Fluke 5606 Full Immersion PRT
- Operate in temperatures ranging from -200 °C to 160 °C.
- Accuracy of ± 0.05 °C.
- Both the transition junction and lead wires of these PRTs can withstand the full PRT temperature range.
- Unique probe seal prevents moisture ingress.
- Come with NVLAP-accredited calibration.
- Suitable for calibration/validation of sensors in laboratory or bio freezers, autoclaves, ovens, test chambers, IQ/OQ/PQ qualification procedures, temperature mapping, and data logging in temperature-controlled spaces, and more.
- Description
- Additional information
- Specifications
Description
The Fluke 5606 is a Full Immersion Platinum Resistance Thermometer (PRT) designed for use in demanding temperature environments.
Key Features:
- Operate in temperatures ranging from -200 °C to 160 °C.
- Accuracy of ± 0.05 °C.
- Both the transition junction and lead wires of these PRTs can withstand the full PRT temperature range.
- Unique probe seal prevents moisture ingress.
- Come with NVLAP-accredited calibration.
- Suitable for calibration/validation of sensors in laboratory or bio freezers, autoclaves, ovens, test chambers, IQ/OQ/PQ qualification procedures, temperature mapping, and data logging
- in temperature-controlled spaces, and more.
- Wide Temperature Range: Operates from -200°C to 160°C, making it suitable for a wide range of applications.
- Full Immersion Capability: The entire probe, including the transition junction and lead wires, can be fully immersed in the medium being measured, ensuring accurate temperature readings.
- High Accuracy: Provides accurate temperature measurements with a specified calibration accuracy.
- Robust Construction: Designed to withstand harsh environments and provide reliable performance.
Here are some key application areas:
- Laboratory Freezers and Refrigerators:
- Calibration and validation of temperature sensors in laboratory freezers, refrigerators, and cold rooms.
- Ensuring proper temperature control for sensitive samples and reagents.
- Autoclaves and Sterilizers:
- Monitoring and verifying sterilization temperatures in autoclaves and other sterilization equipment.
- Ensuring the effectiveness of sterilization processes.
- Ovens and Furnaces:
- Calibrating and validating temperature sensors used in ovens, furnaces, and other high-temperature equipment.
- Ensuring accurate temperature control in industrial processes.
- Stability Test Chambers:
- Monitoring and controlling temperature within stability test chambers used in pharmaceutical and other industries.
- Ensuring product stability and shelf life.
The Fluke 5606 is a high-quality PRT that offers exceptional performance in demanding temperature environments. Its full immersion capability, wide temperature range, and robust construction make it a reliable choice for accurate temperature measurements in various applications.
Additional information
Brand | Fluke Calibration |
---|
Specifications: 5606 Full Immersion PRT
General Specifications | ||
Temperature range | -200°C to 160°C | |
Nominal resistance at 0.01°C | 100 Ω ±0.1 Ω | |
Temperature coefficient | 0.00385 Ω/Ω/°C | |
Accuracy1 | ±0.04°C at 0.01°C ±0.06°C at 160°C |
|
Short-term repeatability2 | ±0.03°C at 0.01°C ±0.04°C at 160°C |
|
Drift3 | ±0.03°C at 0.01°C ±0.04°C at 160°C |
|
Hysteresis | ±0.015°C | |
Sheath length | 50 mm ±5 mm (2 in ±0.2 in) |
|
Sheath diameter | 3.1 mm ±0.1 mm (1/8 in ±0.004 in) | |
Sheath material | 316 SST | |
Transition junction temperature range4 | -200°C to 160°C | |
Transition junction dimensions | No transition | |
Sensor length | 30 mm ±3mm (1.2 in ±0.1 in) | |
Sensor location | 3 mm ±1 mm from tip (1.2 in ±0.04 in) | |
Minimum insulation resistance | 20 MΩ at 23°C | |
Minimum immersion length | Full immersion | |
Maximum immersion depth in liquid medium | Full immersion | |
Maximum immersion depth in dry Medium | Full immersion | |
Response time5 | 12 seconds typical | |
Self heating (in 0°C bath) | ±0.003°C | |
Lead-wire cable type | Enameled copper wire | |
Lead-wire length | 2.4 meters (8 ft) | |
Lead-wire temperature range | 160°C | |
Calibration | Calibration not included. NVLAP accredited calibration available. See ordering information. | |
1. Includes calibration and 100 hr drift (k = 2). 2. Three thermal cycles from min to max temp, includes hysteresis, 95% confidence (k = 2). 3. After 100 hours at max temp, 95% confidence (k = 2). 4. Temperatures outside this range will cause irreparable damage. 5. Per ASTM E 644. |
||
Calibration Uncertainty | ||
Temperature | -197°C | 1924 Calibration Uncertainty6 0.03°C Tolerance w/out Calibration7 0.4°C |
-38°C | 1924 Calibration Uncertainty6 0.03°C Tolerance w/out Calibration7 0.2°C |
|
0°C | 1924 Calibration Uncertainty6 0.03°C Tolerance w/out Calibration7 0.1°C |
|
100°C | 1924 Calibration Uncertainty6 0.045°C Tolerance w/out Calibration7 0.3°C |
|
157°C | 1924 Calibration Uncertainty6 0.05°C Tolerance w/out Calibration7 0.4°C |
|
6. Lab code 200348-0 7. Each PRT includes a certificate with a measured R0 value. An uncalibrated 5606 meets ASTM E 1137 Grade A Classification of Tolerances. Use the coefficients below to achieve the tolerances listed in the “No Calibration” column or calculate the Grade A tolerance using the following formula: Grade A = ±[0.13 + 0.0017|t|]°C where: |t| = value of temperature (°C) without regard to sign. See ASTM E 1137. R0: Enter the R0 value provided A: 3.9083 x 10–3 B: –5.775 x 10–7 C: –4.183 x 10–12 |