- Multi-purpose AC measurement and transfer standard: accuracy of a thermal transfer standard with the ease of use of a digital multimeter.
- Absolute AC voltage measurement uncertainties as low as ± 24 ppm (one year, 23 °C ± 5 °C).
- Alternating voltage range of 600 μV to 1000 V, and a frequency range of 10 Hz to 1 MHz, with options to extend to 30 or 50 MHz.
- Combine with A40B Current Shunts for direct, absolute, or relative current measurements without the need for calculations.
- 6.5-inch, full-color, capacitive touch screen display.
- Use alone or as a transfer standard when compared to external DC source, performing all necessary switching and calculations automatically.
- Description
- Additional information
- Specifications
Description
The Fluke 5790B is a high-precision AC Measurement Standard designed for demanding calibration applications. Here’s an overview:
Key Features:
- Multi-purpose AC measurement and transfer standard: accuracy of a thermal transfer standard with the ease of use of a digital multimeter.
- Absolute AC voltage measurement uncertainties as low as ± 24 ppm (one year, 23 °C ± 5 °C).
- Alternating voltage range of 600 μV to 1000 V, and a frequency range of 10 Hz to 1 MHz, with options to extend to 30 or 50 MHz.
- Combine with A40B Current Shunts for direct, absolute, or relative current measurements without the need for calculations.
- 6.5-inch, full-color, capacitive touch screen display.
- Use alone or as a transfer standard when compared to external DC source, performing all necessary switching and calculations automatically.
- Exceptional Accuracy: Delivers highly accurate AC voltage measurements with uncertainties as low as ± 24 ppm (parts per million).
- Solid-State Thermal RMS Sensor: Employs a patented Fluke Solid-State Thermal RMS Sensor for true RMS measurements, ensuring accurate readings across various waveforms.
- Wide Frequency Range: Measures AC voltages across a broad frequency range, typically from 10 Hz to 1 MHz (with options for extended frequency ranges).
- Versatile Applications: Suitable for a wide range of applications, including calibrating:
- AC voltmeters
- Oscilloscopes
- Power analyzers
- Data acquisition systems
- User-Friendly Interface: Features a user-friendly interface for easy operation and data analysis.
Benefits:
- Enhanced Accuracy: Ensures highly accurate and reliable AC voltage measurements.
- Improved Efficiency: Streamlines the calibration process with its ease of use and automation capabilities.
- Increased Confidence: Provides confidence in the accuracy of calibrated equipment.
- Versatility: Adaptable to various calibration needs across different industries.
In essence, the Fluke 5790B is a precision instrument that plays a critical role in maintaining the accuracy and reliability of AC voltage measurements in various applications.
Additional information
Brand | Fluke Calibration |
---|
Specifications: 5790B AC Measurement Standard
Summary Specifications | |
AC and dc voltage | Best 99%, absolute specification: ± 24 ppm (one year, 23 °C to ± 5 °C |
Inputs | Input 1: 600 µV to 1000 V (5% over-range) Input 2: 600 µV to 1000 V (5% over-range) Aux input: 250 mV to 500 mV WB input: <7 V rms wideband |
Warm-up time | 30 minutes |
Relative humidity | Operating: 45% to 50 °C 75% to 45 °C 95% to 30 °C Storage: <95% non-condensing |
Altitude | Operating: 3,050 meters (10,000 feet) Non-operating: 12,200 meters (40,000 feet) |
Temperature | Operating: 0 °C to °50 C Calibration: 15 °C to 35 °C Storage: -40 °C to 70 °C |
Electromagnetic Compatibility (EMC) | |
International | IEC 61326-1: Controlled Electromagnetic Environment – CISPR 11: Group 1, Class A |
Korea (KCC) | Class A Equipment (Industrial Broadcasting & Communication Equipment) |
USA (FCC) | 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103. |
Surge | ANSI C62.41-1980, Category A |
Reliability | MIL-T-2880D, paragraph 3.13.3 |
General Specifications | |
Dimensions | Height: 17.8 cm (7 in) standard rackmount + 1.5 cm (0.6 in) Width: 43.2 cm (17 in) Depth: 62 cm (24.8 in) |
Maximum power requirements | 5790B: 100 VA |
Weight | 5790B: 24 kg (53 lb) With wideband: 24.5 kg (54 lb) |
Line power | 50 Hz or 60 Hz; 100 V – 120 V, 220 V – 240 V 100V, 110 V, 115 V, 120 V, 200 V, 220 V, 230 V, 240 V |
Safety | Complies with UL1244 and IEC 348-1976 and IEC 1010 and CSA C22.2 No. 231 and ANSI/ISA S82 |
Remote interfaces | RS-232, IEEE-488 |
Confidence level | 99% unless otherwise specified |
5790B/AF | |
The 5790B/AF has enhanced absolute accuracy to ± 0.23 % of voltage (1 year after factory calibration, 23 °C ± 3 °C, at 95 % confidence level) at 50 MHz and 223.61 mV, using a serialized one meter cable, characterized at that point for RF power meter 0 dBm reference output measurement. In the RF power measurement mode, the accuracy is specified to the end of the serialized cable with a curve fit for the other ranges and frequencies that is a first order correction for cable loss. |