Fluke 725 Multifunction Process Calibrator
- Calibration of both temperature and pressure transmitters
- Measure/source/simulate mA loop parameters
- Measure and source mA, Volts DC, frequency, and resistance
- Built-in 24 V loop power supply
- Store and recall frequently used calibration set ups
- Description
- Additional information
- Specifications
Description
The Fluke 725 Multifunction Process Calibrator is a versatile tool designed to simplify the calibration of various process parameters. It combines the ability to both generate and simulate signals with precise measurement functions, making it a valuable asset for technicians and engineers across diverse industries.
This calibrator excels at sourcing and simulating signals such as milliamps (mA), volts, temperature (for both Resistance Temperature Detectors – RTDs and thermocouples), frequency, and ohms. It also provides highly accurate measurement capabilities for these parameters.
The 725 is designed with user-friendliness in mind. Its intuitive interface features clear and concise displays, making it easy to navigate and operate, even for those with limited experience in calibration. Built to withstand the demands of industrial environments, the 725 is robust and reliable, ensuring consistent performance in challenging conditions.
Furthermore, the 725 incorporates features that streamline the calibration process, such as fast and accurate measurements, which significantly improve efficiency.
Applications:
- Process Plants: Calibration of pressure transmitters, flow meters, temperature sensors, and other critical process instrumentation.
- Oil and Gas: Calibration of equipment in refineries, pipelines, and offshore platforms.
- Power Generation: Calibration of power plant instrumentation and control systems.
- Manufacturing: Calibration of production equipment and quality control instruments.
The Fluke 725 empowers technicians and engineers to perform accurate and efficient calibrations, ensuring the optimal performance and reliability of critical equipment across various industries.
Disclaimer: This overview is based on general information about the Fluke 725. For the most accurate and up-to-date information, please refer to the official Fluke product documentation.
Additional information
Brand | Fluke Calibration |
---|
Specifications: Fluke 725 Multifunction Process Calibrator
Measurement Accuracy | ||
Voltage DC | 30.000 V | 0.02% + 2 counts (upper display) |
30.000 V | 0.02% + 2 counts (lower display) | |
100.00 mV | 0.02% + 2 counts | |
-10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
Current DC | 24.000 mA | 0.02% + 2 counts |
Resistance | 0.0 to 400.0 Ω | 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire) |
401 to 1500 Ω | 0.5 Ω (4-wire), 1 Ω (2- and 3-wire) | |
1500 to 3200 Ω | 1 Ω (4-wire), 1.5 Ω (2- and 3-wire) | |
Frequency | 2.0 to 1000.0 CPM | 0.05% + 1 count |
1.0 to 1100.0 Hz | 0.05% + 1 count | |
1.00 to 10.00 kHz | 0.05% + 1 count | |
Sensitivity | 1 V peak-to-peak-minimum | |
Pressure | Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure. | |
Source Accuracy | ||
Voltage DC | 100.00 mV | 0.02% + 2 counts |
10.000 V | 0.02% + 2 counts | |
-10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
Current DC | 24.000 mA (source) | 0.02% + 2 counts |
24.000 mA (simulate) | 0.02% + 2 counts | |
Resistance | 15.0 to 400.0 Ω | 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
401 to 1500 Ω | 0.5 Ω (excitation current 0.05 to 0.8 mA) | |
1500 to 3200 Ω | 1 Ω (excitation current 0.05 to 0.4 mA) | |
Frequency | 2.0 to 1000.0 CPM | 0.05% |
1.0 to 1100.0 Hz | 0.05% | |
1.00 to 10.00 kHz | 0.25% | |
Waveform | 5 V p-p squarewave, -0.1 V offset | |
RTDs and Thermocouples | ||
Measure accuracy | NI-120 | 0.2°C |
PT-100 (385) | 0.33°C | |
PT-100 (393) | 0.3°C | |
PT-100 (JIS) | 0.3°C | |
PT-200 (385) | 0.2°C | |
PT-500 (385) | 0.3°C | |
PT-1000 (385) | 0.2°C | |
Resolution | 0.1°C | |
J | 0.7°C | |
K | 0.8°C | |
T | 0.8°C | |
E | 0.7°C | |
R | 1.8°C | |
S | 1.5°C | |
B | 1.4°C | |
L | 0.7°C | |
U | 0.75°C | |
N | 0.9°C | |
Resolution | J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F | |
XK | 0.6°C | |
BP | 1.2°C | |
Source accuracy | NI-120 | 0.2°C |
PT-100 (385) | 0.33°C | |
PT-100 (393) | 0.3°C | |
PT-100 (JIS) | 0.3°C | |
PT-200 (385) | 0.2°C | |
PT-500 (385) | 0.3°C | |
PT-1000 (385) | 0.2°C | |
Resolution | 0.1°C | |
Accuracy stated for 4-wire measurement. | ||
J | 0.7°C | |
K | 0.8°C | |
T | 0.8°C | |
E | 0.7°C | |
R | 1.4°C | |
S | 1.5°C | |
B | 1.4°C | |
L | 0.7°C | |
U | 0.75°C | |
N | 0.9°C | |
Resolution | J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C | |
XK | 0.6°C | |
BP | 1.2°C | |
Specifications | ||
Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature |
Ramps | Slow ramp, Fast ramp, 25% step-ramp | |
Loop power function | Voltage | 24 V |
Accuracy | 10% | |
Maximum current | 22 mA, short circuit protected | |
Step functions | Source functions | Voltage, current, resistance, frequency, temperature |
Steps | 25% of range, 100% of range | |
Environmental Specifications | ||
Operating temperature | -10°C to +55°C | |
Storage temperature | -20°C to 71°C | |
Operating altitude | 3000 m | |
Safety Specifications | ||
Agency approvals | EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 | |
Mechanical and General Specifications | ||
Size | 130 x 236 x 61 mm | |
Weight | 0.65 kg | |
Batteries | 4 AA alkaline batteries | |
Battery replacement | Separate battery compartment, accessable without breaking calibration seal | |
Side port connections | Pressure module connector, also used for remote real-time programming |