Fluke 730G Digital Pressure Calibrator

  • Pressure measurement from ±10 inH2O/2.5 kPa to 10000 psi/69 MPa
  • Four absolute pressure measurement ranges: 15 PSIA (103 kPa), 30 PSIA (207 kPa), 100 PSIA (690 kPa), 300 PSIA (2 MPa)
  • Pressure measurement accuracy 0.05% of full scale, 0.02% optional
  • Current measurement range: 0-24mA
  • Voltage measurement range: 0-30VDC
  • 0.01% current and voltage accuracy
  • Powerful HART communication, supports HART 7
  • 24V loop power, no external power supply needed
  • Lithium battery life: 25 hours continuously (without 24 V loop power), or 8 hours continuously (24 V loop power)
  • Storage up to 128 sets of calibration data
  • Ingress protection: IP54
  • 1-meter drop test

Description

The Fluke 730G is a high-performance digital pressure calibrator designed for a variety of industrial applications.

Key Features:

  • Pressure measurement from ±10 inH2O/2.5 kPa to 10000 psi/69 MPa
  • Four absolute pressure measurement ranges: 15 PSIA (103 kPa), 30 PSIA (207 kPa), 100 PSIA (690 kPa), 300 PSIA (2 MPa)
  • Pressure measurement accuracy 0.05% of full scale, 0.02% optional
  • Current measurement range: 0-24mA
  • Voltage measurement range: 0-30VDC
  • 0.01% current and voltage accuracy
  • Powerful HART communication, supports HART 7
  • 24V loop power, no external power supply needed
  • Lithium battery life: 25 hours continuously (without 24 V loop power), or 8 hours continuously (24 V loop power)
  • Storage up to 128 sets of calibration data
  • Ingress protection: IP54
  • 1-meter drop test
  • Precision Pressure Measurement: Offers high-accuracy pressure measurements for precise calibration of pressure instruments.
  • Built-in HART Communication: Allows direct adjustment of HART devices, simplifying calibration processes.
  • Loop Calibration: Functions as a loop calibrator, enabling testing and calibration of process transmitters and other field devices.
  • Rugged and Durable: Designed to withstand harsh industrial environments for reliable performance.
  • User-Friendly Interface: Intuitive controls and a clear display make operation easy and efficient.

Applications:

  • Pressure Transmitter Calibration: Calibrates a wide range of pressure transmitters used in various industries, including oil and gas, chemical processing, and power generation.
  • Process Control: Used to test and calibrate instruments involved in process control systems, ensuring accurate and reliable process operation.
  • Field Service: A valuable tool for field technicians performing on-site calibration and troubleshooting of pressure instruments.

In essence, the Fluke 730G combines the precision of a pressure gauge with the functionality of a loop calibrator and HART communication capabilities, making it a versatile and efficient tool for pressure calibration professionals.

Additional information

Brand

Fluke Calibration

Specifications: Fluke 730G Digital Pressure Calibrator

Technical specifications
Pressure specifications: 0.02% of full scale (730RG), 0.05% of full scale (730G)
Current specifications: 0.01% of reading
Model Range Resolution Accuracy1 Burst pressure
730G01 -10 to +10 inH2O
-20 to 20 mbar
-2.49 to 2.49 kPa
0.001 inH2O
0.001 mbar
0.0001 kPa
Positive pressure, ±0.2% of range2
Vacuum, ±0.2% of range2
3 psi,
210 mbar
20.6 kPa
730G02 -1 to +1 psi
-70 to 70 mbar
-6.89 to 6.89 kPa
0.0001 psi
0.001 mbar
0.0001 kPa
5 psi,
350 mbar
34.4 kPa
730G04 -14 to 15 psi
-0.97 to 1 bar
-97 to 103 kPa
0.001 psi
0.0001 bar
0.01 kPa
Positive pressure, ±0.05% of range
Vacuum, ±0.1% of range
60 psi
4 bar
413.6 kPa
730G05 -14 to 30 psi
-0.97 to 2 bar
-97 to 206 kPa
0.001 psi
0.0001 bar
0.01 kPa
120 psi
8 bar
827.3 kPa
730G06 -12 to 100 psi
-0.83 to 6.9 bar
-83 to 690 kPa
0.01 psi, 0.0001 bar 400 psi
26 bar
2.757 MPa
730G27 -12 to 300 psi
-0.83 to 20 bar
-83 to 2 MPa
0.01 psi
0.001 bar
0.1 kPa
1200 psi
80 bar
8.273 MPa
730G07 -12 to 500 psi
-0.83 to 34 bar
-83 to 3.4 MPa
0.01 psi
0.001 bar
0.1 kPa
2000 psi
138 bar
13.789 MPa
730G08 -14 to 1000 psi,
-0.97 to 69 bar
-97 to 6900 kPa
0.01 psi
0.001 bar
0.1 kPa
4000 psi
266 bar
27.579 MPa
730G10 -14 to 2000 psi,
-0.97 to 140 bar
-97 to 14 MPa
0.1 psi
0.01 bar
1 kPa
8000 psi
550 bar
55.158 MPa
730G29 -14 to 3000 psi,
-0.97 to 200 bar
-97 to 20 MPa
0.1 psi
0.01 bar
1 kPa
10000 psi
690 bar
68.947 MPa
730G30 -14 to 5000 psi,
-0.97 to 340 bar
-97 to 34 MPa
0.1 psi
0.01 bar
1 kPa
10000 psi
690 bar
68.947 MPa
730G31 -14 to 10000 psi,
-0.97 to 690 bar
-97 to 69 MPa
1 psi
0.01 bar
1 kPa
15000 psi
1035 bar
103.421 MPa
730GA4 0 to 15 psia,
0 to 1 bar
0 to 100 kPa absolute pressure
0.001 psi
0.0001 bar
0.01 kPa
±0.05% of range4 60 psi
4 bar
413.6 kPa
730GA5 0 to 30 psia
0 to 2 bar
0 to 200 kPa absolute pressure
0.001 psi
0.0001 bar
0.01 kPa
120 psi
8 bar
827.3 kPa
730GA6 0 to 100 psia,
0 to 6.9 bar
0 to 690 kPa absolute pressure
0.01 psi
0.0001 bar
0.01 kPa
400 psi
27 bar
2.757 MPa
730GA27 0 to 300 psia
0 to 20 bar
0 to 2 MPa absolute
0.01 psi
0.001 bar
.01 kPa
1200 psi
80 bar
8.273 MPa
730RG04 -14 to 15 psi,
-0.97 to 1 bar
-97 to 100 kPa
0.001 psi
0.0001 bar
0.0001 kPa
Positive pressure, ±0.02 % of range3
Vacuum, ±0.05 % of range3
45 psi
3 bar
310.2 kPa
730RG05 -14 to 30 psi,
-0.97 to 2 bar
-97 to 200 kPa
0.001 psi
0.0001 bar
0.01 kPa
90 psi
6 bar
620.5 kPa
730RG06 -12 to 100 psi,
-0.83 to 6.9 bar
-83 to 690 kPa
0.01 psi
0.0001 bar
0.01 kPa
1000 psi
69 bar
6.894 MPa
730RG27 –12 psi to 300 psi
-0.83 to 20 bar
–80 kPa to 2 MPa
0.01 psi
0.001 bar
0.1 kPa
2000 psi
138 bar
14 MPa
730RG07 –12 psi to 500 psi
-0.83 to 34 bar
–80 kPa to 3.5 MPa
0.01 psi
0.001 bar
0.1 kPa
2000 psi
138 bar
13.789 MPa
730RG08 -14 to 1000 psi,
-0.97 to 69 bar
-97 to 6900 kPa
0.1 psi
0.001 bar
0.1 kPa
10000 psi
690 bar
68.947 MPa
730RG29 -14 to 3000 psi
-0.97 to 200 bar
-97 to 20 MPa
0.1 psi
0.01 bar
1 kPa
10000 psi
690 bar
68.947 MPa
730RG30 -14 to 5000 psi,
-0.97 to 340 bar
-97 to 34 MPa
0.1 psi
0.01 bar
1 kPa
10000 psi
690 bar
68.947 MPa
730RG31 -14 to 10000 psi,
-0.97 to 690 bar
-97 to 69 MPa
0.1 psi
0.01 bar
1 kPa
15000 psi
1035 bar
103.421 MPa
Accuracy specifications above have considered the effects of linearity, repeatability, resolution, reference accuracy, 1-year stability, and temperature effect.
2 For 730G01/730G02, accuracy is ±0.3% of full scale @ 0 °C to 15 °C and 35 °C to 50 °C.
3 For all 730RG model, add 0.003% FS/°C @ 0 °C to 18 °C and 28 °C to 50 °C。
4 For 730GA4, add 0.006% FS/°C @ -10 °C to 15 °C and 35 °C to 55 °C; for all other 730GA model, add 0.003% FS/°C @ -10 °C to 15 °C and 35 °C to 55 °C.
Media Compatibility
730G01, 730G02, 730G04, 730G05, 730RG04, 730RG05: any clean dry non-corrosive gas
Above 100 psi: any liquids or gases compatible with 316 stainless steel

 

Electrical specifications (All specifications are valid to 110% of range)
Current 0 mA to 24 mA
Voltage 0 V DC to 30 V DC
Current range 1 μA
Voltage range 1 mV
Accuracy 0.01% ± 2 LSD all ranges @ 23°C ±5 °C
Temperature compensation From 20% FS/°C @ -10 °C to +18 °C and 28 °C to 50 °C
Loop Compliance Voltage 24 V DC @ 20 mA
Temperature measurement accuracy inside the product ±0.5 ℃
Drive Capability 1200 Ω (without HART resistor), 950 Ω (with internal HART resistor)

 