Fluke 9132/9133 Infrared Calibrator with Blackbody Source
- Accurate, reliable performance from –15 °C to 500 °C
- Small, compact design in a blackbody calibrator
- These blackbody calibrator units feature a temperature controlled measurement surface with a diameter of 2.25 in (57 mm).
- Intuitive, easy-to-read display that indicates when temperature is stable.
- Description
- Additional information
- Specifications
Description
The Fluke 9132 and 9133 are portable infrared calibrators designed to accurately calibrate infrared thermometers and thermal imagers.
Key Features:
- Accurate, reliable performance from –15 °C to 500 °C
- Small, compact design in a blackbody calibrator
- These blackbody calibrator units feature a temperature controlled measurement surface with a diameter of 2.25 in (57 mm).
- Intuitive, easy-to-read display that indicates when temperature is stable.
- High Accuracy: Provide accurate and stable temperature control for reliable calibration results.
- Temperature Range:
- 9132: 50°C to 500°C
- 9133: -30°C to 150°C
- Portability: Compact and lightweight, making them easy to transport for on-site calibrations.
- User-Friendly Interface: Intuitive controls and a clear display for simple operation.
- Blackbody Source: Incorporate a stable blackbody source for accurate infrared temperature simulations.
Applications:
- Calibration of Infrared Thermometers: Calibrating a wide range of infrared thermometers used in various industries.
- Thermal Imaging: Calibrating thermal imagers for accurate temperature measurements and analysis.
- Industrial Maintenance: Used for preventive maintenance and troubleshooting of equipment.
- Research and Development: Accurate temperature measurements in research and development laboratories.
- Quality Control: Ensuring product quality and consistency through accurate temperature measurements.
The Fluke 9132 and 9133 are valuable tools for ensuring the accuracy of infrared temperature measurements. Their portability, accuracy, and user-friendly design make them suitable for a wide range of applications in various industries.
Additional information
Brand | Fluke Calibration |
---|
Specifications: Fluke 9132/9133 Portable Infrared Calibrator with Blackbody Source
Specifications | 9132 | 9133 |
Temperature range | 50°C to 500°C (122°F to 932°F) |
–30°C to 150°C at 23°C ambient (–22°F to 302°F at 73°F ambient) |
Accuracy | ±0.5°C at 100°C (±0.9°F at 212°F) ±0.8°C at 500°C (±1.4°F at 932°F) |
±0.4°C (±0.72°F) |
Stability | ±0.1°C at 100°C (±0.18°F at 212°F) ±0.3°C at 500°C (±0.54°F at 932°F) |
±0.1°C (±0.18°F) |
Target size | 57 mm (2.25 in) | 57 mm (2.25 in) |
Target emissivity | 0.95 (±0.02 from 8 to 14 mm) | 0.95 (±0.02 from 8 to 14 mm) |
Resolution | 0.1° | 0.1° |
Heating time | 30 minutes (50°C to 500°C) | 15 minutes (25°C to 150°C) |
Cooling time | 30 minutes (500°C to 100°C) | 15 minutes (25°C to –20°C) |
Computer interface | RS-232 I/O included | |
Power | 115 V AC (±10%), 3 A or 230 V AC (±10%), 1.5 A, switchable, 50/60 Hz, 340 W |
115 V AC (±10%), 1.5 A, or 230 V AC (±10%), 1.0 A, switchable, 50/60 Hz, 200 W |
Size (H x W x D) | 102 x 152 x 178 mm (4 x 6 x 7 in ) |
152 x 286 x 267 mm (6 x 11.25 x 10.5 in) |
Weight | 1.8 kg (4 lb) | 4.6 kg (10 lb) |
NIST-traceable contact calibration | Data at 50°C, 100°C, 200°C, 250°C, 300°C, 400°C, and 500°C | Data at –30°C, 0°C, 25°C, 75°C, 100°C, 125°C, and 150°C |