.188 in)
5615-12: 305 mm x 6.35 mm (12.0 in x 0.250 in)
Sheath diameter tolerance ± 0.127 mm (± 0.005 in)
Sheath material Inconel™ 600
Minimum insulation resistance 1000 MΩ at 23 °C
Transition junction temperature range[4] –50 °C to 200 °C
Transition junction dimensions 71 mm x 13 mm dia (2.8 in x 0.5 in)
Maximum immersion length 5615-6: 102 mm (4 in)
5615-9: 178 mm (7 in)
5615-12: 254 mm (10 in)
Response time[5] 9 seconds typical
Self heating (in 0 °C bath) 50 mW/°C
Lead-wire cable type PTFE insulated with PTFE jacket, 22 AWG
Lead-wire length 183 cm (72 in or 6 ft)
Lead-wire temperature range –50 °C to 200 °C
Calibration NVLAP-accredited calibration included, lab code 200706-0. Please see calibration uncertainty table and its explanation of changeable uncertainties.
[1] Includes calibration and 100 hr drift (k=2)

[2] Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k=2)

[3] After 100 hrs at max temp, 95 % confidence (k=2)

[4] Temperatures outside this range will cause irreparable damage. For best performance, transition junction should not be too hot to touch.

[5] Per ASTM E 644

NVLAP† Calibration Uncertainty
Temperature Expanded Uncertainty (k=2)
–196 °C 0.024 °C
–38 °C 0.011 °C
0 °C 0.010 °C
200 °C 0.018 °C
420 °C 0.029 °C
Note: Calibration uncertainties depend on the uncertainties of the lab performing the calibration. Subsequent calibrations of this probe performed with different processes, at different facilities, or with changed uncertainty statements may state different uncertainties. †Lab code 200706-0 ‡5615-6 excluded

 