Fluke 9150 Thermocouple Furnace

  • Stability of ±0.5 °C, it has a temperature range to 1200 °C and a display accuracy of ±5 °C across its entire range.
  • With interchangeable temperature blocks, you can check thermocouples as small as 1/16 of an inch in diameter or as large as 1/2 inch.
  • Low-cost thermocouple furnace.
  • NIST-traceable calibration included.
  • RS-232 port standard

Description

The Fluke 9150 Thermocouple Furnace is a high-temperature calibration instrument designed for on-site calibration of thermocouples.

Key Features:

  • Stability of ±0.5 °C, it has a temperature range to 1200 °C and a display accuracy of ±5 °C across its entire range.
  • With interchangeable temperature blocks, you can check thermocouples as small as 1/16 of an inch in diameter or as large as 1/2 inch.
  • Low-cost thermocouple furnace.
  • NIST-traceable calibration included.
  • High Temperature Range: Operates within a temperature range of 150°C to 1200°C, allowing for calibration of thermocouples used in high-temperature applications.
  • Stability: Offers good temperature stability, ensuring accurate and reliable calibration results.
  • Interchangeable Blocks: Accommodates various thermocouple sizes and types with interchangeable temperature blocks.
  • Portability: Relatively compact and portable for on-site calibration needs.
  • Ease of Use: Features a user-friendly interface for simple operation and temperature setting.
  • RS-232 Interface: Allows for data logging and integration with other systems.

Applications:

  • Industrial Processes: Calibration of thermocouples used in high-temperature industrial processes, such as metallurgy, ceramics, and power generation.
  • Research and Development: Supports temperature control and calibration in high-temperature research and development.
  • Quality Control: Ensures product quality and consistency through accurate thermocouple calibration.

The Fluke 9150 Thermocouple Furnace is a valuable tool for calibrating thermocouples used in high-temperature applications. Its high-temperature range, stability, and portability make it suitable for various industrial and laboratory environments.

Additional information

Brand

Fluke Calibration

Specifications: Fluke Calibration 9150 Thermocouple Furnace

General Specifications
Temperature range 150°C to 1200°C
(302°F to 2192°F)
Display resolution 0.1° to 999.9°
1° above 1000°
Stability ±0.5°C
Display accuracy ±5°C
Well diameter 1.25 in(32 mm)
Well depth 140 mm (5.5 in); (101 mm [4 in] removable insert plus 38 mm [1.5 in] insulator)
Heating time 35 minutes to 1200°C
Cooling time 140 minutes with block
Well-to-well uniformity ±2.5°C (Insert “C” at 1200°C)
Stabilization 20 minutes
Power 115 V AC (±10%), 10.5 A or 230 V AC (±10%), 5.2 A, switchable, 50/60 Hz, 1200 W
Size (H x W x D) 315 x 208 x 315 mm
(12.4 x 8.2 x 12.4 in)
Weight 13 kg (28 lb)
NIST-traceable calibration Data at 150°C, 300°C, 450°C, 600°C, 800°C, 1000°C, and 1200°C

 