Additel ADT762W Automated Hydraulic Pressure Calibrator

  • Automated pressure generation and control to 15,000 psi (1035 Bar)
  • Accuracy to 0.01%FS
  • Dual-range with a manual or auto-selected mode
  • Designed for use with deionized water
  • Control stability ≤ 0.005%FS
  • Portable – designed for use in the field and in the lab
  • Supports two external pressure modules
  • Wi-Fi, LAN, Bluetooth, USB and Ethernet communication
  • Full HART field communicator
  • HART and profibus communication
  • Data logging and task management
  • Patented electric pump technology
  • ISO 17025-accredited calibration w/data included

Description

The Additel ADT762W is a portable and automated hydraulic pressure calibrator designed to simplify and streamline pressure calibration processes. This innovative device offers a wide range of features and benefits, making it an essential tool for professionals in various industries.

Key Features:

  • High-Pressure Capability: Handles pressure ranges up to 15,000 psi (1035 Bar), making it suitable for a wide range of applications.
  • High Accuracy: Delivers exceptional accuracy up to 0.01% FS, ensuring precise calibration results.
  • Dual-Range Capability: Offers both manual and auto-selected dual-range operation for flexibility and efficiency.
  • Water-Based Operation: Uses environmentally friendly deionized water as the working medium, eliminating the need for hazardous fluids.
  • Built-in Auto-Purge System: Automatically removes air from the system, ensuring accurate and efficient operation.
  • Wireless Connectivity: Wi-Fi, Bluetooth, USB, and Ethernet connectivity options enable remote control and data transfer.
  • Full HART Field Communicator: Supports communication with HART-enabled devices for seamless integration.
  • Data Logging and Task Management: Records calibration data and supports automated test sequences for efficient workflow.
  • Compact and Portable Design: Lightweight and portable design for easy transportation and use in various field and laboratory settings.

Applications:

  • Calibration Laboratories: Calibrating a wide range of pressure instruments, including transmitters, gauges, and switches.
  • Manufacturing Industries: Testing and calibrating pressure sensors and equipment in production processes.
  • Oil and Gas Industry: Calibrating pressure instruments used in oil and gas exploration and production.
  • Aerospace and Automotive Industries: Calibrating pressure sensors and systems used in aircraft and vehicles.
  • Field Service and Maintenance: Calibrating and troubleshooting pressure instruments in the field.

The Additel ADT762W is a powerful and versatile tool that simplifies the calibration process, improves efficiency, and ensures accurate and reliable results. Its innovative design and advanced features make it an indispensable tool for professionals in various industries.

Additional information

Brand

Additel

Pressure Specifications

Model

Specification

 

ADT762W Automated Hydraulic Pressure Calibrator

Pressure Range 15-10,000 psi (1-700 Bar) and 15-15,000 psi (1-1,035 Bar)
Range Selection Manual 3K psi, 10K psi or auto-range for the GP10K and Manual 6k psi, 15K psi or auto-range for the GP15K
 

 

Accuracy

GP15K: 0~15,000 psi, 0.01%FS [2]

GP15K: 0~6,000 psi, 0.01%FS [2]

GP15K: 0~15,000 psi, 0.02%FS [2]

GP15K: 0~6,000 psi, 0.02%FS [2]

GP10K: 0~10,000 psi, 0.01%FS [2]

GP10K: 0~3,000 psi, 0.01%FS [2]

GP10K: 0~10,000 psi, 0.02%FS [2]

GP10K: 0~3,000 psi, 0.02%FS [2]

Resolution 7 digits 6 digits
 

Measurement Units

Pa, hPa, kPa, mPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, inH2O@20°C, inH2O@68°F, kgf/cm2, mtorr, torr, lb/ft2, tsi, custom
Maximum External Load Capacity Max: 80 ml@60 MPa,60 ml@100 MPa, 50 ml recommend
Reservoir Max: 350 ml, built-in filter
Control Stability [1] 0.005%FS from 100 to 15,000 psi
Stability Duration > 5 min
Pressure Module Built-in one module with dual range
External Control Pressure Module See the following “External  Control Pressure Module Specification and Compatibility” table
External Measurement Pressure Module All ADT161 pressure modules

[1] Control Stability is based on the range selection or external module

[2] Pressure range capabilities are dependent on model. See ordering info.

Electrical Specifications

Model

Specification

 

Range

 

Resolution

 

Accuracy

 

Note

 

mA Measure

-25 to 25 mA 0.1 μA ± (0.008%RD + 1.0 μA)  

Impedance <10 Ω

-50 to 50 mA 0.1 μA ± (0.008%RD + 2.0 μA)
V Measure -300 to 300 mV 1 μV ± (0.008%RD + 6 μV) Impedance <1 GΩ
 

V Measure

(Auto-ranging)

-5 to 5 V 20 μV ± (0.008%RD + 100 μV)  

 

Impedance >1 MΩ

-12 to 12 V 100 μV ± (0.008%RD + 320 μV)
-30 to 30 V 100 μV ± (0.008%RD + 600 μV)
Loop Power Source 24 V N/A ±1 V 50 mA (Max Loading)
 

mA Source

 

0 to 2.5 mA or

2.5 to 25 mA

 

0-2.5 mA: 0.05μA

0-25 mA:0.5μA

 

0-2.5 mA: 0.008%RD + 0.2 μA

0-25 mA: 0.008%RD + 1.0 μA

 

20 mA @ 1 KΩ

Power Source 16 to 30 V 1 V ±1 V 70 mA (Max Loading)
 

V Source

 

0 to 16 V

 

0.25 mV

 

0.008%RD + 500 μV

 

Pressure Switch

Mechanical Switch, Live Mechanical Switch, NPN Switch, PNP Switch  

N/A

 

N/A

Response time<10 ms.

If the switch is live, voltage range will be (3-30) V

Temperature Compensation 18 °C to 28°C
 

Temperature Coefficient

 

Outside of 18 °C to 28 °C: <± 0.0005%RD + 0.00005%FS/°C

 

Misuse Protection

 

Up to 30 V on any two sockets

Pressure Switch Test
 

HART / PROFIBUS PA