Fluke P3200 Hydraulic Deadweight Tester

  • Pressure ranges to 10,000 psi (700 bar)
  • 0.015% of reading accuracy standard (0.008% optional)
  • Single or dual piston formats
  • Built-in hand pumps standard
  • Mounted spirit level with adjustable feet
  • Units can be trimmed to local gravity FOC

Description

The Fluke Calibration P3200 Series are high-accuracy, hydraulic deadweight testers designed for calibrating a wide range of pressure instruments.

Key Features:

  • Pressure ranges to 10,000 psi (700 bar)
  • 0.015% of reading accuracy standard (0.008% optional)
  • Single or dual piston formats
  • Built-in hand pumps standard
  • Mounted spirit level with adjustable feet
  • Units can be trimmed to local gravity FOCHigh Accuracy: Offers exceptional accuracy, typically better than 0.015% of reading, ensuring precise and reliable pressure measurements.
  • Versatility: Available in various configurations to cover a wide range of pressure ranges, from low to high pressures.
  • Rugged and Durable: Designed for demanding laboratory and field environments, with robust construction and reliable performance.
  • Traceability: Calibrations are traceable to national and international standards, ensuring the accuracy and reliability of your measurements.
  • Easy to Use: User-friendly design and intuitive operation simplify the calibration process.

Applications:

  • Calibration Laboratories: Primary standard for calibrating pressure instruments, including gauges, transmitters, transducers, and switches.
  • Industrial Environments: Used for on-site calibration of critical process instrumentation in various industries, such as oil and gas, chemical, and manufacturing.
  • Research and Development: Employed in research and development laboratories for precise pressure measurements and calibrations.

The Fluke Calibration P3200 Series are high-performance pressure standards that provide exceptional accuracy and reliability for critical pressure calibration applications.

Additional information

Brand

Fluke Calibration

Specifications: Fluke Calibration P3200 Hydraulic Deadweight Tester

Pressure Ranges
Distilled water operated Ranges to 10,000 psi (700 bar)
Accuracy
Accuracy1 ±0.015% of reading (±0.008% optional)
1. Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range.
Materials of Construction
Standard weight material Series 3 non-magnetic austenitic stainless steel
Weight density 7.8 g/cm³
Optional fractional weights Solution heat treated aluminum
Weight density 2.7 g/cm³
Piston material Tungsten carbide with nickel binder
Density- 15.0 g/cm³
Cylinder material Hardened martensitic steel
Tungsten carbide (water above 500 psi, 35 bar)
Thermal coefficients of expansion Oil piston/cylinder 16.5 ppm/°C
Water (above 500 psi, 35 bar) 11 ppm/°C
General
Test port adaptors 1/8, 1/4, 3/8 and 1/2
NPT or BSP
Instrument weight 36 lbs (16 kg)
Instrument size (W x D x H) 17.5 x 12 x 8.5 in (440 x 300 x 215 mm)
Mass set weight (typical) 80 lbs (36 kg)
Reservoir volume 150 cc
Screw press displacement 5.5 cc
Pump displacement 4.7 cc per stroke
O-ring seal materials Buna N as standard, Viton and EPDM available
Weight Increments
Minimum standard weight increments Up to 500 psi (35 bar): 1 psi (0.1 bar)
500 to 5000 psi (35 to 350 bar): 10 psi (1 bar)
5000 to 20,000 psi (350 to 1400 bar): 20 psi (2 bar)
Model P3112: 1 psi (0.1 bar)
Optional fractional weights Up to 500 psi (35 bar): 0.1 psi (0.01 bar)
500 to 5000 psi (35 to 350 bar): 1 psi (0.1 bar)
5000 to 20,000 psi (350 to 1400 bar): 2 psi (0.2 bar)
Operating Fluids
Operating fluids Distilled or deionized water
Options
PressCal software Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments
Fluids Instruments can be supplied for use with Skydrol or brake fluids. Due to the aggressive nature of these fluids the standard nitrile seals and the acrylic reservoir tube will be replaced by either Viton or EPDM seals (as applicable) and an aluminum reservoir tube. For these or any other specialty fluid applications, please consult the factory.

 