Fluke 1551a Stik Thermometer & Temperature Calibrator

  • Intrinsically-safe, battery operated, portable reference thermometer
  • Provides accuracy of ±0.05°C (±0.09°F) over full range
  • Features intrinsically-safe operation (ATEX and IECEx compliant)
  • Measures -50 °C to 160 °C (-58 °F to 320 °F)
  • Provides user-configurable temperature with Trend/Stability indicator
  • Displays temperature in °C or °F

Description

The Fluke 1551A is a high-precision, intrinsically safe “Stik” thermometer designed for accurate temperature measurements in various industrial and field applications.

Key features and benefits:

  • Intrinsically-safe, battery operated, portable reference thermometer
  • Provides accuracy of ±0.05°C (±0.09°F) over full range
  • Features intrinsically-safe operation (ATEX and IECEx compliant)
  • Measures -50 °C to 160 °C (-58 °F to 320 °F)
  • Provides user-configurable temperature with Trend/Stability indicator
  • Displays temperature in °C or °F
  • High accuracy: Delivers repeatable temperature measurements accurate to ±0.05°C over its full range.
  • Intrinsically safe: Rated intrinsically safe for use in potentially explosive atmospheres, ensuring safe operation in hazardous environments.
  • Portability and durability: Compact and rugged design makes it easy to transport and use in the field.
  • Long battery life: Provides up to 300 hours of continuous operation on a single battery charge.
  • Easy to use: Simple interface and intuitive operation make it easy to learn and use.

Applications:

  • Calibration:
    • Temperature Sensors: Calibrating a wide range of temperature sensors, including thermocouples, RTDs, and thermistors.
    • Temperature Transmitters: Verifying the accuracy of temperature transmitters used in process control systems.
  • Process Control:
    • Monitoring Critical Processes: Monitoring temperatures in critical processes such as chemical reactions, food processing, and pharmaceutical manufacturing.
    • Ensuring Product Quality: Maintaining accurate temperature control to ensure product quality and consistency.
  • Industrial Maintenance:
    • Troubleshooting Equipment: Diagnosing equipment malfunctions related to temperature deviations.
    • Preventive Maintenance: Conducting regular temperature checks to prevent equipment failures.
  • Hazardous Area Applications:
    • Oil and Gas: Measuring temperatures in potentially explosive atmospheres, such as oil refineries and gas processing plants.
    • Chemical Processing: Monitoring temperatures in hazardous environments where flammable gases or vapors may be present.

Overall, the Fluke 1551A is a reliable and accurate temperature measurement tool suitable for a wide range of applications, especially in hazardous environments.

Additional information

Brand

Fluke Calibration

Specifications: Fluke 1551a Stik Thermometer & Temperature Calibrator

Specifications
Temperature range -50°C to 160°C
(-58°F to 320°F)
Accuracy (1 year) ±0.05°C (±0.09°F)
Temperature resolution Selectable: 0.1, 0.01, 0.001 (default 0.01)
Sample rate Selectable: 0.5 seconds, 1 second, 2 seconds (default 1 second)
Readout temperature coefficient Add ±10 ppm/°C of full scale temperature from −10 °C to 18 °C and 28 °C to 50 °C
Probe temperature coefficient 0.00385 Ω/Ω/°C nominal
Nominal probe resistance at 0 °C 100 Ω
Probe hysteresis ±0.01°C
Response time 4.8 mm diameter probe (3/16 in): 14 seconds
6.35 mm diameter probe (1/4 in): 21 seconds
Operating temperature range readout -10°C to 50°C (14°F to 122°F)
Storage temperature -20°C to 60°C (-4°F to 140°F) 0% to 95% RH, non-condensing
Display units °C, °F
Sensor type 100 Ω thin-film RTD
Sensor length ≤ 10 mm
Sensor position (from sheath tip) 3 mm (0.1 in)
Minimum immersion depth Wet medium (liquid bath, thermowell with thermal compound) 70 mm (2.8 in)
Dry medium (dry-well, thermowell without thermal compound) 120 mm (4.8 in)
Maximum immersion depth 50 mm (2 in) less than the total length. Make sure the last 50 mm of probe, nearest to the internal electronics of the device, are not put into the heat source.
Probe sheath material Stainless steel
Optional data logging1 Up to 10,000 time-stamped readings stored to internal memory
Logging intervals1 2 seconds, 5 seconds, 10 seconds, 30 seconds, or 60 seconds
Damping Moving average of the most recent 2 to 10 readings
(ON/OFF, 2, 5, 10)
Communications RS-232 stereo jack (access calibration parameters only)
AC power None
DC power 3-AAA alkaline batteries
See Users Manual for approved batteries
Battery life Approximately 300 hours (12.5 days) of continuous operation with the backlight off.
Battery save (auto-off) range Selectable from 1 to 30 minutes or can be disabled
Safety IEC 60079-0, IEC 60079-11: Ex ib IIB T4 Gb; IEC 61010-1: Pollution Degree 2
Electromagnetic compatibility International:
IEC 61326-1: Portable Equipment;
IEC 61326-2-2; CISPR 11: Group 1, Class B
Group 1: Equipment has intentionally generated and/or uses conductively coupled radio frequency energy that is necessary for the internal function of the equipment itself.
Class B: Equipment is suitable for use in domestic establishments and in establishments directly connected to a low voltage power supply network which supplies buildings used for domestic purposes.
USA (FCC):
47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Size (readout only) 125 mm x 62 mm x 35 mm (5 in x 2.5 in x 1.4 in)
Probe Size 1551A-9: 4.8 mm x 229 mm (3/16 in x 9 in)
1551A-12: 6.35 mm x 305 mm (1/4 in x 12 in)
1551A-20: 6.35 mm x 508 mm (1/4 in x 20 in)
Weight 196 g (6.9 oz)
Calibration (included) NVLAP-accredited, NIST-traceable
Characterization CVD
1. See ordering information for optional data logging configurations

 

Related Products