Description
The Fluke 4180 and 4181 are precision infrared calibrators designed for accurate calibration of infrared thermometers and thermal imagers. They offer a combination of accuracy, ease of use, and portability, making them ideal for both field and laboratory applications.
Key Features:
- High Accuracy: Both models provide high accuracy, ensuring reliable calibration results.
- Wide Temperature Range: The 4180 covers a temperature range from -15°C to 120°C, while the 4181 covers a wider range from 35°C to 500°C.
- Portability: Compact and lightweight, making them easy to transport and use in various locations.
- Ease of Use: Intuitive interface and simple operation make calibration quick and efficient.
- Accredited Calibration: Comes with an accredited calibration from a trusted laboratory, ensuring traceability and confidence in measurement results.
Applications:
- Calibration of Infrared Thermometers: Used to calibrate a wide range of infrared thermometers, from general-purpose models to high-precision instruments.
- Thermal Imaging: Calibrating thermal imagers for accurate temperature measurements and analysis.
- Industrial Maintenance: Used for preventive maintenance and troubleshooting of equipment.
- Research and Development: Accurate temperature measurements in research and development laboratories.
- Quality Control: Ensuring product quality and consistency through accurate temperature measurements.
Overall, the Fluke 4180 and 4181 are valuable tools for anyone who needs to ensure the accuracy of their infrared temperature measurements.
Additional information
Brand | Fluke Calibration |
---|
Specifications: Fluke Calibration 4180/4181 Precision IR Calibrators
Specifications | 4180 | 4181 |
Temperature range (@ 23°C ambient, 0.95 emissivity) | –15°C to 120°C | 35°C to 500°C |
Display accuracy1 | ±0.40°C at –15°C ±0.40°C at 0°C ±0.50°C at 50°C ±0.50°C at 100°C ±0.55°C at 120°C |
±0.35°C at 35°C ±0.50°C at 100°C ±0.70°C at 200°C ±1.20°C at 350°C ±1.60°C at 500°C |
Stability | ±0.10°C at –15°C ±0.05°C at 0°C ±0.10°C at 120°C |
±0.05°C at 35°C ±0.20°C at 200°C ±0.40°C at 500°C |
Uniformity (5.0 in dia of center of target)2 | ±0.15°C at –15°C ±0.10°C at 0°C ±0.25°C at 120°C |
±0.10°C at 35°C ±0.50°C at 250°C ±1.00°C at 500°C |
Uniformity (2.0 in dia of center of target)2 | ±0.10°C at -15°C ±0.10°C at 0°C ±0.20°C at 120°C |
±0.10°C at 35°C ±0.25°C at 250°C ±0.50°C at 500°C |
Heating time | 15 min: –15°C to 120°C 14 min: 23°C to 120°C |
45 min: 35⁰ C to 500⁰ C |
Cooling time | 15 min: 120°C to 23°C 20 min: 23°C to –15°C |
100 min: 500°C to 35°C 40 min: 500°C to 100°C |
Stabilization time | 10 minutes | 10 minutes |
Nominal emissivity3 | 0.95 | 0.95 |
Thermometer emissivity compensation | 0.9 to 1.0 | 0.9 to 1.0 |
Target diameter | 152.4 mm (6 in) | 152.4 mm (6 in) |
Computer interface | RS-232 | RS-232 |
Power | 115 V AC (±10%), 6.3 A, 50/60 Hz, 630 W 230 V AC (±10%), 3.15 A, 50/60 Hz, 630 W |
115 V AC (±10%), 10 A, 50/60 Hz, 1000 W 230 V AC (±10%), 5 A, 50/60 Hz, 1000 W |
Fuse(s) | 115 V AC 6.3 A, 250 V, slow blow 230 V AC 3.15 A, 250 V, T |
115 V AC 10 A, 250 V, fast blow 230 V AC 5 A, 250 V, F |
Size (H x W x D) | 356 x 241 x 216 mm (14 x 9.5 x 8.5 in) |
356 x 241 x 216 mm (14 x 9.5 x 8.5 in) |
Weight | 9.1 kg (20 lb) | 9.5 kg (21 lb) |
1. For 8 um to 14 um spectral band thermometers with emissivity set between 0.9 and 1.0 2. The uniformity specification refers to how IR thermometers with different spot sizes both focused at the center of the targer will measure the same temperature. 3. The target has a nominal emissivity of 0.95, however it is radiometrically calibrated to minimize emissivity related uncertainties. |