Fluke 714B Thermocouple Calibrator
- Highly accurate, easy-to-use thermocouple temperature calibrator
- Measures and simulates 17 thermocouple types and millivolts
- Measures 4 to 20 mA signals while sourcing a temperature signal
- Comes with a magnetic hanger for convenient hands-free operation
- Features configurable source settings for quick linearity checks
- Description
- Additional information
- Specifications
Description
The Fluke 714B is a versatile and robust handheld calibrator designed specifically for temperature calibration professionals. It combines precision, ease of use, and durability to streamline calibration tasks across various industries.
Core Capabilities:
- Comprehensive Thermocouple Support:
- Measures and sources signals for 17 common thermocouple types, including Type K, J, T, E, N, R, S, B, C, and more. This broad compatibility ensures adaptability to diverse temperature measurement scenarios.
- Millivolt Source Function:
- Enables calibration of linear thermocouple transmitters by directly sourcing millivolt signals, offering precise control over the output.
- 4-20 mA Signal Measurement:
- Simultaneously measures 4-20 mA signals while sourcing a temperature signal, providing valuable insights into the overall system behavior.
- Dual Inputs and Backlit Display:
- Facilitates clear and simultaneous viewing of both input measurements and sourced signals, enhancing operational efficiency.
- Intuitive Interface:
- The 714B boasts an easy-to-navigate interface with clearly labeled buttons and a well-organized display, minimizing learning curves and maximizing productivity.
- Rugged Construction:
- Designed to withstand the demands of field environments, the 714B is built with durable materials and robust construction, ensuring reliable performance in challenging conditions.
- Portability and Convenience:
- Compact and lightweight, the 714B is easily transportable, making it ideal for on-site calibrations and field service applications.
Applications:
- Calibration of Thermocouples and RTDs:
- Forms the cornerstone of accurate temperature measurement in various industries, including manufacturing, process control, and research.
- Process Control:
- Ensures optimal process performance by verifying the accuracy of temperature sensors used in critical control loops.
- Industrial Maintenance:
- Facilitates proactive maintenance of temperature-sensitive equipment by identifying and addressing potential calibration issues.
In essence, the Fluke 714B stands as a reliable and indispensable tool for temperature calibration professionals. Its comprehensive capabilities, user-friendly design, and rugged construction make it a valuable asset in optimizing temperature measurement accuracy and ensuring the integrity of critical processes.
Additional information
Brand | Fluke Calibration |
---|
Specifications: Fluke 714B Thermocouple Calibrator
General Specifications | ||
Maximum voltage applied between any terminal and earth ground or between any two terminals | 30 V | |
Operating temperature | -10°C to 50°C | |
Storage temperature | -30°C to 60°C | |
Operating altitude | 2,000 m | |
Storage altitude | 12,000 m | |
Relative humidity (% RH operating without condensation) | Non condensing 90% (10°C to 30°C) 75% (30°C to 40°C) 45% (40°C to 50°C) (Without condensation) |
|
Vibration requirements | MIL-T-28800E, Class 2 | |
Drop text requirements | 1 m | |
IP Rating | IEC 60529: IP52 | |
Electromagnetic environment | IEC 61326-1, Portable | |
Safety | IEC 61010-1, Max 30 V to earth, Pollution Degree 2 | |
Power supply | 4 AA NEDA 1.5A IEC LR6 batteries | |
Size (H x W x L) | 52.5 x 84 x 188.5 mm | |
Weight | 515 g | |
DC mA Measurement | ||
Resolution | Range | Accuracy (% of reading + counts) |
0-24 mA | 0.001 mA | 0.010% + 2 μA |
Temperature coefficient | ±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C) | |
Millivolt Measurement and Source | ||
Resolution | Range | Accuracy (% of reading +counts) |
-10 mV to 75 mV | 0.01 mV | 0.015% + 10 μA |
Temperature coefficient | ±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C) |
Thermocouple Input and Output | |||||
TC Type | Range (°C) | Measure (°C) | Source (°C) | ||
1 year | 2 year | 1 year | 2 year | ||
E | -250 to 200°C | 1.3 | 2.0 | 0.6 | 0.9 |
-200 to -100°C | 0.5 | 0.8 | 0.3 | 0.4 | |
-100 to 600°C | 0.3 | 0.4 | 0.3 | 0.4 | |
600 to 1000°C | 0.4 | 0.6 | 0.2 | 0.3 | |
N | -200 to -100°C | 1.0 | 1.5 | 0.6 | 0.9 |
-100 to 900°C | 0.5 | 0.8 | 0.5 | 0.8 | |
900 to 1300°C | 0.6 | 0.9 | 0.3 | 0.4 | |
J | -210 to -100°C | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800°C | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 1200°C | 0.5 | 0.8 | 0.3 | 0.3 | |
K | -200 to -100°C | 0.7 | 1.0 | 0.4 | 0.6 |
-100 to 400°C | 0.3 | 0.4 | 0.3 | 0.4 | |
400 to 1200°C | 0.5 | 0.8 | 0.3 | 0.4 | |
1200 to 1372°C | 0.7 | 1.0 | 0.3 | 0.4 | |
T | -250 to -200°C | 1.7 | 2.5 | 0.9 | 1.4 |
-200 to 0°C | 0.6 | 0.9 | 0.4 | 0.6 | |
0 to 400°C | 0.3 | 0.4 | 0.3 | 0.4 | |
B | 600 to 800°C | 1.3 | 2.0 | 1.0 | 1.5 |
800 to 1000°C | 1.0 | 1.5 | 0.8 | 1.2 | |
1000 to 1820°C | 0.9 | 1.3 | 0.8 | 1.2 | |
R | -20 to 0°C | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 100°C | 1.5 | 2.2 | 1.1 | 1.7 | |
100 to 1767°C | 1.0 | 1.5 | 0.9 | 1.4 | |
S | -20 to 0°C | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 200°C | 1.5 | 2.1 | 1.1 | 1.7 | |
200 to 1400°C | 0.9 | 1.4 | 0.9 | 1.4 | |
1400 to 1767°C | 1.1 | 1.7 | 1.0 | 1.5 | |
C | 0 to 800°C | 0.6 | 0.9 | 0.6 | 0.9 |
800 to 1200°C | 0.8 | 1.2 | 0.7 | 1.0 | |
1200 to 1800°C | 1.1 | 1.6 | 0.9 | 1.4 | |
1800 to 2316°C | 2.0 | 3.0 | 1.3 | 2.0 | |
L | -200 to -100°C | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800°C | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 900°C | 0.5 | 0.8 | 0.2 | 0.3 | |
U | -200 to 0°C | 0.6 | 0.9 | 0.4 | 0.6 |
0 to 600°C | 0.3 | 0.4 | 0.3 | 0.4 | |
BP | 0 to 1000°C | 1.0 | 1.5 | 0.4 | 0.6 |
1000 to 2000°C | 1.6 | 2.4 | 0.6 | 0.9 | |
2000 to 2500°C | 2.0 | 3.0 | 0.8 | 1.2 | |
XK | -200 to 300°C | 0.2 | 0.3 | 0.2 | 0.5 |
300 to 800°C | 0.4 | 0.6 | 0.3 | 0.6 | |
G | 100 to 300°C | 1.6 | 2.4 | 1.2 | 1.8 |
300 to 1500°C | 1.0 | 1.5 | 1.0 | 1.5 | |
1500 to 2320°C | 2.0 | 3.0 | 1.6 | 2.4 | |
D | 0 to 300°C | 1.6 | 2.4 | 1.2 | 1.8 |
300 to 1500°C | 1.0 | 1.5 | 1.0 | 1.5 | |
1500 to 2315°C | 2.0 | 3.0 | 1.6 | 2.4 | |
P | 0 to 1000°C | 1.6 | 2.4 | 0.6 | 0.9 |
1000 to 1395°C | 2.0 | 3.0 | 0.8 | 1.2 | |
M | -50 to 100°C | 1.0 | 1.5 | 0.4 | 0.6 |
100 to 1000°C | 1.6 | 2.4 | 0.6 | 0.9 | |
1000 to 1410°C | 2.0 | 3.0 | 0.8 | 1.2 |