Fluke 6109A / 7109A Portable Calibration Baths

  • Calibrate up to four tri-clamp sanitary sensors per batch with ± 0.1 °C temperature display accuracy.
  • Covers a wide temperature range: 6109A: 35 °C to 250 °C
  • 7109A: -25 °C to 140 °C.
  • Stainless steel casing that withstands harsh cleaning solutions; it’s easy to transport and maintain.

Description

The Fluke Calibration 6109A and 7109A are high-performance Portable Calibration Baths designed for calibrating sanitary temperature sensors, particularly those used in the food, beverage, and pharmaceutical industries.

Key Features:

  • Calibrate up to four tri-clamp sanitary sensors per batch with ± 0.1 °C temperature display accuracy.
  • Covers a wide temperature range:
    • 6109A: 35 °C to 250 °C
    • 7109A: -25 °C to 140 °C.
  • Stainless steel casing that withstands harsh cleaning solutions; it’s easy to transport and maintain.
  • High Accuracy: Offer high accuracy and stability, ensuring reliable calibration results.
  • High Throughput: Can calibrate up to four tri-clamp sanitary sensors simultaneously, significantly increasing efficiency.
  • Sanitary Design: Constructed with stainless steel for easy cleaning and decontamination, meeting stringent hygiene requirements.
  • Portability: Compact and lightweight, making them suitable for on-site calibrations in production facilities.
  • User-Friendly:
    • Feature intuitive controls and a clear display for simple operation.

Applications:

  • Food and Beverage Industry: Calibration of temperature sensors used in food processing, storage, and packaging.
  • Pharmaceutical Industry: Calibration of temperature sensors used in pharmaceutical manufacturing, storage, and distribution.
  • Biotechnology: Calibration of temperature sensors used in bioreactors, fermentation tanks, and other bioprocessing equipment.
  • Chemical Processing: Calibration of temperature sensors used in chemical processes where hygiene and sanitation are critical.

The Fluke Calibration 6109A and 7109A are specialized calibration baths designed for the unique needs of the food, beverage, and pharmaceutical industries. Their high throughput, sanitary design, and portability make them valuable tools for ensuring accurate temperature measurements and maintaining product quality and safety.

Additional information

Brand

Fluke Calibration

Specifications: Fluke Calibration 6109A / 7109A Portable Calibration Baths

General specifications
Size Height: 382 mm (15 in)
Width: 242 mm (9.5 in)
Depth: 400 mm (15.7 in)
Weight 6109A: 16 kg (35 lb)
7109A: 20 kg (44 lb)
Fluid volume 2.5 liters (0.66 gallons)
Fluid working area 75 mm x 75 mm (3 in x 3 in)
Maximum fluid depth 154 mm (6.1 in)
Factory calibration Traceable NVLAP-accredited calibration included

 

Temperature specifications
The temperature specifications describe the Absolute Instrumental Uncertainty at 95 % level of confidence (coverage factor k = 2) within one year from calibration. The specifications include environmental temperature effects from 13 °C to 33 °C.
6109A 7109A
Range (at 25 °C ambient temperature) 35 °C to 250 °C –25 °C to 140 °C (–15 °C cover off)
Accuracy (maximum permissible error) 0.1 °C 0.1 °C
Display resolution 0.01 °, 0.001° 0.01 °, 0.001°
Stability 0.015 °C 0.01 °C
Typical uniformity 0.03 °C up to 200 °C
0.04 °C above 200 °C
0.02 °C
Repeatability 0.04 °C 0.04 °C
Typical heating time 35 °C to 100 °C: 25 minutes
100 °C to 250 °C: 45 minutes
–25 °C to 25 °C: 35 minutes
25 °C to 140 °C: 55 minutes
Typical cooling time 250 °C to 100 °C: 35 minutes
100 °C to 35 °C: 55 minutes
140 °C to 25 °C: 45 minutes
25 °C to –25 °C: 75 minutes
Typical settling time 15 minutes 10 minutes

 

Process Input Module specifications (-P models)
The Input Module specifications describe the Absolute Instrumental Uncertainty at 95 % level of confidence (coverage factor k = 2) within one year from calibration. The specifications include calibration uncertainty, linearity, repeatability, resolution, stability, and environmental temperature effects from 13 °C to 33 °C.
Function Range Accuracy (Maximum Permissible Error)
Reference input resistance 0 Ω to 42 Ω
42 Ω to 400 Ω
0.0025 Ω
0.006 %
Reference input temperature -25 °C to 250 °C 0.007 % + 0.015 °C
Resistance sensing current 1 mA 8 %
DUT 4-wire resistance 0 Ω to 31 Ω
31 Ω to 400 Ω
0.0025 Ω
0.008%
DUT 3-wire resistance accuracy 0 Ω to 400 Ω 0.12 Ω
Thermocouple mV -10 mV to 100 mV 0.025 % + 0.01 mV
Reference junction temperature 0 °C to 40 °C 0.35 °C
Thermocouple temperature -25 °C to 250 °C J: 0.44 °C
E: 0.44 °C
L: 0.42 °C
K: 0.49 °C
N: 0.57 °C
U: 0.48 °C
T: 0.53 °C
M: 0.48 °C
0 °C to 250 °C R: 1.92 °C
D: 1.12 °C
S: 1.88 °C
G/W: 3.34 °C
C: 0.84 °C
mA range 0 mA to 22 mA 0.02 % + 0.002 mA
Loop power 24 V dc ± 6 V

 

Related Products