Fluke P3000 Pneumatic Deadweight Testers
- 0.015% of reading accuracy standard (0.008% optional)
- 1-30 inHg vacuum (0.03-1 bar vacuum)
- 3-500 psi (0.2-35 bar) pressure
- inH2O and mbar ranges available
- Series 3 non-magnetic stainless steel weights
- Units can be trimmed to local gravity FOC
- Description
- Additional information
- Specifications
Description
The Fluke Calibration P3000 Series are high-accuracy, piston-cylinder-based pressure standards used for calibrating a wide range of pressure instruments.
Key Features:
- 0.015% of reading accuracy standard (0.008% optional)
- 1-30 inHg vacuum (0.03-1 bar vacuum)
- 3-500 psi (0.2-35 bar) pressure
- inH2O and mbar ranges available
- Series 3 non-magnetic stainless steel weights
- Units can be trimmed to local gravity FOC
- High Accuracy: Offers exceptional accuracy, typically better than 0.015% of reading, ensuring precise and reliable pressure measurements.
- Versatility: Available in various configurations to cover a wide range of pressure ranges, from low to high pressures.
- Rugged and Durable: Designed for demanding laboratory and field environments, with robust construction and reliable performance.
- Traceability: Calibrations are traceable to national and international standards, ensuring the accuracy and reliability of your measurements.
- Easy to Use: User-friendly design and intuitive operation simplify the calibration process.
Applications:
- Calibration Laboratories: Primary standard for calibrating pressure instruments, including gauges, transmitters, transducers, and switches.
- Industrial Environments: Used for on-site calibration of critical process instrumentation in various industries, such as oil and gas, chemical, and manufacturing.
- Research and Development: Employed in research and development laboratories for precise pressure measurements and calibrations.
In essence:
The Fluke Calibration P3000 Series represents the pinnacle of pressure calibration technology, offering exceptional accuracy, reliability, and versatility for a wide range of applications. These instruments are essential for maintaining the integrity and accuracy of pressure measurements in critical applications.
Additional information
Brand | Fluke Calibration |
---|
Specifications: Fluke Calibration P3000 Pneumatic Deadweight Testers
Accuracy | ||
Accuracy1 | ±0.015% of reading (±0.008% optional) | |
1. Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range. | ||
Mass and PCU Materials of Construction | ||
Standard weight material | Series 3 non-magnetic austenitic stainless steel | |
Weight density | 7.8 g/cm³ | |
Optional fractional weights | Solution heat treated aluminum | |
Weight density | 2.7 g/cm³ | |
Models P3011, P3012, P3013, P3022, P3023, P3025 (V) | Piston material | Ceramic |
Cylinder material | Martensitic steel | |
Coefficient of expansion | 11 ppm/°C | |
Models P3014, P3015, P3025 (P) P3031, P3032 (P & V) | Piston material | Tungsten carbide |
Cylinder material | Martensitic steel | |
Coefficient of expansion | 16.5 ppm/°C | |
Model P3016 | Piston material | Tungsten carbide |
Cylinder Material | Tungsten carbide | |
Coefficient of Expansion | 11 ppm/°C | |
General | ||
Test port adaptors | 1/8, 1/4, 3/8 and ½ NPT or BSP | |
Instrument weight | 24 lbs | |
Instrument size (W x D x H) | 17.5 x 12 x 8.5 | |
Mass set weight (typical) | 65 lbs | |
Minimum Standard Weight Increments | ||
P3012, P3022 Pressure | 1 inH2O, (5 mbar) | |
P3013, P3023 Pressure | 2 inH2O, (10 mbar) | |
P3014, P3015, P3025 Pressure | 1 psi, (0.1 bar) | |
P3016, P3031, P3032 | 1 psi, (0.1 bar) | |
P3011, P3022, P3023, P3025 Vacuum | 0.2 inHg (10 mbar) | |
Optional Fractional Weights | ||
P3014, P3015, P3025 Pressure | 0.1 psi (0.01 bar) | |
Built-in Hand Pumps | ||
Pressure mode | Max pressure 300 psi (20 bar) | |
Vacuum mode | To 90% vacuum | |
Operating Fluid for Liquid Lubricated Models | ||
55-655 | Mineral oil (standard) | |
Krytox | For oxygen-safe applications (optional) | |
Options | ||
PressCal Software | Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments. | |
Krytox | As operating fluid for liquid lubricated gas operated models. |