Fluke 712B RTD Calibrators
- Highly accurate, full-featured , easy-to-use single function RTD temperature calibrator
- Measures and simulates 13 different RTD types and resistance
- Measures 4 to 20 mA signals while simultaneously sourcing a temperature signal
- Comes with a magnetic hanging tool designed in for convenient hands-free operation
- Features configurable 0% and 100% source settings for quick 25% linearity checks
- Description
- Additional information
- Specifications
Description
The Fluke 712B is a specialized handheld calibrator designed for professionals working with RTD (Resistance Temperature Detectors) temperature sensors.
Key Features and Benefits:
- Highly accurate, full-featured , easy-to-use single function RTD temperature calibrator
- Measures and simulates 13 different RTD types and resistance
- Measures 4 to 20 mA signals while simultaneously sourcing a temperature signal
- Comes with a magnetic hanging tool designed in for convenient hands-free operation
- Features configurable 0% and 100% source settings for quick 25% linearity checks
- RTD Focus: Specifically designed for measuring and simulating 13 different RTD types, making it ideal for calibrating a wide range of RTD-based temperature sensors.
- Resistance Measurement: Measures resistance directly, providing accurate readings for RTD calibration.
- 4-20 mA Signal Measurement: Allows simultaneous measurement of 4-20 mA signals while sourcing a temperature signal, facilitating comprehensive testing of process control loops.
- User-Friendly Interface: Intuitive design with easy-to-use controls and a clear display for efficient operation.
- Rugged and Portable: Built to withstand the demands of field environments, ensuring reliable performance in challenging conditions.
Applications:
- Calibration of RTD temperature sensors in various industries, including manufacturing, process control, and research.
- Testing and maintenance of RTD-based equipment.
- Process control optimization.
- Research and development.
In essence, the Fluke 712B is a valuable tool for professionals who require a dedicated and efficient solution for calibrating RTD temperature sensors.
Additional information
| Brand | Fluke Calibration |
|---|
Specifications: Fluke 712B RTD Calibrators
| General Specifications | |
| Maximum voltage applied between any terminal and earth ground or between any two terminals | 30 V |
| Operating temperature | -10°C to 50°C |
| Storage temperature | -30°C to 60°C |
| Operating altitude | 2,000 m |
| Storage altitude | 12,000 m |
| Relative humidity (% RH operating without condensation) | Non condensing |
| 90% (10°C to 30°C) | |
| 75% (30°C to 40°C) | |
| 45% (40°C to 50°C) | |
| (Without condensation) | |
| Vibration requirements | MIL-T-28800E, Class 2 |
| Drop text requirements | 1 m |
| IP Rating | IEC 60529: IP52 |
| Electromagnetic environment | IEC 61326-1, Portable |
| Safety | IEC 61010-1, Max 30 V to earth, Pollution Degree 2 |
| Power supply | 4 AA NEDA 1.5A IEC LR6 batteries |
| Size (H x W x L) | 52.5 x 84 x 188.5 mm |
| Weight | 515 g |
| DC mA Measurement | |
| Resolution | 0-24 mA |
| Range | 0.001 mA |
| Accuracy (% of reading + counts) | 0.010% + 2 μA |
| Temperature coefficient | ± (0.002% of reading + 0.002% of range) /°C ( < 18°C or >28°C) |
| Ohms Measurement | |
| Ohms range | Accuracy (% of reading +counts) |
| 0.00 Ω to 400.00 Ω | 0.015% + 0.05 Ω |
| 400.0 Ω to 4000.0 Ω | 0.015% + 0.5 Ω |
| Read accuracy is based on 4-wire input. For 3-wire ohm measurements, assuming all three leads are matched, add 0.05 Ω (0.00 Ω~400.00 Ω), 0.2 Ω (400.0 Ω~4000.0 Ω) to the specifications. | |
| Temperature coefficient | ±(0.002% of reading + 0.002% of range) / °C (< 18°C or > 28°C) |
| Ohms Source | ||
| Ohms range | 1.0 Ω to 400.0 Ω | |
| 1.00 Ω to 400.00 Ω | ||
| 400.0 Ω to 1500.0 Ω | ||
| 1500.0 Ω to 4000.0 Ω | ||
| Excitation current from measurement device | 0.1 mA to 0.5 mA | |
| 0.5 mA to 3 mA | ||
| 0.05 mA to 0.8 mA | ||
| 0.05 mA to 0.4 mA | ||
| Accuracy (% of reading + counts) | 0.015% + 0.1 Ω | |
| 0.015% + 0.05 Ω | ||
| 0.015% + 0.5 Ω | ||
| 0.015% + 0.5 Ω | ||
| Resolution | 0.00 Ω to 400.00 Ω | 0.01 Ω |
| 400.0 Ω to 4000.0 Ω | 0.1 Ω | |
| Temperature coefficient | ± (0.002% of reading + 0.002% of range) /°C ( <18°C or >28°C) Supports pulsed transmitters and PLCs with pulse times as short as 5 ms |
|
| RTD Input and Output | ||||||
| RTD Type (α) | Range (°C) | Measure (°C) | Source (°C) | |||
| 1-year | 2-year | Source current | 1-year | 2-year | ||
| 10 Ω Pt (385) | -200 to 100°C | 1.5°C | 3°C | 1 mA | 1.5°C | 3°C |
| 100 to 800°C | 1.8°C | 3.6°C | 1 mA | 1.8°C | 3.6°C | |
| 50 Ω Pt (385) | -200 to 100°C | 0.4°C | 0.7°C | 1 mA | 0.4°C | 0.7°C |
| 100 to 800°C | 0.5°C | 0.8°C | 1 mA | 0.5°C | 0.8°C | |
| 100 Ω Pt (385) | -200 to 100°C | 0.2°C | 0.4°C | 1 mA | 0.2°C | 0.4°C |
| 100 to 800°C | 0.015% +0.18°C |
0.03% +0.36°C |
1 mA | 0.015% +0.18°C |
0.03% +0.36°C |
|
| 200 Ω Pt (385) | -200 to 100°C | 0.2°C | 0.4°C | 500 μA | 0.2°C | 0.4°C |
| 100 to 630°C | 0.015% +0.18°C |
0.03% +0.36°C |
500 μA | 0.015% +0.18°C |
0.03% +0.36°C |
|
| 500 Ω Pt (385) | -200 to 100°C | 0.3°C | 0.6°C | 250 μA | 0.3°C | 0.6°C |
| 100 to 630°C | 0.015% +0.28°C |
0.03% +0.56°C |
250 μA | 0.015% +0.28°C |
0.03% +0.56°C |
|
| 1000 Ω Pt (385) | -200 to 100°C | 0.2°C | 0.4°C | 250 μA | 0.2°C | 0.4°C |
| 100 to 630°C | 0.015% +0.18°C |
0.03% +0.36°C |
250 μA | 0.015% +0.18°C |
0.03% +0.36°C |
|
| 100 Ω Pt (3916) | -200 to 100°C | 0.2°C | 0.4°C | 1 mA | 0.2°C | 0.4°C |
| 100 to 630°C | 0.015% +0.18°C |
0.03% +0.36°C |
1 mA | 0.015% +0.18°C |
0.03% +0.36°C |
|
| 100 Ω Pt (3926) | -200 to 100°C | 0.2°C | 0.4°C | 1 mA | 0.2°C | 0.4°C |
| 100 to 630°C | 0.015% +0.18°C |
0.03% +0.36°C |
1 mA | 0.015% +0.18°C |
0.03% +0.36°C |
|
| 10 Ω Cu (427) | -100 to 260°C | 1.5°C | 3°C | 1 mA | 1.5°C | 3°C |
| 120 Ω Ni (672) | -80 to 260°C | 0.15°C | 0.3°C | 1 mA | 0.15°C | 0.3°C |
| 50 Ω Cu (427) | -180 to 200°C | 0.4°C | 0.7°C | 1 mA | 0.4°C | 0.7°C |
| 100 Ω Cu (427) | -180 to 200°C | 0.2°C | 0.4°C | 1 mA | 0.2°C | 0.4°C |
| YSI400 | 15 to 50°C | 0.2°C | 0.4°C | 250 μA | 0.2°C | 0.4°C |
| 1. Sensor inaccuracies not included. 2. Resolution: 0.1°C. 3. Read accuracy is based on 4-wire input. For 3-wire RTD measurements, assuming all three RTD leads are matched, add 1.0°C (Pt10 and Cu10), 0.6°C (Pt50 and Cu50), 0.4°C (Other RTD types) to the specifications. 4. Source Accuracy in source mode is based on 0.5 mA~3 mA (1.00 Ω~400.00 Ω), 0.05 mA~0.8 mA (400.0 Ω~1500.0 Ω), 0.05 mA~0.4 mA (1500.0 Ω ~4000.0 Ω), excitation current (0.25 mA for Pt1000 range). 5. Temperature Coefficient: ±0.05°C /°C for measure, ±0.05°C /°C (< 18°C or > 28°C) for source. 6. Supports pulsed transmitters and PLCs with pulse times as short as 5 ms. |
||||||
Related Products
- Dual-zone control
- Full HART field communicator
- Process calibrator option provides a multi-channel readout for a reference thermometer, RTDs and TCs, task documentation, and HART Communication
- Set point control by reference
- Self-calibration feature
- ISO 17025-accredited calibration w/data included
- Temperature control from 100°C to 1210°C
- Two models to choose from: Reference (ADT878) and Standard (ADT875)
- Display Accuracy of ±1.5°C (ADT878)
- 4 on-board measurement channels (PC option)
- Process calibrator option provides a multi-channel readout for TCs, switches and transmitters, including task documentation and HART communication
- ISO 17025-accredited calibration w/data included
- Measure, graph, and record PRTs, thermocouples, and thermistors, supporting a wide range of temperature sensors.
- Accuracy up to ±0.002 °C.
- Quick sample readings in as little as 0.3 seconds.
- 1524 has two channels for simultaneous measurements, doubling productivity.
- Intrinsically-safe, battery operated, portable reference thermometer
- Provides accuracy of ±0.05°C (±0.09°F) over full range
- Features intrinsically-safe operation (ATEX and IECEx compliant)
- Measures -80 °C to 300 °C ( -112 °F to 572 °F)
- Displays temperature in °C or °F
- Measure thermocouples, PRTs, thermistors, DC voltage, DC current, and resistance, with up to 40 isolated universal inputs.
- Precision measurements (PRTs ± 0.005 °C, thermocouples: ± 0.5 °C, thermistors: ± 0.002 °C).
- Scan up to 10 channels per second, great for high-speed applications.
- Four operation modes – Scan, Monitor, Measure, Digital Multimeter (DMM), and features real-time color trending for up to four channels.
- Records up to 20 MB of data and setup files to internal non-volatile memory or an external USB drive, with administrator and user profiles for data protection and test traceability.
- Highly accurate thermo hygrometer and humidity monitor.
- Measure temperature to ± 0.125 °C and humidity to ± 1.5% on two channels.
- Offers networking capabilities via Ethernet or RS-232, connect multiple Fluke 1620s.
- Powerful logging and analysis tools, including the LogWare III software for real-time logging, data storage, and integration with MET/CAL® software.
- Accredited radiometric calibration included
- Accurate, reliable performance from –15 °C to 500 °C
- Large target size of 152 mm (6 in) is required for calibrating most thermometers.
- Light instrument weight of 8.6 kg (19 lbs) makes it easy to lift and carry.
- Operate in temperatures ranging from -200 °C to 160 °C.
- Accuracy of ± 0.05 °C.
- Both the transition junction and lead wires of these PRTs can withstand the full PRT temperature range.
- Unique probe seal prevents moisture ingress.
- Come with NVLAP-accredited calibration.
- 5608: –200 °C to 500 °C (80 mm minimum immersion)
- 5609: –200 °C to 670 °C (100 mm minimum immersion)
- ± 0.01 short-term stability
- Calibration not included, NVLAP-accredited calibration optional, lab code 200348-0
- Dry-well reference with 90 ° bend
- Highly accurate, easy-to-use thermocouple temperature calibrator
- Measures and simulates 17 thermocouple types and millivolts
- Measures 4 to 20 mA signals while sourcing a temperature signal
- Comes with a magnetic hanger for convenient hands-free operation
- Features configurable source settings for quick linearity checks
- Calibration of both temperature and pressure transmitters
- Measure/source/simulate mA loop parameters
- Measure and source mA, Volts DC, frequency, and resistance
- Built-in 24 V loop power supply
- Intrinsically safe multifunction process calibrator
- Meets I.S. Class I, Division 1 Groups B-D, 171°C compliance standards
- Measures volts DC, mA, RTDs, thermocouples, frequency, and ohms
- Sources or simulates volts DC, mA, RTDs, thermocouples, frequency and ohms
- Precision multifunction, temp, frequency, voltage, and mA calibrator
- Calibrates thermocouples and RTD temperature instruments
- Stores up to eight calibration results in memory for analysis
- Built-in 24 V Loop Power Supply
- Measures volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
- Sources and simulate volts, mA, pressure, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters
- Powers transmitters during test using loop supply with simultaneous mA measurement
- Complete pressure, temperature, and mA loop calibrator
- Troubleshoot and calibrate HART smart digital transmitters
- Create calibration procedures and automatically document results
- Exceptional stability (±0.005 °C) and axial uniformity (±0.02 °C over a 60 mm zone).
- Immersion depth to 203 mm (8 in)
- Optional ITS-90 reference input reads PRTs to ±0.006 °C
- Wide Temperature range from –45 °C to 700 °C
- Exceptional stability (±0.005 °C) and axial uniformity (±0.02 °C over a 60 mm zone).
- Immersion depth to 203 mm (8 in)
- Optional ITS-90 reference input reads PRTs to ±0.006 °C
- Wide Temperature range from –45 °C to 700 °C
- Exceptional stability (±0.005 °C) and axial uniformity (±0.02 °C over a 60 mm zone).
- Immersion depth to 203 mm (8 in)
- Optional ITS-90 reference input reads PRTs to ±0.006 °C
- Wide Temperature range from –45 °C to 700 °C
- Wide temperature range: –95 °C to 140 °C.
- Excellent accuracy
- Accuracy using built-in reference thermometer readout: ± 0.05 °C full range.
- Display accuracy: ± 0.2 °C full range.
- Best-in-class stability: ± 0.015 °C full range
- Wide Temperature Range: Covers a broad temperature range from -95°C to 140°C, suitable for a wide variety of applications, including those requiring low-temperature calibration.
- Portability: Remains compact and portable, allowing for on-site calibrations at various locations.