Fluke P3200 Hydraulic Deadweight Tester
- Pressure ranges to 10,000 psi (700 bar)
- 0.015% of reading accuracy standard (0.008% optional)
- Single or dual piston formats
- Built-in hand pumps standard
- Mounted spirit level with adjustable feet
- Units can be trimmed to local gravity FOC
- Description
- Additional information
- Specifications
Description
The Fluke Calibration P3200 Series are high-accuracy, hydraulic deadweight testers designed for calibrating a wide range of pressure instruments.
Key Features:
- Pressure ranges to 10,000 psi (700 bar)
- 0.015% of reading accuracy standard (0.008% optional)
- Single or dual piston formats
- Built-in hand pumps standard
- Mounted spirit level with adjustable feet
- Units can be trimmed to local gravity FOCHigh Accuracy: Offers exceptional accuracy, typically better than 0.015% of reading, ensuring precise and reliable pressure measurements.
- Versatility: Available in various configurations to cover a wide range of pressure ranges, from low to high pressures.
- Rugged and Durable: Designed for demanding laboratory and field environments, with robust construction and reliable performance.
- Traceability: Calibrations are traceable to national and international standards, ensuring the accuracy and reliability of your measurements.
- Easy to Use: User-friendly design and intuitive operation simplify the calibration process.
Applications:
- Calibration Laboratories: Primary standard for calibrating pressure instruments, including gauges, transmitters, transducers, and switches.
- Industrial Environments: Used for on-site calibration of critical process instrumentation in various industries, such as oil and gas, chemical, and manufacturing.
- Research and Development: Employed in research and development laboratories for precise pressure measurements and calibrations.
The Fluke Calibration P3200 Series are high-performance pressure standards that provide exceptional accuracy and reliability for critical pressure calibration applications.
Additional information
Brand | Fluke Calibration |
---|
Specifications: Fluke Calibration P3200 Hydraulic Deadweight Tester
Pressure Ranges | |
Distilled water operated | Ranges to 10,000 psi (700 bar) |
Accuracy | |
Accuracy1 | ±0.015% of reading (±0.008% optional) |
1. Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range. | |
Materials of Construction | |
Standard weight material | Series 3 non-magnetic austenitic stainless steel |
Weight density | 7.8 g/cm³ |
Optional fractional weights | Solution heat treated aluminum |
Weight density | 2.7 g/cm³ |
Piston material | Tungsten carbide with nickel binder Density- 15.0 g/cm³ |
Cylinder material | Hardened martensitic steel Tungsten carbide (water above 500 psi, 35 bar) |
Thermal coefficients of expansion | Oil piston/cylinder 16.5 ppm/°C Water (above 500 psi, 35 bar) 11 ppm/°C |
General | |
Test port adaptors | 1/8, 1/4, 3/8 and 1/2 NPT or BSP |
Instrument weight | 36 lbs (16 kg) |
Instrument size (W x D x H) | 17.5 x 12 x 8.5 in (440 x 300 x 215 mm) |
Mass set weight (typical) | 80 lbs (36 kg) |
Reservoir volume | 150 cc |
Screw press displacement | 5.5 cc |
Pump displacement | 4.7 cc per stroke |
O-ring seal materials | Buna N as standard, Viton and EPDM available |
Weight Increments | |
Minimum standard weight increments | Up to 500 psi (35 bar): 1 psi (0.1 bar) 500 to 5000 psi (35 to 350 bar): 10 psi (1 bar) 5000 to 20,000 psi (350 to 1400 bar): 20 psi (2 bar) Model P3112: 1 psi (0.1 bar) |
Optional fractional weights | Up to 500 psi (35 bar): 0.1 psi (0.01 bar) 500 to 5000 psi (35 to 350 bar): 1 psi (0.1 bar) 5000 to 20,000 psi (350 to 1400 bar): 2 psi (0.2 bar) |
Operating Fluids | |
Operating fluids | Distilled or deionized water |
Options | |
PressCal software | Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments |
Fluids | Instruments can be supplied for use with Skydrol or brake fluids. Due to the aggressive nature of these fluids the standard nitrile seals and the acrylic reservoir tube will be replaced by either Viton or EPDM seals (as applicable) and an aluminum reservoir tube. For these or any other specialty fluid applications, please consult the factory. |