Description
The Fluke Calibration P3800 Series are high-accuracy, hydraulic deadweight testers designed for calibrating a wide range of high-pressure instruments.
Key Features:
- Pressure ranges to 60,000 psi (4000 bar)
- Select from two accuracy classes; 0.02% or 0.015% of reading
- Includes hand pump and intensifier for generating high pressures
- Tungsten carbide piston and cylinder provides long term stability and durability
- Mass set can be trimmed to local gravity FOC
- High Accuracy: Offers exceptional accuracy, typically better than 0.02% of reading, ensuring precise and reliable pressure measurements at high pressures.
- High Pressure Range: Capable of generating high pressures, typically up to 60,000 psi (4000 bar), making it suitable for calibrating instruments used in demanding applications.
- Rugged and Durable: Designed for demanding laboratory and field environments, with robust construction and reliable performance.
- Traceability: Calibrations are traceable to national and international standards, ensuring the accuracy and reliability of your measurements.
- Easy to Use: User-friendly design and intuitive operation simplify the calibration process.
Applications:
- Calibration Laboratories: Primary standard for calibrating high-pressure instruments, including transducers, transmitters, gauges, and switches used in various industries.
- Industrial Environments: Used for on-site calibration of critical process instrumentation in industries such as oil and gas, aerospace, and power generation.
- Research and Development: Employed in research and development laboratories for precise pressure measurements and calibrations in high-pressure applications.
In essence:
The Fluke Calibration P3800 Series are high-performance pressure standards specifically designed for calibrating instruments operating at high pressures, offering exceptional accuracy and reliability for critical applications.
Specifications: Fluke Calibration P3800 Series Hydraulic Deadweight Tester
| Pressure Ranges |
| P3830 |
500 to 30,000 psi, or 40 to 2,000 bar, or 4 to
200 MPa, or 40 to 2000 kgf/cm2 |
| P3840 |
500 to 40,000 psi, or 40 to 2,600 bar, or 4 to
260 MPa, or 40 to 2600 kgf/cm2 |
| P3860 |
500 to 60,000 psi or 40 to 4,000 bar or 4 to
400 MPa or 40 to 4000 kgf/cm2 |
| Accuracy |
| Accuracy1 |
Standard accuracy is 0.02% of Reading1. Optional accuracy of 0.015% of Reading is vailable. Calibration certificate with traceability to NIST is provided with each instrument. An optional ISO/IEC 17025 accredited calibration certificate is available. |
| 1. Accuracy is based on % of Reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10%, ± (accuracy class) x 10% of the piston range. |
| Materials of Construction |
| Standard weight material |
Series 3 non-magnetic, austenitic, stainless steel |
| Weight density |
7.8 g/cm3 |
| Piston material |
Tungsten carbide with nickel binder |
| Cylinder material |
Tungsten carbide with cobalt binder |
| Thermal coefficient of expansion |
11 ppm/°C |
| General |
| Test port adapters |
9/16 – 18 UNF (Autoclave); 3/8 BSP; 1/2 BSP and 3/4 BSP |
| Instrument weight |
66 lbs. (30 Kg) Instrument Base only |
| Instrument size |
18 x 13.5 x 19 in (455 x 340 x 478 mm) |
| Reservoir volume |
235 cc (14.3 in3) |
| Intensifier ratio |
7 to 1 |
| Seal materials |
Nitrile |
| Operating fluid |
Dioctyl Sebacate (DOS), our reference 55-600 |
| Options |
| Presscal software |
Windows based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Supports additional 12 pressure units (bar, mbar, MPa, kPa, psi, kg/cm2, atm, inH2O, mH2O, mmH2O, inHg, mmHg). |
| Conversion weight sets |
For applications that require nominal pressure increments in measurement units different than the main unit, conversion weight sets are available. Available measurement units are: psi, bar kgf/cm2 or MPa. |