Fluke 5606 Full Immersion PRT

  • Operate in temperatures ranging from -200 °C to 160 °C.
  • Accuracy of ± 0.05 °C.
  • Both the transition junction and lead wires of these PRTs can withstand the full PRT temperature range.
  • Unique probe seal prevents moisture ingress.
  • Come with NVLAP-accredited calibration.
  • Suitable for calibration/validation of sensors in laboratory or bio freezers, autoclaves, ovens, test chambers, IQ/OQ/PQ qualification procedures, temperature mapping, and data logging in temperature-controlled spaces, and more.

Description

The Fluke 5606 is a Full Immersion Platinum Resistance Thermometer (PRT) designed for use in demanding temperature environments.

Key Features:

  • Operate in temperatures ranging from -200 °C to 160 °C.
  • Accuracy of ± 0.05 °C.
  • Both the transition junction and lead wires of these PRTs can withstand the full PRT temperature range.
  • Unique probe seal prevents moisture ingress.
  • Come with NVLAP-accredited calibration.
  • Suitable for calibration/validation of sensors in laboratory or bio freezers, autoclaves, ovens, test chambers, IQ/OQ/PQ qualification procedures, temperature mapping, and data logging
  • in temperature-controlled spaces, and more.
  • Wide Temperature Range: Operates from -200°C to 160°C, making it suitable for a wide range of applications.
  • Full Immersion Capability: The entire probe, including the transition junction and lead wires, can be fully immersed in the medium being measured, ensuring accurate temperature readings.
  • High Accuracy: Provides accurate temperature measurements with a specified calibration accuracy.
  • Robust Construction: Designed to withstand harsh environments and provide reliable performance.

Here are some key application areas:

  • Laboratory Freezers and Refrigerators:
    • Calibration and validation of temperature sensors in laboratory freezers, refrigerators, and cold rooms.
    • Ensuring proper temperature control for sensitive samples and reagents.
  • Autoclaves and Sterilizers:
    • Monitoring and verifying sterilization temperatures in autoclaves and other sterilization equipment.
    • Ensuring the effectiveness of sterilization processes.
  • Ovens and Furnaces:
    • Calibrating and validating temperature sensors used in ovens, furnaces, and other high-temperature equipment.
    • Ensuring accurate temperature control in industrial processes.
  • Stability Test Chambers:
    • Monitoring and controlling temperature within stability test chambers used in pharmaceutical and other industries.
    • Ensuring product stability and shelf life.

The Fluke 5606 is a high-quality PRT that offers exceptional performance in demanding temperature environments. Its full immersion capability, wide temperature range, and robust construction make it a reliable choice for accurate temperature measurements in various applications.

Additional information

Brand

Fluke Calibration

Specifications: 5606 Full Immersion PRT

General Specifications
Temperature range -200°C to 160°C
Nominal resistance at 0.01°C 100 Ω ±0.1 Ω
Temperature coefficient 0.00385 Ω/Ω/°C
Accuracy1 ±0.04°C at 0.01°C
±0.06°C at 160°C
Short-term repeatability2 ±0.03°C at 0.01°C
±0.04°C at 160°C
Drift3 ±0.03°C at 0.01°C
±0.04°C at 160°C
Hysteresis ±0.015°C
Sheath length 50 mm ±5 mm
(2 in ±0.2 in)
Sheath diameter 3.1 mm ±0.1 mm (1/8 in ±0.004 in)
Sheath material 316 SST
Transition junction temperature range4 -200°C to 160°C
Transition junction dimensions No transition
Sensor length 30 mm ±3mm (1.2 in ±0.1 in)
Sensor location 3 mm ±1 mm from tip (1.2 in ±0.04 in)
Minimum insulation resistance 20 MΩ at 23°C
Minimum immersion length Full immersion
Maximum immersion depth in liquid medium Full immersion
Maximum immersion depth in dry Medium Full immersion
Response time5 12 seconds typical
Self heating (in 0°C bath) ±0.003°C
Lead-wire cable type Enameled copper wire
Lead-wire length 2.4 meters (8 ft)
Lead-wire temperature range 160°C
Calibration Calibration not included. NVLAP accredited calibration available. See ordering information.
1. Includes calibration and 100 hr drift (k = 2).
2. Three thermal cycles from min to max temp, includes hysteresis, 95% confidence (k = 2).
3. After 100 hours at max temp, 95% confidence (k = 2).
4. Temperatures outside this range will cause irreparable damage.
5. Per ASTM E 644.
Calibration Uncertainty
Temperature -197°C 1924 Calibration Uncertainty6 0.03°C
Tolerance w/out Calibration7 0.4°C
-38°C 1924 Calibration Uncertainty6 0.03°C
Tolerance w/out Calibration7 0.2°C
0°C 1924 Calibration Uncertainty6 0.03°C
Tolerance w/out Calibration7 0.1°C
100°C 1924 Calibration Uncertainty6 0.045°C
Tolerance w/out Calibration7 0.3°C
157°C 1924 Calibration Uncertainty6 0.05°C
Tolerance w/out Calibration7 0.4°C
6. Lab code 200348-0
7. Each PRT includes a certificate with a measured R0 value.
An uncalibrated 5606 meets ASTM E 1137 Grade A Classification of Tolerances.
Use the coefficients below to achieve the tolerances listed in the “No Calibration” column or calculate the Grade A tolerance using the following formula:
Grade A = ±[0.13 + 0.0017|t|]°C
where: |t| = value of temperature (°C) without regard to sign. See ASTM E 1137.
R0: Enter the R0 value provided
A: 3.9083 x 10–3
B: –5.775 x 10–7
C: –4.183 x 10–12

 