CONTACT US

This email address is being protected from spambots. You need JavaScript enabled to view it.

 

Jakarta Office : 021-45850618


   Surabaya Office : 031-8421264

 

Additel 160 Intelligent Digital Pressure Modules

Features

The 160A series Digital Pressure Modules (IDPM) are specially designed as external pressure module for 761 automated pressure calibrator, 222A Multifunction Process Calibrators and 223A Documenting Process Calibrators.

Application

  • Gauge calibration
  • Precision pressure measurement
  • Pressure transmitters calibration
  • Pressure switch testing
  • Safety relief valve resting
  • Pressure regulator testing
  • Hydraulic leak testing

Specs

• Environmental
Compensated Temperature: 14°F to 122°F(-10°C to 50°C )
Operating Temperature*: 14°F to 122°F(-10°C to 50°C )
Storage Temperature: -4°F to 158°F (-20°C to 70°C )
Humidity: <95%

• Pressure Connections
1/4NPT male, 1/4BSP male, M20X1.5 male
Other Connections available per request

• Enclosure
Case material: SST
Wetted parts: 316L SS
Dimension: Ø1.3" x 4.84" (Ø33mm x 123mm)
Weight: 0.99 lb (0.4kg)

• Pressure Range and Accuracy

Gauge Pressure [1]

P/N

Pressure Range

Media[2]

Accuracy(%FS)

Burst
Pressure

(psi)

(bar)

V15

-15

-1.0

G

0.02

GP2

2

0.16

G

0.05

GP5

5

0.35

G

0.05

GP10

10

0.7

G

0.02

GP15

15

1.0

G

0.02

GP30

30

2.0

G

0.02

GP50

50

3.5

G , L

0.02

GP100

100

7.0

G , L

0.02

GP300

300

20

G , L

0.02

GP500

500

35

G , L

0.02

GP600

600

40

G , L

0.02

GP1K

1,000

70

G , L

0.02

GP2K

2,000

140

G , L

0.02

GP3K

3,000

200

G , L

0.02

GP5K

5,000

350

G , L

0.02

GP10K

10,000

700

G , L

0.02

GP15K

15,000

1,000

G , L

0.05

GP20K

20,000

1,400

G , L

0.05

1.5×

GP25K

25,000

1,600

G , L

0.1

1.5×

GP30K

30,000

2,000

G , L

0.1

1.5×

GP36K

36,000

2,500

G , L

0.1

1.5×

GP40K

40,000

2,800

G , L

0.1

1.1×

Note:
[1]. Sealed gauge pressure for above 1000 psi
[2]. G=Gas, L=Liquid

Compound Pressure

P/N

Pressure Range

Media

Accuracy(%FS)

Burst
Pressure

(psi)

(bar)

CP2

±2

±0.16

G

0.05

CP5

±5

±0.35

G

0.02

CP10

±10

±0.7

G

0.02

CP15

±15

±1

G

0.02

CP30

-15 to 30

-1 to 2

G

0.02

CP100

-15 to 100

-1 to 7

G, L

0.02

CP300

-15 to 300

-1 to 20

G, L

0.02

CP500

-15 to 500

-1 to 35

G, L

0.02

CP600

-15 to 600

-1 to 40

G, L

0.02

CP1K

-15 to1,000

-1 to 70

G, L

0.02

CP2K

-15 to 2,000

-1 to 140

G, L

0.02

CP3K

-15 to 3,000

-1 to 200

G, L

0.02

CP5K

-15 to 5,000

-1 to 350

G, L

0.02

CP10K

-15 to 10,000

-1 to 700

G, L

0.02

 

Absolute Pressure

P/N

Pressure Range

Media

Accuracy(%FS)

Burst Pressure

(psi)

(bar)

AP5

5

0.35

G

0.1

AP10

10

0.7

G

0.1

AP15

15

1.0

G

0.1

AP30

30

2.0

G

0.1

AP50

50

3.5

G

0.1

AP100

100

7.0

G , L

0.05 (0.1)

AP300

300

20

G , L

0.05 (0.1)

AP500

500

35

G , L

0.05 (0.1)

AP1K

1,000

70

G , L

0.05 (0.1)

AP3K

3,000

200

G , L

0.05 (0.1)

AP5K

5,000

350

G , L

0.05 (0.1)

 

Precision Quartz Sensors[1] - Absolute Pressure

P/N

Pressure Range

Media

Accuracy

Burst
Pressure

psia

bar.a

AP45Q

0 to 45

2.0

G

0.005%rdg+0.005%FS

AP100Q

0 to 100

7.0

G

0.005%rdg+0.005%FS

AP400Q

0 to 400

28

G

0.005%rdg+0.005%FS

AP500Q

0 to 500

35

G

0.005%rdg+0.005%FS

AP1KQ

0 to 1,000

70

G

0.005%rdg+0.005%FS

AP2KQ

0 to 2,000

140

G

0.005%rdg+0.005%FS

AP3KQ

0 to 3,000

200

G

0.005%rdg+0.005%FS

[1] Contact Additel for other range options.

Differential Pressure

P/N

Pressure Range

Media

Accuracy(%FS)

Burst
Pressure

(inH2O)

(mbar)

DP1

±1

±2.5

G

0.05

100×

DP2

±2

±5.0

G

0.05

100×

DP5

±5

±10

G

0.05

50×

DP10

±10

±25

G

0.05

20×

DP20

±20

±50

G

0.05

20×

DP30

±30

±75

G

0.05

15×

DP50

±50

±160

G

0.05

DP150

±150

±350

G

0.02

DP300

±300

±700

G

0.02

* One year accuracy (including yearly stability) except DP1 and DP2 range
which is 0.05%FS calibration accuracy and 0.05%FS yearly stability.

 

Models

ADT160A - 02 - GP300 - PSI - N

Note :

ADT 160 - Model

02 - Accuracy
(02 = 0.02% of full scale)
(05 = 0.05% of full scale)
(10 = 0.1% of full scale)

GP300 - Pressure range P/N (see pressure range table

PSI - Range type
PSI = range by psi)
(BAR = range by bar)

N - Pressure port type
(N = 1/4 NPT male)
(N2 = 1/2 NPT male)
(B = 1/4 BSP male)
(M = M20 x 1.5 male)

 

Accessories

Model Name Description Price
9060 Additel Pressure Module Connection Cable Get a Quote >>

 