Fluke 9142 Field Metrology Well

  • Lightweight, portable, and fast
  • Built-in two-channel readout for PRT, RTD, thermocouple, 4-20 mA current
  • True reference thermometry with accuracy to ±0.01 °C
  • On-board automation and documentation
  • Metrology performance in accuracy, stability, uniformity, and loading

Description

The Fluke Calibration 9142 Field Metrology Well is a portable, high-performance dry-block calibrator designed for accurate temperature calibration in various industrial and field applications.

Key Features and Benefits:

  • High Accuracy: Provides exceptional accuracy with stability of ±0.01°C across the entire temperature range, ensuring reliable and trustworthy temperature measurements.
  • Versatility: Capable of calibrating a wide range of temperature sensors, including PRTs, RTDs, thermocouples, and 4-20 mA transmitters.
  • Portability: Compact and lightweight design makes it easy to transport and use in the field, minimizing downtime and increasing efficiency.
  • Fast Heating and Cooling: Reaches target temperatures quickly, significantly reducing calibration time.
  • Built-in Readout: Features a built-in two-channel readout for PRT, RTD, thermocouple, and 4-20 mA current, eliminating the need for additional equipment.
  • On-Board Automation: Supports on-board automation and documentation, streamlining calibration processes and enhancing data management.

Applications:

  • Calibration of temperature sensors and transmitters in various industries, including manufacturing, process control, and research.
  • On-site calibration of process control equipment.
  • Industrial maintenance and troubleshooting.
  • Research and development.
  • Quality control and assurance.

In essence, the Fluke Calibration 9142 is a powerful and versatile tool for professionals who require accurate, efficient, and portable temperature calibration solutions.

In summary, the Fluke Calibration 9142 Field Metrology Well combines high accuracy, portability, and user-friendly features to provide a reliable and efficient solution for temperature calibration needs in diverse industrial and field settings.

Additional information

Brand

Fluke Calibration

Specifications: Fluke Calibration 9142 Field Metrology Well

Base Unit Specifications
Temperature range at 23°C –25°C to 150°C (–13°F to 302°F)
Display accuracy ±0.2°C full range
Stability ±0.01°C full range
Axial uniformity at 40 mm (1.6 in) ±0.05°C full range
Radial uniformity ±0.01°C full range
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes) ±0.006°C full range
Hysteresis 0.025
Operating conditions 0°C to 50°C, 0% to 90% RH (non-condensing)
Environmental conditions (for all specifications except temperature range) 13°C to 33°C
Immersion (well) Depth 150 mm (5.9 in)
Insert OD 30 mm (1.18 in)
Heating time 16 min: 23°C to 140°C
23 min: 23°C to 150°C
25 min: –25°C to 150°C
Cooling time 15 min: 23°C to –25°C
25 min: 150°C to –23°C
Resolution 0.01°
Display LCD, °C or °F user-selectable
Size (H x W x D) 290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in)
Weight 8.16 kg (18 lb)
Power requirements 100 V to 115 V (±10%) 50/60 Hz, 635 W 230 V (±10%) 50/60 Hz, 575 W
Computer interface RS-232 interface
Calibration NVLAP accredited calibration included
Automation Use the 1586A Super-DAQ to automate temperature sensor calibration
-P Specifications
Built-in reference thermometer readout accuracy (4-Wire Reference Probe)1 ±0.010°C at -25°C
±0.015°C at 0°C
±0.020°C at 50°C
±0.025°C at 150°C
±0.030°C at 200°C
±0.040°C at 350°C
±0.050°C at 420°C
±0.070°C at 660°C
Reference resistance range 0 ohms to 400 ohms
Reference resistance accuracy2 0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading
Reference characterizations ITS-90, CVD, IEC-751, Resistance
Reference measurement capability 4-wire
Reference probe connection 6-pin Din with Infocon Technology
Built-in RTD thermometer readout accuracy NI-120: ±0.015°C at 0°C
PT-100 (385): ±0.02°C at 0°C
PT-100 (3926): ±0.02°C at 0°C
PT-100 (JIS): ±0.02°C at 0°C
RTD resistance range 0 ohms to 400 ohms
RTD resistance accuracy2 0 ohms to 25 ohms: ±0.002 ohms
25 ohms to 400 ohms: ±80 ppm of reading
RTD characterizations PT-100 (385),(JIS),(3926), NI-120, Resistance
RTD measurement capability 4-wire RTD (2-,3-wire RTD w/ Jumpers only)
RTD connection 4 terminal input
Built-in TC thermometer readout accuracy Type J: ±0.7°C at 660°C
Type K: ±0.8°C at 660°C
Type T: ±0.8°C at 400°C
Type E: ±0.7°C at 660°C
Type R: ±1.4°C at 660°C
Type S: ±1.5°C at 660°C
Type M: ±1.4°C at 660°C
Type L: ±0.7°C at 660°C
Type U: ±0.75°C at 600°C
Type N: ±0.9°C at 660°C
Type C: ±1.1°C at 660°C
TC millivolt range –10 mV to 75 mV
Voltage accuracy 0.025% of reading + 0.01 mV
Internal cold junction compensation accuracy ±0.35°C (ambient of 13°C to 33°C)
TC connection Small connectors
Built-in mA readout accuracy 0.02% of reading + 2 mV
mA range Cal 4-22 mA, Spec 4-24 mA
mA connection 2 terminal input
Loop power function 24 V DC loop power
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C) ±0.005% of range per °C
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors.
2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.

 

Related Products