Fluke 9190A Ultra-Cool Field Metrology Well
- Wide temperature range: –95 °C to 140 °C.
- Excellent accuracy
- Accuracy using built-in reference thermometer readout: ± 0.05 °C full range.
- Display accuracy: ± 0.2 °C full range.
- Best-in-class stability: ± 0.015 °C full range
- Description
- Additional information
- Specifications
Description
The Fluke 9190A Ultra-Cool Field Metrology Well is a portable, dry-block temperature calibrator designed for on-site calibration of temperature sensors, thermocouples, and thermometers. It’s particularly well-suited for applications requiring accurate temperature control at low temperatures.
Key features and specifications:
- Wide temperature range: –95 °C to 140 °C.
- Excellent accuracy
- Accuracy using built-in reference thermometer readout: ± 0.05 °C full range.
- Display accuracy: ± 0.2 °C full range.
- Best-in-class stability: ± 0.015 °C full range
- Wide temperature range: Covers a broad range from -95°C to 140°C, ideal for various applications in industries like pharmaceuticals, food processing, and medical.
- High accuracy and stability: Offers high accuracy and stability, ensuring reliable calibration results.
- Portable and lightweight: Compact and easy to transport, making it suitable for on-site calibrations.
- Fast heating and cooling: Reaches set points quickly, saving time during calibration processes.
- Built-in reference thermometry: Includes a reference thermometer for accurate temperature measurement.
- User-friendly interface: Easy to operate with intuitive controls.
Applications:
- Pharmaceutical Industry:
- Calibration of temperature sensors used in pharmaceutical manufacturing, storage, and transportation
- Validation of medical freezers, refrigerators, and cold rooms to ensure product stability and safety
- Calibration of thermometers used in quality control and research and development
- Food Processing:
- Calibration of temperature sensors used in food processing and storage
- Validation of refrigeration equipment to maintain food safety and quality
- Calibration of thermometers used in food safety inspections
- Medical Industry:
- Calibration of medical thermometers, blood bank equipment, and other temperature-sensitive medical devices
- Validation of laboratory equipment and cold storage units
- Research and development in medical fields
Overall, the Fluke 9190A is a versatile and reliable temperature calibrator that provides accurate and efficient temperature control for a wide range of applications.
Additional information
Brand | Fluke Calibration |
---|
Specifications: Fluke 9190A Ultra-Cool Field Metrology Well
Specifications | |
Temperature range at 23°C | –95°C to 140°C (–139°F to 284°F) |
Display accuracy | ±0.2°C full range |
Accuracy with external reference3 | ±0.05°C full range |
Stability | ±0.015°C full range |
Axial uniformity at 40 mm (1.6 in) | ±0.05°C full range |
Radial gradient | ±0.01°C full range |
Loading effect | (with a 6.35 mm reference probe and three 6.35 mm probes) |
±0.006°C full range | |
(versus display with one 6.35 mm probe) | |
±0.25°C at –95°C | |
±0.10°C at 140°C | |
Operating conditions | 0°C to 35°C, 0% to 90% |
RH (non-condensing) < 2000 m altitude | |
Environmental conditions for all specifications except temperature range | 13°C to 33°C |
Immersion (well) depth | 160 mm (6.3 in) |
Well diameter | 30 mm (1.18 in) |
Heating time1 | –95°C to 140°C: 40 min |
Cooling time1 | 23°C to –90°C: 80 min |
23°C to –95°C: 90 min | |
140°C to 23°C: 60 min | |
Stabilization time2 | 15 min |
Resolution | 0.01° |
Display | LCD, °C or °F user selectable |
Size (H x W x D) | 480 x 205 x 380 mm (18.8 x 8.0 x 14.9 in) |
Weight | 16 kg (35 lb) |
Power requirements | 100 V to 115 V (±10%) 50/60 Hz, 575 W |
200 V to 230 V (±10%) 50/60 Hz, 575 W | |
System fuse ratings | 115 V: 6.3 A T 250 V |
230 V: 3.15 A T 250 V | |
4–20 mA fuse (-P model only) | 50 mA F 250 V |
Computer interface | RS-232, and USB Serial included |
Safety | IEC 61010-1, Installation Category II, Pollution degree 2 |
Electromagnetic environment | IEC 61326-1: Basic |
Refrigerants | R32 (Difluoromethane) |
< 20 g, ASHRAE safety group A2L | |
R704 (Helium) | |
< 20 g, ASHRAE safety group A1 | |
-P Specifications | |
Built-in reference thermometer readout accuracy (4-wire reference probe)3 | ±0.010°C at –95°C |
±0.013°C at –25°C | |
±0.015°C at 0°C | |
±0.020°C at 50°C | |
±0.025°C at 140°C | |
Reference resistance range | 0 Ω to 400 Ω |
Reference resistance accuracy4 | 0 Ω to 42 Ω: ±0.0025 Ω 42 Ω to 400 Ω: ±60 ppm of reading |
Reference characterizations | ITS-90, CVD, IEC-751, resistance |
Reference measurement capability | 4 wire |
Reference probe connection | 6-pin din with INFO-CON technology |
Built-in RTD thermometer readout accuracy | NI-120: ±0.015°C at 0°C |
PT-100 (385): ±0.02°C at 0°C | |
PT-100 (3926): ±0.02°C at 0°C | |
PT-100 (JIS): ±0.02°C at 0°C | |
RTD resistance range | 0 Ω to 400 Ω |
Resistance accuracy4 | 0 Ω to 25 Ω: ±0.002 Ω |
25 Ω to 400 Ω: ±80 ppm of reading | |
RTD characterizations | PT-100 (385), (JIS), (3926), NI-120, resistance |
RTD measurement capability | 2-wire, 3-wire, and 4-wire RTD with jumpers only |
RTD connection | 4-terminal input |
Built-in TC thermometer readout accuracy5 | Type J: ±0.70°C at 140°C |
Type K: ±0.75°C at 140°C | |
Type T: ±0.60°C at 140°C | |
Type E: ±0.60°C at 140°C | |
Type R: ±1.60°C at 140°C | |
Type S: ±1.60°C at 140°C | |
Type M: ±0.65°C at 140°C | |
Type L: ±0.65°C at 140°C | |
Type U: ±0.70°C at 140°C | |
Type N: ±0.75°C at 140°C | |
Type C: ±1.00°C at 140°C | |
TC millivolt range | –10 mV to 75 mV |
Voltage accuracy | 0.025% of reading +0.01 mV |
Internal cold junction compensation accuracy | ±0.35°C (ambient of 13°C to 33°C) |
TC connection | Miniature connectors (ASTM E1684) |
Built-in mA readout accuracy | 0.02% of reading + 0.002 mA |
mA range | Cal 4-22 mA, Spec 4-24 mA |
mA connection | 2 terminal input |
Loop power function | 24 V DC loop power |
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C) |
±0.005% of range per °C |
1. For ambient temperature of 23°C. 2. Time from when the SETPOINT is reached to when the unit is with in stability specification. 3. The temperature range may be limited by the reference probe connected to the readout. The built-in reference accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 4. Measurement accuracy specifications apply within the operating range and assume 4 wires for PRTs. With 3-wire RTDs add 0.05 Ω to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires. 5. The thermocouple input readout is sensitive to EM fields in the frequency range of 500 MHz to 700 MHz. |