Fluke 5322A Electrical Safety Tester Calibrator
- Calibrates a wide range of equipment from Hipot testers to portable appliance testers.
- Many functions in one instrument. Variable high voltage resistance outputs, Active Loop Compensation, a 600 V source, 4-wire low resistance sources, and optional high voltage load resistors.
- Simulate RCD, display Ripple coefficient and THD measurements, with multiple models available.
- Automate with MET/CAL software for improved consistency.
- Description
- Additional information
- Specifications
Description
The Fluke 5322A is a versatile and powerful calibration tool designed to simplify and streamline the calibration process for a wide range of electrical safety testers. Here’s an overview:
Key Features & Capabilities:
Multifunctionality:
- Calibrates a wide range of equipment from Hipot testers to portable appliance testers.
- Many functions in one instrument. Variable high voltage resistance outputs, Active Loop Compensation, a 600 V source, 4-wire low resistance sources, and optional high voltage load resistors.
- Simulate RCD, display Ripple coefficient and THD measurements, with multiple models available.
- Automate with MET/CAL software for improved consistency.
- Hipot Testing: Calibrates AC and DC hipot testers with high accuracy.
- Insulation Resistance Testing: Simulates various resistance values for calibrating insulation testers (megohmeters).
- Loop Impedance & Ground Bond Testing: Provides precise resistance sources for calibrating loop impedance testers and ground bond testers.
- RCD Testing: Simulates residual current devices (RCDs) for calibration.
- High Accuracy: Ensures accurate and reliable calibration results for your electrical safety equipment.
Benefits:
- Improved Efficiency: Streamlines calibration processes, saving time and resources.
- Enhanced Accuracy: Ensures accurate and reliable calibration results, leading to more accurate and safer electrical safety testing.
- Reduced Costs: Can help reduce the overall cost of ownership by minimizing downtime and improving equipment performance.
- Increased Productivity: Enables faster turnaround times for calibrations, improving overall productivity.
The Fluke 5322A is a valuable asset for calibration laboratories, electrical safety professionals, and anyone involved in maintaining and calibrating electrical safety equipment. Its comprehensive capabilities, high accuracy, and user-friendly interface make it a reliable and efficient solution for ensuring the accuracy and reliability of critical safety equipment.
Additional information
Brand | Fluke Calibration |
---|
Specifications: 5322A Electrical Safety Tester Calibrator
Complete 5322A specifications
General specifications | |
Specifications confidence levell | 99Â % |
Specifications interval | 1 year |
Power line | 115/230 V ac (50/60 Hz) +10 % / -14 %, with the maximum voltage difference between neutral and protective earth not exceeding 15 V. Operation with power line between -10 % and -14 % has limitations in burden current for voltage outputs. See AC/DC Voltage Calibrator (VLC option) below. |
Power consumption | 1250 VA maximum |
Fuse protection | AC mains input: 2 A, 250 V for 230 V, time delay (T2L250 V – 5 mm x 20 mm) 4 A, 250 V for 115 V, time delay (T4L250 V – 5 mm x 20 mm) |
RCD input: 3.15 A, 250 V, fast (F3.15H250 V – 5 mm x 20 mm) | |
Meter amps (A) input: 20 A, 500 V, time delay (F20H500 V – 6.3 mm x 32 mm) | |
Loop/Line impedance input: 4 A, 500 V, time delay (T4H500 V – 6.3 mm x 32 mm) | |
Leakage current input: 100 mA, 250 V, fast (F100 mL250 V – 5 mm x 20 mm) | |
Environment | |
Warm-up time | 15Â minutes |
Temperature performance | Operating temperature: 18 °C to 28 °C |
Calibration temperature (tcal): 23 °C | |
Temperature coefficient for temperatures outside of tcal ± 5 °C is 0.1 / °C of the specification, for temperatures between 5 °C to 40 °C. | |
Storage temperature : -10 °C to 50 °C | |
Storage recovery time: Typically <24 hours at operating temperature | |
Relative humidity (operating) | <80 % to 28 °C (resistance outputs >10 GΩ specified for <70 % to 28 °C) |
Relative humidity (storage) | 90 % non-condensing 0 °C to 50 °C |
Altitude | Operating: 2000 m (6561 ft) |
Storage: 12 200 m (40 000 ft) | |
Dimensions and weight | |
Dimensions | 430 mm x 555 mm x 170 mm (16.9 in x 21.8 in x 6.7 in) |
Weight | 20 kg (44.1 lb) |
Compliance | |
Safety | Mains: IEC 61010-1: Overvoltage category II, pollution degree 2 |
Measurement: IEC 61010-2-030: 5000 V (not category rated) | |
Electromagnetic compatibility (EMC) | |
International:
|
|
Korea (KCC):
|
|
USA (FCC):
|
|
Electrical specifications | |
Low resistance source | |
Range | 100 mΩ to 10 kΩ + 10 mΩ single value selection, dc and line frequency (50/60 Hz) |
1.5 kV high resistance source (DC only) | |
Range | 10 kΩ to 10 GΩ + 100 GΩ single value selection |
5.5 kV high resistance source (DC only) (5322A with /5 option) | |
Range | 10 kΩ to 100 GΩ |
Test voltage measurement | |
Range | 0 V dc to 5500 V dc |
Ground bond resistance source | |
Resistance mode | |
Range | 1 mΩ to 1700 Ω, dc and line frequency (50/60 Hz) |
Line/loop impedance source | |
Range | 25 mΩ to 1700 Ω |
Correction manual/scan mode | |
Residual impedance range | 0 Ω to 10 Ω |
Resolution | 1 mΩ |
Correction COMP mode (active loop compensation) (5322AVLC Option) | |
Maximum compensated impedance | 0 Ω to 2 Ω |
Leakage current source | |
Range | 0.1 to 30 mA |
Test voltage | Passive mode: 60 V ac to 250 V ac rms |
Differential mode: 60 V ac to 250 V ac rms | |
Substitute mode:10 V ac to 250 V ac rms | |
Active mode (5322A/VLC only)[1]: 50 V ac to 100 V ac rms | |
Uncertainty | Passive mode: ± (0.3% setting + 2 μA) |
Differential mode: ± (0.3% setting + 2 μA) | |
Test uncertainty can be influenced by power line voltage instability | |
Substitute mode: ± (0.3% setting + 2 μA) | |
Active mode (5322A/VLC only)[1]: ± (0.3% setting + 1 μA) | |
[1]Â The Active Mode outputs are synchronized with the ac mains frequency to suppress interference between the calibrator and external noise sources. | |
Human body simulation (for substitute leakage currently only | Resistance range: 0 Ω to 10 000 Ω |
Resolution: 1 Ω | |
RCD (residual current device) (for installation testers) | |
Trip current range | 0.5 X I and 1 X I Mode: 5 to 30 mA in 1 mA steps |
1.4 X I and 2 X I Mode: 14 to 60 mA in 1 mA steps | |
5 X I Mode: 50 to 150 mA in 1 mA steps | |
Trip current measurement resolution | 1 μA below 30 mA |
10 μA in range from 30 mA to 150 mA | |
100 μA in range from 300 mA to 3 A | |
Trip current measurement uncertainty | |
Trip current | ± 1 % of nominal current (I) setting |
Trip time range | 10 to 5000 ms |
Trip time uncertainty | (0.02 % setting + 0.25 ms) |
Touch/line voltage | Touch voltage range: 50 V |
Touch voltage setting: in discrete points depending on setup trip current value | |
Touch series resistance: 0.02 Ω, 0.05 Ω, 0.10 Ω, 0.35 Ω, 0.50 Ω, 0.96 Ω, 1.7 Ω, 4.7 Ω, 9 Ω, 17 Ω, 47 Ω, 90 Ω, 170 Ω, 470 Ω, 900 Ω, 1700 Ω | |
Line voltage range: 250 V | |
Line voltage uncertainty: ± (5 % reading + 3 V) | |
User selectable nominal line voltage: 100 V/115 V/120 V/220 V/230 V/240 V/250 V or real | |
Post-trip delayed power restore mode: user selectable | |
RCD (Residual current device (for PATs) | |
Trip current range | 0.5 X I to 5 X I Mode: 5 to 150 mA in 1 mA steps |
Trip current measurement resolution | 1 μA |
Trip current measurement uncertainty | Trip current: ± 1 % of nominal current (I) setting |
Trip time range | 10 to 5000 ms |
Trip time uncertainty | (0.02 % setting + 0.25 ms) |
Line voltage | Reconnection delay: 2.5 s |
AC/DC voltage calibrator (5322A with VLC Option) | |
Range | 0.03 V to 600 V, ac or dc (.5 – 1%) |
Internal ranges | AC mode: 0.3 V to 600 V |
DC mode: 0.3 V to 600 V | |
Multimeter | |
Maximum withstand voltage | HV terminal to COM terminal: 5000 V rms |
V terminal to COM terminal: Â 1100 V rms | |
COM terminal to Protective Earth: 2200 V pk | |
AC/DC voltage range | V (1100 V) Input:  0 V dc to ±1100 V dc 10 mV to 1100 V ac rms |
HV (5000 V) Input: 0 Vdc to ±5000 Vdc 5 V to 5000 V ac rms | |
AC/DC current | |
Range | 0 A to 20 A continuous, 20 A to 30 A for up to 5 minutes, ac rms or dc |
AC power | |
Range | 0 kVA ac to 33 kVA ac |
Type | Apparent, active, reactive |
Phase indication | Phase angle (φ), Power Factor (PF) |
Phase uncertainty (dφ) | ± 0.1 ° |
DC power | |
Range | 0 to 33 kVA dc |
Hipot ac voltage distortion measurement | |
Frequency range | 45 Hz to 65 Hz |
Number of harmonics | 25 |
Hipot dc voltage ripple coefficient measurement | |
Voltage range | 100 V dc to 5000 V dc |
Ripple coefficient range | 10 % |
Flash test voltage measurement (using flash LC or flash V mode) | |
Class I Voltage range | 2000 V ac rms |
Uncertainty | ± (0.3 % of reading + 6 V) |
Class II Voltage range | 3000 V ac rms |
Uncertainty | ± (1 % of reading value + 6 V) |
Flash leakage current measurement (using flash LC mode) | |
Range | 0 mA ac rms or dc to 300 mA ac rms or dc |
Resolution | 4.5 digits |