Fluke 9132/9133 Infrared Calibrator with Blackbody Source

  • Accurate, reliable performance from –15 °C to 500 °C
  • Small, compact design in a blackbody calibrator
  • These blackbody calibrator units feature a temperature controlled measurement surface with a diameter of 2.25 in (57 mm).
  • Intuitive, easy-to-read display that indicates when temperature is stable.

Description

The Fluke 9132 and 9133 are portable infrared calibrators designed to accurately calibrate infrared thermometers and thermal imagers.

Key Features:

  • Accurate, reliable performance from –15 °C to 500 °C
  • Small, compact design in a blackbody calibrator
  • These blackbody calibrator units feature a temperature controlled measurement surface with a diameter of 2.25 in (57 mm).
  • Intuitive, easy-to-read display that indicates when temperature is stable.
  • High Accuracy: Provide accurate and stable temperature control for reliable calibration results.
  • Temperature Range:
    • 9132: 50°C to 500°C
    • 9133: -30°C to 150°C
  • Portability: Compact and lightweight, making them easy to transport for on-site calibrations.
  • User-Friendly Interface: Intuitive controls and a clear display for simple operation.
  • Blackbody Source: Incorporate a stable blackbody source for accurate infrared temperature simulations.

Applications:

  • Calibration of Infrared Thermometers: Calibrating a wide range of infrared thermometers used in various industries.
  • Thermal Imaging: Calibrating thermal imagers for accurate temperature measurements and analysis.
  • Industrial Maintenance: Used for preventive maintenance and troubleshooting of equipment.
  • Research and Development: Accurate temperature measurements in research and development laboratories.
  • Quality Control: Ensuring product quality and consistency through accurate temperature measurements.

The Fluke 9132 and 9133 are valuable tools for ensuring the accuracy of infrared temperature measurements. Their portability, accuracy, and user-friendly design make them suitable for a wide range of applications in various industries.

Additional information

Brand

Fluke Calibration

Specifications: Fluke 9132/9133 Portable Infrared Calibrator with Blackbody Source

Specifications 9132 9133
Temperature range 50°C to 500°C
(122°F to 932°F)
–30°C to 150°C at 23°C ambient
(–22°F to 302°F at 73°F ambient)
Accuracy ±0.5°C at 100°C
(±0.9°F at 212°F)
±0.8°C at 500°C
(±1.4°F at 932°F)
±0.4°C (±0.72°F)
Stability ±0.1°C at 100°C
(±0.18°F at 212°F)
±0.3°C at 500°C
(±0.54°F at 932°F)
±0.1°C (±0.18°F)
Target size 57 mm (2.25 in) 57 mm (2.25 in)
Target emissivity 0.95 (±0.02 from 8 to 14 mm) 0.95 (±0.02 from 8 to 14 mm)
Resolution 0.1° 0.1°
Heating time 30 minutes (50°C to 500°C) 15 minutes (25°C to 150°C)
Cooling time 30 minutes (500°C to 100°C) 15 minutes (25°C to –20°C)
Computer interface RS-232 I/O included
Power 115 V AC (±10%), 3 A or
230 V AC (±10%), 1.5 A,
switchable, 50/60 Hz, 340 W
115 V AC (±10%), 1.5 A, or
230 V AC (±10%), 1.0 A,
switchable, 50/60 Hz, 200 W
Size (H x W x D) 102 x 152 x 178 mm
(4 x 6 x 7 in )
152 x 286 x 267 mm
(6 x 11.25 x 10.5 in)
Weight 1.8 kg (4 lb) 4.6 kg (10 lb)
NIST-traceable contact calibration Data at 50°C, 100°C, 200°C, 250°C, 300°C, 400°C, and 500°C Data at –30°C, 0°C, 25°C, 75°C, 100°C, 125°C, and 150°C

 