Fluke 725 Multifunction Process Calibrator

  • Calibration of both temperature and pressure transmitters
  • Measure/source/simulate mA loop parameters
  • Measure and source mA, Volts DC, frequency, and resistance
  • Built-in 24 V loop power supply
  • Store and recall frequently used calibration set ups

Description

The Fluke 725 Multifunction Process Calibrator is a versatile tool designed to simplify the calibration of various process parameters. It combines the ability to both generate and simulate signals with precise measurement functions, making it a valuable asset for technicians and engineers across diverse industries.

This calibrator excels at sourcing and simulating signals such as milliamps (mA), volts, temperature (for both Resistance Temperature Detectors – RTDs and thermocouples), frequency, and ohms. It also provides highly accurate measurement capabilities for these parameters.

The 725 is designed with user-friendliness in mind. Its intuitive interface features clear and concise displays, making it easy to navigate and operate, even for those with limited experience in calibration. Built to withstand the demands of industrial environments, the 725 is robust and reliable, ensuring consistent performance in challenging conditions.

Furthermore, the 725 incorporates features that streamline the calibration process, such as fast and accurate measurements, which significantly improve efficiency.

Applications:

  • Process Plants: Calibration of pressure transmitters, flow meters, temperature sensors, and other critical process instrumentation.
  • Oil and Gas: Calibration of equipment in refineries, pipelines, and offshore platforms.
  • Power Generation: Calibration of power plant instrumentation and control systems.
  • Manufacturing: Calibration of production equipment and quality control instruments.

The Fluke 725 empowers technicians and engineers to perform accurate and efficient calibrations, ensuring the optimal performance and reliability of critical equipment across various industries.

Disclaimer: This overview is based on general information about the Fluke 725. For the most accurate and up-to-date information, please refer to the official Fluke product documentation.

Additional information

Brand

Fluke Calibration

Specifications: Fluke 725 Multifunction Process Calibrator

Measurement Accuracy
Voltage DC 30.000 V 0.02% + 2 counts (upper display)
30.000 V 0.02% + 2 counts (lower display)
100.00 mV 0.02% + 2 counts
-10.00 mV to 75.00 mV 0.025% + 1 count (via TC connector)
Current DC 24.000 mA 0.02% + 2 counts
Resistance 0.0 to 400.0 Ω 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)
401 to 1500 Ω 0.5 Ω (4-wire), 1 Ω (2- and 3-wire)
1500 to 3200 Ω 1 Ω (4-wire), 1.5 Ω (2- and 3-wire)
Frequency 2.0 to 1000.0 CPM 0.05% + 1 count
1.0 to 1100.0 Hz 0.05% + 1 count
1.00 to 10.00 kHz 0.05% + 1 count
Sensitivity 1 V peak-to-peak-minimum
Pressure Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC 100.00 mV 0.02% + 2 counts
10.000 V 0.02% + 2 counts
-10.00 mV to 75.00 mV 0.025% + 1 count (via TC connector)
Current DC 24.000 mA (source) 0.02% + 2 counts
24.000 mA (simulate) 0.02% + 2 counts
Resistance 15.0 to 400.0 Ω 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
401 to 1500 Ω 0.5 Ω (excitation current 0.05 to 0.8 mA)
1500 to 3200 Ω 1 Ω (excitation current 0.05 to 0.4 mA)
Frequency 2.0 to 1000.0 CPM 0.05%
1.0 to 1100.0 Hz 0.05%
1.00 to 10.00 kHz 0.25%
Waveform 5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure accuracy NI-120 0.2°C
PT-100 (385) 0.33°C
PT-100 (393) 0.3°C
PT-100 (JIS) 0.3°C
PT-200 (385) 0.2°C
PT-500 (385) 0.3°C
PT-1000 (385) 0.2°C
Resolution 0.1°C
J 0.7°C
K 0.8°C
T 0.8°C
E 0.7°C
R 1.8°C
S 1.5°C
B 1.4°C
L 0.7°C
U 0.75°C
N 0.9°C
Resolution J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F
XK 0.6°C
BP 1.2°C
Source accuracy NI-120 0.2°C
PT-100 (385) 0.33°C
PT-100 (393) 0.3°C
PT-100 (JIS) 0.3°C
PT-200 (385) 0.2°C
PT-500 (385) 0.3°C
PT-1000 (385) 0.2°C
Resolution 0.1°C
Accuracy stated for 4-wire measurement.
J 0.7°C
K 0.8°C
T 0.8°C
E 0.7°C
R 1.4°C
S 1.5°C
B 1.4°C
L 0.7°C
U 0.75°C
N 0.9°C
Resolution J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C
XK 0.6°C
BP 1.2°C
Specifications
Ramp functions Source functions Voltage, current, resistance, frequency, temperature
Ramps Slow ramp, Fast ramp, 25% step-ramp
Loop power function Voltage 24 V
Accuracy 10%
Maximum current 22 mA, short circuit protected
Step functions Source functions Voltage, current, resistance, frequency, temperature
Steps 25% of range, 100% of range
Environmental Specifications
Operating temperature -10°C to +55°C
Storage temperature -20°C to 71°C
Operating altitude 3000 m
Safety Specifications
Agency approvals EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992
Mechanical and General Specifications
Size 130 x 236 x 61 mm
Weight 0.65 kg
Batteries 4 AA alkaline batteries
Battery replacement Separate battery compartment, accessable without breaking calibration seal
Side port connections Pressure module connector, also used for remote real-time programming

 