Additel 209/210 Series Loop Calibrator

  • Accuracy to 0.01% of reading
  • Measure, source, or simulate loop current
  • Measure DC volts
  • Simultaneously mA and % span display
  • Switch functionality
  • Selectable ramp and step functions
  • HART 250Ω resistor in series with 24V loop
  • ISO 17025-accredited calibration w/data included

Description

The Additel Additel 209/210 are both versatile multifunction process calibrators designed for testing and calibrating a wide range of industrial instruments. While both models share a common foundation, the ADT210 offers a significant upgrade in performance and capabilities compared to the ADT209.

Additel ADT209:

  • Entry-Level Workhorse: The ADT209 is positioned as an entry-level model, offering a robust and reliable solution for a wide range of basic calibration tasks.
  • Key Features:
    • Essential Measurements: Measures and sources a variety of electrical signals, including mA, V, RTD, and TC.
    • Loop Power Supply: Provides loop power for powering process transmitters during calibration.
    • User-Friendly Interface: Features an intuitive interface for easy operation and navigation.
    • Rugged and Durable: Designed to withstand the demands of industrial environments with a durable construction.
    • Applications: Ideal for technicians and engineers who require a reliable and versatile tool for routine calibration tasks, such as calibrating basic transmitters, troubleshooting process
    • loops, and performing general instrument checks.

Additel ADT210:

  • Advanced Performance: The ADT210 represents a significant step up in performance and capabilities. It offers enhanced accuracy, advanced functionality, and a more robust feature set.
  • Key Features:
    • Superior Accuracy: Delivers higher accuracy for both electrical and pressure measurements, ensuring precise and reliable calibration results.
    • Expanded Functionality: Includes advanced features such as HART communication, pressure measurement and sourcing, and more comprehensive data logging capabilities.
    • Enhanced User Experience: Offers a more sophisticated user interface with advanced features like data trending, graphical display, and customizable settings.
    • Rugged and Reliable: Built to withstand harsh industrial environments with an IP67 rating for protection against dust and water ingress.

Applications Additel 209/210: Ideal for professionals who require a high-performance calibrator for demanding applications, such as calibrating complex transmitters, performing advanced troubleshooting, and ensuring the highest level of accuracy in critical processes.

Additional information

Brand

Additel

Electrical Specifications

Range Resolution ADT209 Accuracy ADT210 Accuracy
Voltage DC1 0 to 30 V 1 mV 0.03%RD + 2mV 0.01%RD + 2mV
Current DC2 0 to 24 mA 1 μA 0.03%RD + 2μA 0.01%RD + 2μA
 

Switch test

Input resistance more than  500 MΩ

Trigger level: low level <0.3V; high level: >2V

Electrical Source Specifications

Range Resolution ADT209 Accuracy ADT210 Accuracy
Current DC 0 to 24 mA 1 μA 0.03%RD + 2 μA 0.01%RD + 2 μA
Source mode 700Ω/20 mA maximum
Sink mode External loop voltage nominal 24 V, maximum 30 V, minimum 12 V

General Specifications

Voltage limit 30 V between terminals or between terminals and ground
Measurement Functions Auto step, auto ramp, span step
Display VA LCD display. 2.04 x 2.04 in (52 x 52 mm)
Loop power 24 V
Over-voltage protection 30 V DC (240 V AC)
Overload current protection 33 mA DC
Storage temperature -20°C to 70°C
Working Environment -10 to 50°C, 95%RH
Working Altitude <3,000 m
 

Vibration/shock

Random 2G 5 to 500 Hz

1 meter drop

 

Power

One 9 V alkaline battery (ANSI/NEDA 1604A or IEC) DC9 V optional adapter available
 

Battery life (typical)

Output  mode:  18 hours  (12 mA/500Ω)

Measure mode: 50 hours

Size (LxWxH) 163 x 83 x 41 mm
Weight 350 g
Calibration Certification ISO 17025 Accredited Calibration with data
 

Compliance Certification

ADT209: CE ADT210: CE
Warranty 3 years