Additel ADT760 Automatic Handheld Pressure Calibrator
- Fully automatic calibrator with built-in pump and controller
- Switchable internal pressure modules for expandable ranges
- Accuracy as low as 0.01%FS
- External pressure modules available (measure only)
- Less than 4 lbs (1.8 kg) for handheld operation
- Source pressure, measure pressure and electrical
- 4 channels
- Optional HART communications
- Optional data logging and task documenting
- Bluetooth (new), USB, and Wi-Fi communication
- ISO 17025-accredited calibration w/data included
- Description
- Additional information
- Specifications
Description
The Additel ADT760 is a powerful and versatile automatic handheld pressure calibrator designed to simplify and streamline pressure calibration processes. This compact and lightweight device offers a range of features and benefits, making it an essential tool for professionals in various industries.
Key Features:
- Built-in Pump and Controller: The 760 features a built-in pump and controller, eliminating the need for external pressure sources.
- Switchable Internal Pressure Modules: The device allows for easy switching between internal pressure modules, covering a wide range of pressure ranges.
- High Accuracy: Delivers exceptional accuracy up to 0.01% FS, ensuring precise calibration results.
- External Pressure Module Compatibility: Supports the use of external pressure modules for additional measurement capabilities.
- Compact and Portable Design: Lightweight and portable for easy transportation and use in various settings.
- Intuitive User Interface: The user-friendly interface simplifies operation and configuration.
- Data Logging and Reporting: Records calibration data and generates comprehensive reports.
- HART Communication: Supports communication with HART-enabled devices for seamless calibration.
Applications:
- Calibration Laboratories: Calibrating a wide range of pressure instruments, including transmitters, gauges, and switches.
- Manufacturing Industries: Testing and calibrating pressure sensors and equipment in production processes.
- Oil and Gas Industry: Calibrating pressure instruments used in oil and gas exploration and production.
- Aerospace and Automotive Industries: Calibrating pressure sensors and systems used in aircraft and vehicles.
- Field Service and Maintenance: Calibrating and troubleshooting pressure instruments in the field.
The Additel ADT760 is a powerful and versatile tool that simplifies the calibration process, improves efficiency, and ensures accurate and reliable results. Its innovative design and advanced features make it an indispensable tool for professionals in various industries.
Additional information
Brand | Additel |
---|
Pressure Specifications
Specification | 760-LLP | 760-D | 760-MA |
Max Pressure Range | ±30 inH2O (75 mbar) | -12.5 to 35 psi
(-0.86 to 2.5 bar) |
-12.5 to 300 psig
(-0.86 to 20 bar.g) |
Accuracy | 0.05%FS[1] | 0.01%FS or 0.02%[1] | 0.01%FS or 0.02%FS[1][3] |
Stability | 0.005%FS[2] | 0.005%FS[2] | 0.005%FS[2] |
Pressure Type | Differential, Gauge | Differential, Gauge | Gauge, Absolute |
Over Range Indication | 120% | ||
Resolution | 6 digits | ||
Measurement Units |
Pa, hPa, kPa, mPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, mH2O@68°F, inH2O@20°C, kgf/cm2, mtorr, torr, lb/ ft2, tsi, custom | ||
Barometric Accuracy | N/A | N/A | 55 Pa[4] |
Connection | Barb fitting | Hose, Â 5 ft (1.5 m), with built-in filter to
1/4BSPF, 1/4NPTF, and M20F adapters |
Hose, Â 5 ft (1.5 m), with built-in filter to
1/4BSPF, 1/4NPTF, and M20F adapters |
Pressure Output Rate | <30 Seconds (30 inH2O / 100 ml) | <10 Seconds (35 psi / 5 ml) | <90 Seconds (300 psi / 5 ml) |
[1] FS specification applies to the span of the module range.
[2] Stability based on FS of the internal pressure module. Stability is 0.005%FS or 0.05 pa whichever is greater. Internal module is switchable. [3] Specification based on gauge measurement. An additional 55 pa uncertainty will need to be included when measuring in absolute mode.
[4] 55 Pa uncertainty (k=2) includes calibration uncertainty, linearity, and long-term stability (<30 Pa per year). Barometer range of 60 to 110 kPa.
|
Electrical Specifications
Specification | Range | Resolution | Accuracy | Note |
mA Measure | ± 30 mA | 0.0001 mA | 0.01%RD + 1.5 μA | Impedance <10 Ω |
V Measure | ± 30 V | 0.0001 V | 0.01%RD + 1.5 mV | Impedance >1 MΩ |
mA Source | 24 mA | 0.001 mA | 0.01%RD + 1.2 μA | 20 mA @ 1 K |
Loop Power Source | 24 V | N/A | ± 1 V | 50 mA (Max Loading) |
Pressure Switch | Open, close. Support for mechanical switches and NPN/PNP digital switches. | |||
Temperature Compensation | 8 °C to 38 °C |