Additel 850-1200 Laboratory Thermocouple Calibration Furnace
- Temperature control from 300°C to 1200°C
- 3-in-1 furnace with 9 unique modes
- Stability of ±0.1°C
- Radial uniformity of ±0.25°C @ 1200°C
- Axial uniformity of ±0.25°C @ 1200°C
- Multi-zone temperature control
- Quick cool technology
- Sliding probe holder provides mechanical stability and precise probe depth control
- Pivoting color touchscreen display
- Alumina and metal inserts available
- Patent pending EMF shielding technology
- Advanced safety control
- Wi-Fi Communications
- Description
- Additional information
- Specifications
Description
The Additel 850-1200 is a versatile and high-performance thermocouple calibration furnace designed to meet a wide range of calibration needs. It offers exceptional temperature control, stability, and uniformity, making it an ideal choice for calibration laboratories and industries requiring accurate temperature measurements.
Key Features:
- Wide Temperature Range: Operates from 300°C to 1200°C, covering a broad spectrum of temperature calibration requirements.
- 3-in-1 Furnace: Offers nine unique modes, including long cup, short cup, and multi-hole configurations, providing flexibility for various thermocouple types and lengths.
- Exceptional Stability: Maintains a stability of ±0.1°C, ensuring precise temperature control.
- Excellent Uniformity: Delivers radial and axial uniformity of ±0.25°C at 1200°C, ensuring accurate measurements across the furnace chamber.
- Multi-Zone Temperature Control: Advanced temperature control system for precise temperature regulation.
- Quick Cool Technology: Reduces cool-down time, increasing efficiency.
- Sliding Probe Holder: Provides mechanical stability and precise probe depth control.
- User-Friendly Interface: Intuitive touch screen display for easy operation.
- Advanced Safety Features: Built-in safety features to protect the user and the instrument.
- Wi-Fi Connectivity: Enables remote control and data transfer.
Applications:
- Calibration of Thermocouples: Calibrates a wide range of thermocouple types, including Type K, Type J, Type T, and more.
- Process Calibration: Verifies the accuracy of temperature measurement devices in process control systems.
- Research and Development: Supports research and development activities requiring precise temperature control.
- Quality Assurance: Ensures the accuracy of temperature measurements in quality control processes.
The Additel 850-1200 is a reliable and efficient tool for thermocouple calibration. Its combination of versatility, accuracy, and user-friendly features makes it a valuable asset for any calibration laboratory or field service technician.
Additional information
Brand | Additel |
---|
General Specifications
|
Cup Style Furnace Insert (Long version – Metal)
Specification | ADT850 |
Temperature Range | 300°C to 1200°C |
Heating Time | (23°C~1200°C) 40 mins, (empty well) |
Cooling Time | (1200°C~300°C) 90 mins, (empty well) |
Operating Conditions | 0°C to 50°C, 0-90%RH  (0°C~50°C), non-condensing, <2000  m altitude |
Storage Temperature | -20°C to 70°C |
Display Screen | 7 in (178 mm) color touch screen |
Display Resolution | 0.01°C |
Display Accuracy
(Long empty chamber mode) |
±5°C |
Performance Specifications
Long (Deep) Immersion | |||
Mode | Long empty chamber mode | Long cup mode / Long insert mode | Long alumina tube mode |
Application | Noble and base metal TC calibration | Base metal TC calibration | Noble metal TC calibration |
Configuration (insert) | Empty chamber, without insert | Long cup insert or multi-hole insert | 20 mm (ID) alumina  tube |
Insert Dimension |
N/A |
Cup insert: 36.5 X 28.5 X 80 mm
Block insert: 36.5 X 80 mm |
26 mm (OD) X 20 mm (ID) X 630 mm (L) |
Immersion Depth | 310 to 370 mm (geometrical center: 340 mm) | 370 mm to the bottom of insert | 310 to 370 mm(geometrical center: 340 mm) |
Stability | ±0.1°C full range | ±0.1°C full range | ±0.1°C full range |
Axial Uniformity |
±0.2°C full range
(within ±30 mm axial length from geometrical center) |
±0.2°C full range (within 60 mm from bottom of the insert) |
±0.2°C full range (within ±30 mm axial length from geometrical center) |
Radial Uniformity |
±0.2°C @ 300°C
±0.2°C @ 700°C ±0.2°C @ 1200°C (within 14 mm from geometrical center) |
±0.1°C @ 300°C
±0.15°C  @ 700°C ±0.2°C @ 1200°C (within 14 mm from geometrical center) |
N/A |