Additel ADT227 Multifunction Process Calibrator
- Sourcing, simulating and measuring pressure, temperature and electrical signals
- Built-in Full Hart Communicator (ADT227-HART)
- Built-in Barometer
- Intrinsically safe models available (Ex)
- Large Smartphone Like Touchscreen User Experience
- USB Type-C and Bluetooth communications
- IP67 Rated
- High Voltage Measurement Capability (±300V AC)
- True RMS Voltage Meter Capability
- Dual channel pressure module ports
- High Static Differential Pressure Measurement 0.002% FS
- ISO 17025-accredited calibration w/data included
- Description
- Additional information
- Specifications
Description
The Additel ADT227 is a cutting-edge, compact multifunction process calibrator designed for efficiency and precision in demanding industrial environments. This instrument represents a significant advancement in calibration technology,
Key Features:
- Advanced Functionality: Measures and sources a wide range of electrical signals (mA, V, RTD, TC, frequency, resistance), pressure (gauge, differential, absolute), and more.
- Exceptional Accuracy: Delivers high accuracy for both electrical and pressure measurements, ensuring reliable calibration results.
- Compact and Portable: Designed for ease of use in the field with a small and lightweight form factor.
- User-Friendly Interface: Features an intuitive touchscreen interface for easy navigation and operation.
- Built-in HART Communicator: Enables seamless communication and calibration of HART devices.
- Rugged and Durable: Built to withstand harsh industrial environments with its robust construction and IP67 rating.
- Data Logging and Reporting: Stores calibration data and generates comprehensive reports for documentation and traceability.
Applications:
- Calibration of process transmitters (pressure, temperature, flow, level)
- Maintenance and troubleshooting of process control systems
- Commissioning and startup of new installations
- Field service and on-site instrument repairs
The Additel ADT227 is a powerful and versatile tool for professionals who require a reliable and efficient process calibrator. Its advanced features, compact design, and user-friendly interface make it an ideal choice for various industrial applications.
Additional information
Brand | Additel |
---|
Source Accuracy | ||||||
Specifications |
ADT227 | ADT227Ex | ||||
Range | Resolution | Accuracy | Range | Resolution | Accuracy | |
Voltage DC |
-150 to 150 mV | 5 μV | 0.005%RDG + 15 μV |
0 to 10.5 V |
0.2 mV |
0.01%RDG + 0.5 mV |
-1.5 to 1.5 V | 0.05 mV | 0.005%RDG + 0.15 mV | ||||
-15 to 15 V | 0.5 mV | 0.005%RDG + 1.5 mV | ||||
Current DC | 0 to 25 mA | 0.5 μA | 0.01%RDG + 1.2 μA | 0 to 25 mA | 0.5 μA | 0.01%RDG + 1.2 μA |
Resistance |
0 to 400 Ω | 10 mΩ | 0.005%RDG + 20 mΩ | 0 to 400 Ω | 10 mΩ | 0.01%RDG + 20 mΩ |
0 to 4000 Ω | 100 mΩ | 0.01%RDG + 200 mΩ | 0 to 4000 Ω | 100 mΩ | 0.01%RDG + 200 mΩ | |
Frequency (Square wave) |
(0.01 ~ 5) Hz | 0.00001 Hz | 0.002%RDG + 0.00002 Hz | (0.01 ~ 5) Hz | 0.00001 Hz | 0.002%RDG + 0.00002 Hz |
(5 ~ 50) Hz | 0.0001 Hz | 0.002%RDG + 0.0002 Hz | (5 ~ 50) Hz | 0.0001 Hz | 0.002%RDG + 0.0002 Hz | |
(50 ~ 500) Hz | 0.001 Hz | 0.002%RDG + 0.002 Hz | (50 ~ 500) Hz | 0.001 Hz | 0.002%RDG + 0.002 Hz | |
(500 ~ 5000) Hz | 0.01 Hz | 0.002%RDG + 0.02 Hz | (500 ~ 5000) Hz | 0.01 Hz | 0.002%RDG + 0.02 Hz | |
(5000 ~ 50000) Hz | 0.1 Hz | 0.002%RDG + 0.2 Hz | (5000 ~ 50000) Hz | 0.1 Hz | 0.002%RDG + 0.2 Hz |
Source Accuracy |
||||||
Specifications |
ADT227 |
ADT227Ex |
||||
Range |
Resolution |
Accuracy |
Range |
Resolution |
Accuracy |
|
Frequency (Sine wave & Triangular wave) |
(0.1~ 50) Hz |
0.001 Hz |
0.002 Hz |
N/A |
||
(50 ~ 500) Hz |
0.01 Hz |
0.02 Hz |
||||
(500 ~ 5000) Hz |
0.1 Hz |
0.2 Hz |
||||
(5000 ~ 50000) Hz |
1Hz |
2 Hz |
||||
Duty Cycle |
(1%~99%)@≤10000Hz |
0.05% |
0.1%kHz+0.1% |
Fixed 50%@(0.01~50000)Hz |
||
(5%~99%)@≤50000Hz |
0.5% |
|||||
Voltage mV (TC) |
-10 to 75 mV |
1.5 μV |
0.008%RDG + 3.0 μV |
-10 to 75 mV |
1.5 μV |
0.01%RDG + 3.0 μV |
Pulse |
0 to 9999999 |
1 |
N/A |
0 to 9999999 |
1 |
N/A |
Optional rising edge or falling edge, minimum threshold voltage: 2.5V |
||||||
Loop power (max 25mA) |
24 V |
N/A |
±1 V |
22 V |
N/A |
± 10% |
Measurement Accuracy Cont. | ||||||
Specifications |
ADT227 | ADT227Ex | ||||
Range | Resolution | Accuracy | Range | Resolution | Accuracy | |
Voltage DC |
-300 to 300 mV | 1 μV | 0.005% RDG + 15 μV | -300 to 300 mV | 1 μV | 0.01% RDG + 15 μV |
-30 to 30 V | 0.1 mV | 0.005% RDG + 1.5 mV | -30 to 30 V | 0.1 mV | 0.01% RDG + 1.5 mV | |
Impedance: -300 mV to 300 mV = > 100 MΩ
-30 V to 30 V = >1 MΩ |
||||||
DC High Voltage |
-300 to 300 V | 10 mV | 0.05% RDG + 30 mV |
N/A |
||
Maximum input voltage = 300 V, IEC61010 300V CATII | ||||||
Common mode rejection: >100 dB (at 50 or 60 Hz) | ||||||
Impedance: > 4 MΩ, DC coupling | ||||||
AC High Voltage |
300V (40 to 500 Hz) | 10 mV | 0.5% RDG + 150 mV |
N/A |
||
Maximum input voltage = 300 V, IEC61010 300V CATII | ||||||
9% to 100% of range is suitable for the above accuracy indicators | ||||||
Impedance: >4 MΩ, <100pF, AC coupling | ||||||
Current DC | -30 to 30 mA | 0.1 μA | 0.01% RDG + 1.5 μA | -30 to 30 mA | 0.1 μA | 0.01% RDG + 1.2 μA |
Resistance (4-Wire) |
0 to 400 Ω | 1 mΩ | 0.005% RDG + 20 mΩ | 0 to 400 Ω | 1 mΩ | 0.01% RDG + 20 mΩ |
0 to 4000 Ω | 10 mΩ | 0.01% RDG + 200 mΩ | 0 to 4000 Ω | 10 mΩ | 0.01% RDG + 200 mΩ | |
2-Wire + 50 mΩ, 3-wire+ 10 mΩ | ||||||
Excitation current: 0.2 mA | ||||||
Voltage mV (TC) |
-10 to 75 mV | 0.1uV | 0.008% RDG + 3.0 μV | -10 to 75 mV | 0.1uV | 0.01% RDG + 3.0 μV |
Impedance: >100 MΩ | ||||||
Frequency |
(0.01 ~ 5) Hz | 0.00001 Hz | 0.002%RDG + 0.00002 Hz | (0.01 ~ 5) Hz | 0.00001 Hz | 0.002%RDG + 0.00002 Hz |
(5 ~ 50) Hz | 0.0001 Hz | 0.002%RDG + 0.0002 Hz | (5 ~ 50) Hz | 0.0001 Hz | 0.002%RDG + 0.0002 Hz | |
(50 ~ 500) Hz | 0.001 Hz | 0.002%RDG + 0.002 Hz | (50 ~ 500) Hz | 0.001 Hz | 0.002%RDG + 0.002 Hz | |
(500 ~ 5000) Hz | 0.01 Hz | 0.002%RDG + 0.02 Hz | (500 ~ 5000) Hz | 0.01 Hz | 0.002%RDG + 0.02 Hz | |
(5000 ~ 50000) Hz | 0.1 Hz | 0.002%RDG + 0.2 Hz | (5000 ~ 50000) Hz | 0.1 Hz | 0.002%RDG + 0.2 Hz | |
Minimum threshold voltage: 2.5 V | ||||||
Supported units: Hz, kHz, MHz, CPM, CPH, s, ms, μs | ||||||
Duty Cycle |
(1%~99%)@≤10000Hz | 0.01% |
0.1% kHz + 0.05% |
N/A |
||
(5%~99%)@≤50000Hz | 0.1% | |||||
Pulse |
0 to 9999999 | 1 | N/A | 0 to 9999999 | 1 | N/A |
Optional rising edge or falling edge, minimum threshold voltage: 2.5V | ||||||
Switch | Supports dry or wet switches. Voltage range of 3 to 30 V. Response speed < 10 ms |
General Specifications
Specifications | ADT227 | ADT227Ex |
Operating Temperature | -10°C to 50°C | -20°C to 50°C |
Specification guaranteed
temperature range |
10°C to 30°C | 10°C to 30°C |
Storage Temperature | -30°C to 70°C | -30°C to 70°C |
Humidity | <95%, non-condensing | <95%, non-condensing |
Power supply |
6600mAh, 23.8Wh lithium battery, charging time about 6 hours, battery pack can be charged independently |
4000mAh 14.4Wh Explosion-proof lithium battery packcharging time
6~8 hours, battery pack can be charged independently |
User interface | Icon drive menus | Icon driven menus with navigation buttons |
Ports protection voltage |
50V max
(Only for the top ports) |
30V max |
Display | 5.0 inch 480 x 800 mm TFT LCD capacitive screen | 4.4 inch 640 x 480 mm color display capacitive screen |
Data logger |
500 results, each result x 100,0000 recordings, |
N/A |
Maximum altitude | 3000 meters | |
European Compliance | CE Mark | |
Electrical Connection | Ø4mm sockets and flat mini-jack thermocouple socket | |
Size | 6.97″ x 4.13″ x 2.04″ (177 mm x 105 mm x 52 mm) | |
Weight | 1.6 lb (0.7 kg) | 1.65 lb (0.75Kg) |
Battery | Rechargeable Li-ion battery (included) | |
Battery Life | Typically 12 hours | Typically 35 hours |
Battery Charge | 110V/220V external power adapter included. Battery can be charged external to the unit. Typically charge time is 6-8 hours. | |
External pressure module | Dual channel Serial plug, can connect two digital pressure modules | |
Warm-up time | Full specification performance is achieved after a 10 minute warm-up time. | |
ROHS compliant | Rohs II Directive 2011/65/EU, EN50581:2012 | |
Display rate | 3 readings per second | |
Barometric Accuracy
(Built-in barometer) |
55Pa | |
IP protection level | IP67, 1 meter drop test | |
Communication | Isolate USB-TYPEC (slave), Bluetooth BLE | |
User Interface
Localization |
English, German, French, Italian, Spanish, Portuguese,
Simplified Chinese, Traditional Chinese, Japanese, Russian, Czech, Slovak |
English, Simplified Chinese, Traditional Chinese, Japanese |
Calibration | ISO 17025 accredited calibration with data | |
Warranty | 3 years |