Features

The 780 Pressure Controller is considered to handle a very wide range of applications which may normally require a pressure calibrator. Built-in capability, includes current and voltage measurement capability, 24 volt loop power, HART® and Profibus PA communication, switch measurement capability, selectable analog output and much more.

Specs

Measurement Specifications

Model Number ADT780-1K ADT780-3K
Pressure range -14.5 to 1,000 psi (-0.95 to 70 bar) -14.5 to 3,000 psi (-0.95 to 200 bar)(1)
Control stability 0.003% FS
Precision (includes 1 year stability) See pressure range table
1 Year stability 0.01% FS standard accuracy sensors 0.005% FS precision accuracy sensors
Media clean Gas
Over-range indication 103% to 120% (based on sensor)
Resolution 4, 5, 6, or 7 digits (user selectable)(2)
Pressure type Gauge, Absolute
Warm up time 15 minutes

Typical pressure settling time (within 0.003%FS)

20 seconds (10%FS pressure change assuming 50ml test volume)
Measurement units Pa, hPa, kPa, MPa, bar, mbar, psi, mmHg@0°C,cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, kg/m2, kg/cm2, mtorr, torr, atm, lb/ft2, tsi, user selectable
Minimum control pressure(3) 0.0001 psi 0.001 psi

(1). Gas booster or HP gas supply required to reach 3,000 psi (200 bar). Booster ordered separatel.
(2). Digit resolution for precision model only.
(3). Dependent on pressure module.

Barometric Measurement Specifications

Sensor/Mode Gauge Absolute
ADT160A-02-CPXX N/A 40Pa(1)
ADT160A-01-APXXQ 3Pa(2) N/A

 

Barometer range (60~110)KPa, the accuracy is 40Pa (11Pa is optional)
(1). 40Pa uncertainty (k=2) includes calibration uncertainty, linearity and long term stability (< 30Pa per year). Barometer range of 60 to 120KPa.
(2). Combined linearity, hysteresis and repeatability. Add 3Pa when used in gauge mode. When using the ADT160A-01-APXXQ sensors with the ADT780 controller in gauge mode, regular zeroing will realign the barometric reading to provide the most accurate result.

 

Electrical Measure Specifications

Specification Range Resolution Accuracy
Volts DC -30 to 30 V 0.1 mV ±0.01% rdg + 1.5 mV
-300 to 300 mV 1 uV ±0.01% rdg + 15 uV
Current DC -30 to 30 mA 0.1 uA ±0.01% rdg + 1.5 uA
Switch test If the switch has detected voltage, the range is from 3 – 24 V
DC 24V output 24V ± 0.24V, max 30 mA

Physical Specifications

Specification ADT780-1K & ADT780-3K
Power 100 – 240V, 50/60 Hz
Pressure ports G1/8 F
Storage temperature -20°C to 70°C
Operating environment 0-95% RH non-condensing
Display 7 inch (17.8 cm) color, touch screen display
Weight 33 lbs (15 kg)
Dimensions(DWH) 16.5 x 17.3 x 5.2 inch (419 x 440 x 132 mm)
Mounting Standard desktop, optional rack mount
Shock 4G
Vibration 1G 10Hz~500Hz

Other Specifications

WiFi specifications 802.11 b, g, and n
Stability indicator User selectable
Typical slew rate Typically 20 seconds
Operating modes Control, measure, and vent

Display modes
Controller – show pressure indication and control
Calibrator – shows pressure indication, control, and electrical measurement
Localization English, Chinese (simplified), German, Spanish, French, Italian, Portuguese, Russian, Japanese
Conformity CE
Calibration certification NIST-traceable certificate of calibration with data included

 

Electric Pump and Booster

ADT780-1K-EPUMP Electric Pump

Generate Vacuum to 1,000 psi(700bar)
Built-in filter and liquid trap

Pressure range -13.5 to 1,080 psi (-0.95 to 75 bar)
Weight

69.6 lbs (31.6 kg)

Size(WHD) 17.3 X 9.3 X 21.7 inch (440 X 235X 550 mm)
Power 100 to 240V
Media Air
Outlet port G 1/8F
Storage temperature -20°C to 70°C
Operating environment 0-90% RH non-condensing
Mounting Standard desktop, optional rack mount
Control operation 2 LED displays with pressure limit settings
Typical max pressure time Approx 2 min with ADT780 Controller

 

ADT780-3K-BST Booster
3,000 psi (200bar) Gas booster specification

Minimum inlet pressure 300 psi (20 bar)
Out pressure range 300 to 3,600 psi (250 bar)
Inlet flow rate 500ml/min
Size (WHD) 17.8 X 7.3 X 23.9 inch (451 X 185 X 607 mm)
Weight 74 lbs (33.6 kg)
Power 100 to 240V
Supply media Nitrogen or clean gas

Models

ADT780 - 3K - CP300 - N - HABP

Note :

ADT 780 - Model ADT780 or ADT780S

3K - Maximum range

3K = 3,000 psi (200 bar)
1K = 1,000 psi (70 bar)

N - Manifold type

N = 1/4NPT male
N2 = 1/2NPT male
B = 1/4BSP male
B2 = 1/2BSP male
M = M20x1.5 male

HABP - High Accuracy Barometer

 

 

Standard Accessories

Model Name & Description Qty
ADT127-X, External Manifold 1 pc
Manifold connection hose 1 pc
Power adapter 1 pc
9022 - test leads for calibrator 2 sets (4pcs)
9020 - short circuit cable 1 set (1 pcs)
9060, Pressure module connection cable 1 pc
Adapter set (adapter to fit the ADT78X pot to male fittings) 1 set (2pc G1/8M to G1/4M, 2pcs G1/8M to Festo release valve)
CD Manual 1 pc
O-rings 2 types
Manual 1 pc
Vacuum/vent tubing 2pcs
NIST traceable calibration certificate 1 pc


Accessories

Model Name Description Price
ADT100-761-N4 Hose test kit, 5 feet flexible hose, 0.156"(Ø4mm) tube to 1/4NPT hand-tight quick connector Get a Quote >>
ADT100-761-B4 Hose test kit, 5 feet flexible hose, 0.156"(Ø4mm) tube to 1/4BSP hand-tight quick connector Get a Quote >>
ADT106-N External pressure manifold (with two 1/4NPT hand-tight quick connectors) Get a Quote >>
ADT106-B External pressure manifold (with two 1/4BSP hand-tight quick connectors) Get a Quote >>
ADT106-M External pressure manifold (with two M20X1.5 hand-tight quick connectors) Get a Quote >>
ADT160A Pressure modules Get a Quote >>
ADT121-X External pneumatic pressure manifold 3,000 psi (200 bar) four hand-tight quick connector Get a Quote >>
ADT780-3K-BST 3,000 psi (200 bar) gas booster Get a Quote >>
ADT780-1K-EPUMP 1,000 psi (70 bar) electrical pump Get a Quote >>
ADT780-VACPUMP Vacuum pump Get a Quote >>
9050 RS232 to USB adapter Get a Quote >>
ADT100-FLT-600 Liquid trap Get a Quote >>
9912-780 Shipment case for the ADT780 and ADT 780S controller Get a Quote >>
9506 Additel/Pcal Manual calibration software Get a Quote >>
9530 Additel/ACal Automated calibration software, task and asset management Get a Quote >>
9245-780 Rack mount kit Get a Quote >>
ADT100-780-Kit Adapter set, total 9 pcs
  • 3 pcs, G1/8 M to 1/4NPT M
  • 3[cs, G1/8 M to G1/4 M
  • 3 pcs, G1/8 M to M20x1.5M