Fluke 9190A Ultra-Cool Field Metrology Well

  • Wide temperature range: –95 °C to 140 °C.
  • Excellent accuracy
  • Accuracy using built-in reference thermometer readout: ± 0.05 °C full range.
  • Display accuracy: ± 0.2 °C full range.
  • Best-in-class stability: ± 0.015 °C full range

Description

The Fluke 9190A Ultra-Cool Field Metrology Well is a portable, dry-block temperature calibrator designed for on-site calibration of temperature sensors, thermocouples, and thermometers. It’s particularly well-suited for applications requiring accurate temperature control at low temperatures.

Key features and specifications:

  • Wide temperature range: –95 °C to 140 °C.
  • Excellent accuracy
  • Accuracy using built-in reference thermometer readout: ± 0.05 °C full range.
  • Display accuracy: ± 0.2 °C full range.
  • Best-in-class stability: ± 0.015 °C full range
  • Wide temperature range: Covers a broad range from -95°C to 140°C, ideal for various applications in industries like pharmaceuticals, food processing, and medical.
  • High accuracy and stability: Offers high accuracy and stability, ensuring reliable calibration results.
  • Portable and lightweight: Compact and easy to transport, making it suitable for on-site calibrations.
  • Fast heating and cooling: Reaches set points quickly, saving time during calibration processes.
  • Built-in reference thermometry: Includes a reference thermometer for accurate temperature measurement.
  • User-friendly interface: Easy to operate with intuitive controls.

Applications:

  • Pharmaceutical Industry:
    • Calibration of temperature sensors used in pharmaceutical manufacturing, storage, and transportation
    • Validation of medical freezers, refrigerators, and cold rooms to ensure product stability and safety
    • Calibration of thermometers used in quality control and research and development
  • Food Processing:
    • Calibration of temperature sensors used in food processing and storage
    • Validation of refrigeration equipment to maintain food safety and quality
    • Calibration of thermometers used in food safety inspections
  • Medical Industry:
    • Calibration of medical thermometers, blood bank equipment, and other temperature-sensitive medical devices
    • Validation of laboratory equipment and cold storage units
    • Research and development in medical fields

Overall, the Fluke 9190A is a versatile and reliable temperature calibrator that provides accurate and efficient temperature control for a wide range of applications.

Additional information

Brand

Fluke Calibration

Specifications: Fluke 9190A Ultra-Cool Field Metrology Well

Specifications
Temperature range at 23°C –95°C to 140°C (–139°F to 284°F)
Display accuracy ±0.2°C full range
Accuracy with external reference3 ±0.05°C full range
Stability ±0.015°C full range
Axial uniformity at 40 mm (1.6 in) ±0.05°C full range
Radial gradient ±0.01°C full range
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
±0.006°C full range
(versus display with one 6.35 mm probe)
±0.25°C at –95°C
±0.10°C at 140°C
Operating conditions 0°C to 35°C, 0% to 90%
RH (non-condensing) < 2000 m altitude
Environmental conditions for all specifications except temperature range 13°C to 33°C
Immersion (well) depth 160 mm (6.3 in)
Well diameter 30 mm (1.18 in)
Heating time1 –95°C to 140°C: 40 min
Cooling time1 23°C to –90°C: 80 min
23°C to –95°C: 90 min
140°C to 23°C: 60 min
Stabilization time2 15 min
Resolution 0.01°
Display LCD, °C or °F user selectable
Size (H x W x D) 480 x 205 x 380 mm (18.8 x 8.0 x 14.9 in)
Weight 16 kg (35 lb)
Power requirements 100 V to 115 V (±10%) 50/60 Hz, 575 W
200 V to 230 V (±10%) 50/60 Hz, 575 W
System fuse ratings 115 V: 6.3 A T 250 V
230 V: 3.15 A T 250 V
4–20 mA fuse (-P model only) 50 mA F 250 V
Computer interface RS-232, and USB Serial included
Safety IEC 61010-1, Installation Category II, Pollution degree 2
Electromagnetic environment IEC 61326-1: Basic
Refrigerants R32 (Difluoromethane)
< 20 g, ASHRAE safety group A2L
R704 (Helium)
< 20 g, ASHRAE safety group A1
-P Specifications
Built-in reference thermometer readout accuracy (4-wire reference probe)3 ±0.010°C at –95°C
±0.013°C at –25°C
±0.015°C at 0°C
±0.020°C at 50°C
±0.025°C at 140°C
Reference resistance range 0 Ω to 400 Ω
Reference resistance accuracy4 0 Ω to 42 Ω: ±0.0025 Ω
42 Ω to 400 Ω: ±60 ppm of reading
Reference characterizations ITS-90, CVD, IEC-751, resistance
Reference measurement capability 4 wire
Reference probe connection 6-pin din with INFO-CON technology
Built-in RTD thermometer readout accuracy NI-120: ±0.015°C at 0°C
PT-100 (385): ±0.02°C at 0°C
PT-100 (3926): ±0.02°C at 0°C
PT-100 (JIS): ±0.02°C at 0°C
RTD resistance range 0 Ω to 400 Ω
Resistance accuracy4 0 Ω to 25 Ω: ±0.002 Ω
25 Ω to 400 Ω: ±80 ppm of reading
RTD characterizations PT-100 (385), (JIS), (3926), NI-120, resistance
RTD measurement capability 2-wire, 3-wire, and 4-wire RTD with jumpers only
RTD connection 4-terminal input
Built-in TC thermometer readout accuracy5 Type J: ±0.70°C at 140°C
Type K: ±0.75°C at 140°C
Type T: ±0.60°C at 140°C
Type E: ±0.60°C at 140°C
Type R: ±1.60°C at 140°C
Type S: ±1.60°C at 140°C
Type M: ±0.65°C at 140°C
Type L: ±0.65°C at 140°C
Type U: ±0.70°C at 140°C
Type N: ±0.75°C at 140°C
Type C: ±1.00°C at 140°C
TC millivolt range –10 mV to 75 mV
Voltage accuracy 0.025% of reading +0.01 mV
Internal cold junction compensation accuracy ±0.35°C (ambient of 13°C to 33°C)
TC connection Miniature connectors (ASTM E1684)
Built-in mA readout accuracy 0.02% of reading + 0.002 mA
mA range Cal 4-22 mA, Spec 4-24 mA
mA connection 2 terminal input
Loop power function 24 V DC loop power
Built-in electronics temperature coefficient
(0°C to 13°C, 33°C to 50°C)
±0.005% of range per °C
1. For ambient temperature of 23°C.
2. Time from when the SETPOINT is reached to when the unit is with in stability specification.
3. The temperature range may be limited by the reference probe connected to the readout.
The built-in reference accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors.
4. Measurement accuracy specifications apply within the operating range and assume 4 wires for PRTs. With 3-wire RTDs add 0.05 Ω to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
5. The thermocouple input readout is sensitive to EM fields in the frequency range of 500 MHz to 700 MHz.

 

Related Products