Fluke 724 Temperature Calibrator
- Powerful, easy-to-use, high-accuracy temperature calibrator
- Test and calibrate temperature sensors and transmitters
- Measures 4 to 20 mA loops and can provide loop power
- Measures RTDs, thermocouples, ohms, and volts
- Source/simulates thermocouples, RTDs, volts, and ohms
- Description
- Additional information
- Specifications
Description
The Fluke 724 is a versatile and powerful temperature calibrator designed for professionals working in various industries. It’s known for its ability to efficiently and accurately calibrate a wide range of temperature sensors and transmitters.
Key features and benefits:
- Powerful, easy-to-use, high-accuracy temperature calibrator
- Test and calibrate temperature sensors and transmitters
- Measures 4 to 20 mA loops and can provide loop power
- Measures RTDs, thermocouples, ohms, and volts
- Source/simulates thermocouples, RTDs, volts, and ohms
- Versatile temperature measurement and sourcing: Measures and sources 10 thermocouple types and 7 RTD types, covering a broad spectrum of temperature measurement needs.
- Dual-line display: Allows simultaneous viewing of both input measurements and sourced signals, enhancing efficiency and clarity.
- Loop power and mA measurement: Provides loop power to transmitters and measures the resulting 4-20 mA current, enabling comprehensive testing of process control loops.
- Rugged and durable: Built to withstand the demands of field environments, ensuring reliable performance in challenging conditions.
- Easy to use: Intuitive interface with simple controls, minimizing training time and maximizing productivity.
Applications:
- Calibration of thermocouples and RTDs in various industries, including manufacturing, process control, and research.
- Testing and maintenance of temperature-sensitive equipment.
- Process control optimization.
- Research and development.
Overall, the Fluke 724 is a valuable tool for professionals who require a reliable and efficient solution for temperature calibration and testing.
Additional information
Brand | Fluke Calibration |
---|
Specifications: Fluke 724 Temperature Calibrator
Measurement Accuracy | ||
Voltage DC | 30.000 V | 0.02% + 2 counts (upper display) |
20.000 V | 0.02% + 2 counts (lower display) | |
100.00 mV | 0.02% + 2 counts | |
-10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
Current DC | 24.000 mA | 0.02% + 2 counts |
Resistance | 0.0 to 400.0 Ω | 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire) |
401 to 1500 Ω | 0.5 Ω (4-wire), 1 Ω (2- and 3-wire) | |
1500 to 3200 Ω | 1 Ω (4-wire), 1.5 Ω (2- and 3-wire) | |
Source Accuracy | ||
Voltage DC | 100.00 mV | 0.02% + 2 counts |
10.000 V | 0.02% + 2 counts | |
-10.00 mV to 75.00 mV | 0.025 % + 1 count (via TC connector) | |
Resistance | 15.0 to 400.0 Ω | 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
401 to 1500 Ω | 0.5 Ω (excitation current 0.05 to 0.8 mA) | |
1500 to 3200 Ω | 1 Ω (excitation current 0.05 to 0.4 mA) | |
RTDs and Thermocouples | ||
Measure accuracy | NI-120 | 0.2°C |
PT-100 (385) | 0.33°C | |
PT-100 (393) | 0.3°C | |
PT-100 (JIS) | 0.3°C | |
PT-200 (385) | 0.2°C | |
PT-500 (385) | 0.3°C | |
PT-1000 (385) | 0.2°C | |
Resolution | 0.1°C | |
J | 0.7°C | |
K | 0.8°C | |
T | 0.8°C | |
E | 0.7°C | |
R | 1.8°C | |
S | 1.5°C | |
B | 1.4°C | |
L | 0.7°C | |
U | 0.75°C | |
N | 0.9°C | |
Resolution | J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F | |
XK | 0.6°C | |
BP | 1.2°C | |
Source accuracy | NI-120 | 0.2°C |
PT-100 (385) | 0.33°C | |
PT-100 (393) | 0.3°C | |
PT-100 (JIS) | 0.3°C | |
PT-200 (385) | 0.2°C | |
PT-500 (385) | 0.3°C | |
PT-1000 (385) | 0.2°C | |
Resolution | 0.1°C | |
Accuracy stated for 4-wire measurement. | ||
J | 0.7°C | |
K | 0.8°C | |
T | 0.8°C | |
E | 0.7°C | |
R | 1.4°C | |
S | 1.5°C | |
B | 1.4°C | |
L | 0.7°C | |
U | 0.75°C | |
N | 0.9°C | |
Resolution | J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C | |
XK | 0.6°C | |
BP | 1.2°C | |
Specifications | ||
Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature |
Ramps | Slow ramp, Fast ramp, 25% step-ramp | |
Loop power function | Voltage | 24 V |
Accuracy | 10% | |
Maximum current | 22 mA, short circuit protected | |
Step functions | Source functions | Voltage, current, resistance, frequency, temperature |
Steps | 25% of range, 100% of range | |
Environmental Specifications | ||
Operating temperature | -10°C to 55°C | |
Storage temperature | -20°C to 71°C | |
Humidity (without condensation) | 90% | 10°C to 30°C |
75% | 30°C to 40°C | |
45% | 40°C to 50°C | |
35% | 50°C to 55°C | |
Safety Specifications | ||
Safety rating | CSA C22.2 No. 1010.1:1992 | |
EMC | EN50082-1:1992 and EN55022:1994 Class B | |
Mechanical and General Specifications | ||
Size | 96 x 200 x 47 mm | |
Weight | 650 g | |
Batteries | Four AA alkaline batteries | |
Warranty | Three years | |
Battery life | 25 hours typical | |
Shock and vibration | Random, 2 G, 5-500 Hz |