Fluke 9103/9140 Field Dry-Well

  • Accuracy to ±0.25°C and comes with an RS-232 interface.
  • The 9103 covers below-ambient temperatures as low as –25 °C. The 9103 is stable to ±0.02 °C
  • The 9140 has a temperature range of 35 °C to 350 °C. The 9140 has a stability of ±0.05 °C.
  • Features designed with user convenience and productivity in mind.

Description

The Fluke Calibration 9103 and 9140 are portable, field-ready dry-well calibrators designed for accurate temperature calibration in various industrial and field applications.

Key features and benefits:

  • Accuracy to ±0.25°C and comes with an RS-232 interface.
  • The 9103 covers below-ambient temperatures as low as –25 °C. The 9103 is stable to ±0.02 °C
  • The 9140 has a temperature range of 35 °C to 350 °C. The 9140 has a stability of ±0.05 °C.
  • Features designed with user convenience and productivity in mind.
  • Accuracy: Both models offer high accuracy, ensuring precise temperature calibration for various applications.
  • Portability: Compact and lightweight design makes them easy to transport and use in the field.
  • Versatility: Suitable for calibrating a wide range of temperature sensors, including thermocouples, RTDs, and thermistors.
  • Fast heating and cooling: Reach target temperatures quickly, minimizing calibration time.
  • User-friendly interface: Simple controls and a clear display make them easy to operate.

Key differences between the 9103 and 9140:

  • Temperature range: The 9103 covers a lower temperature range (-25°C to 140°C) compared to the 9140 (35°C to 350°C).
  • Accuracy: The 9103 typically offers slightly higher accuracy than the 9140.

Applications:

  • Calibration of temperature sensors and transmitters
  • Process control
  • Industrial maintenance
  • Research and development
  • Quality control

Overall, the Fluke Calibration 9103 and 9140 are reliable and versatile tools for accurate temperature calibration in various field and industrial settings.

In summary, the Fluke Calibration 9103 and 9140 are portable, accurate, and user-friendly dry-well calibrators that provide efficient and reliable temperature calibration solutions.

Additional information

Brand

Fluke Calibration

Specifications: Fluke Calibration 9103/9140 Field Dry-Well

Specifications 9103 9140
Range –25°C to 140°C
(–13°F to 284°F)
at 23°C ambient
35°C to 350°C
(95°F to 662°F)
Accuracy ±0.25°C (holes greater than 1/4″ [6.35mm]: ±1°C) ±0.5°C (holes greater than 1/4″ [6.35mm]: ±1°C)
Stability ±0.02°C at –25°C
±0.04°C at 140°C
±0.03°C at 50°C
±0.05°C at 350°C
Well-to-well uniformity ±0.1°C between similarly sized wells ±0.1°C with similarly sized wells
Heating times 18 minutes from ambient to 140°C 12 minutes from ambient to 350°C
Cooling times 20 minutes from ambient to –25°C 15 minutes from 350°C to 100°C
Stabilization time 7 minutes 7 minutes
Immersion depth 124 mm (4.875 in) 124 mm (4.875 in)
Inserts Insert A, B, C, or D included (specify when ordering)
Outside insert dimensions 31.8 mm dia. x 124 mm (1.25 x 4.88 in)
Computer interface RS-232 included with free Interface-it software (Model 9930)
Power 115 V AC (±10%), 1.3 A or 230 V AC (±10%), 0.7 A, switchable, 50/60 Hz, 150 W 115 V AC (±10%), 4.4 A or 230 V AC (±10%), 2.2 A, switchable, 50/60 Hz, 500 W
Size (W x H x D) 143 x 261 x 245 mm
(5.63 x 10.25 x 9.63 in)
152 x 86 x 197 mm
(6 x 3.375 x 7.75 in)
Weight 5.7 kg (12 lb) 2.7 kg (6 lb)
NIST-traceable certificate Data at –25°C, 0°C, 25°C, 50°C, 75°C, 100°C, and 140° Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C, and 350°C

 